User's Manual

RXiBreeze PAP System User Manual
28
12.1 Cleaning and Disinfection Process and Schedule for Device Surface
1 Cleaning
WARNING: To avoid electrical shock, always unplug the power cord from the wall
outlet before cleaning the device. Do not immerse the device in any fluids.
1. Please power off and unplug of the device, and throughout cleaning the casing of
the device to removal of soil, stain or oil by using a soft cloth slightly dampened
with neutral detergent (such as, 5% soapsuds)
2. Use a soft cloth slightly dipped in water to wipe and rinse the surface that was
cleaned with soapy water in the previous step for 5 minutes, and repeat the above
steps at least additional 2 times, three times in total.
3. Visual inspection the casing of the device, ensure there is no soil, stain or oil is
seen. If there is determined not to be visually cleaned at the end of the cleaning
steps. You should either repeat the relevant previous cleaning steps, so that a
visible soiled device is not used again.
4. Wait about at lest 1 minutes for drying completely before plugging in the power
cord.
2 Disinfection
1. After the cleaning procedure, use an 75% ethyl alcohol swab or cotton swab
moistened with 75% ethyl alcohol to disinfect the device surface for about 5
minutes.
2. Repeat the above steps at least 3 times, then wipe surface dry with a clean cloth,
and wait about 1 minute for drying after disinfection process.
3 Drying
Allow sufficient time for the device to air dry completely.
Note: Drying is not required after cleaning if disinfection is continued
immediately.
4 Inspection
Perform a visual inspection of the device casing. If any visible deterioration is apparent
(cracking, crazing etc) discontinue use and contact your care provider or Resvent.
12.2 Cleaning and Disinfection Process and Schedule for (Water tank, air outlet
adaptor, air outlet of water tank)
1 Cleaning