User's Manual

RXiBreeze PAP System User Manual
27
Problem
Possible Cause
Action
Air pressure in my
mask seems too high
(it feels like I am
getting too much air).
Ramp may be
turned off.
Use the Ramp Time option.
Air pressure in my
mask seems too low
(it feels like I am not
getting enough air).
Ramp may be in
progress.
Wait for air pressure to build up or turn
Ramp Time off.
I have stopped
therapy, but the
device is still blowing
air.
Device is cooling
down.
Device blows a small amount of air in
order to avoid condensation in the air
tubing. It will stop automatically after a
few minutes.
The device’s display
is erratic (crash blank
or blue screen).
The device has
been dropped or
mishandled.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
My water tank is
leaking.
Water tank may
not be assembled
correctly.
Water tank may
be damaged or
cracked.
Check for damage and reassemble
the water tank correctly. Contact your
care provider for a replacement.
Key exception
(non-responsive or
insensitive).
Program crashes
or key
misalignment.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
The knob is
insensitive.
Encoder is
damaged
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
12 Cleaning and Maintaining
The machine needs to be cleaned and disinfected regularly so that these germs and
contaminants do not grow inside of your equipment and make you sick. Dust and dirt
can also cause problems with the machine, making it more likely to break or need
replacement. Please strictly follow the cleaning and disinfection process below: