User's Manual

RXiBreeze PAP System User Manual
29
1. Remove all detachable parts from the device. Make a solution of a neutral
detergent (such as, soapsuds) and water.
2. Soaked all components for about 5-10 minutes. Agitate the component in the
cleaning solution to ensure there are no air bubbles
3. Clean the inside and outside of all components with a soft bristle brush while
soaking in a detergent solution. Pay particular attention to all crevices and cavities
4. Soaking all the components in the solution,and through rinse each component in
the rinsing water to ensure there are no air bubbles. Repeat the rinse procedure
two additional times using fresh water for a total of three rinses.
5. Visual inspection to ensure there is no soil, stain or oil is seen. If there is
determined not to be visually cleaned at the end of the cleaning steps. You should
either repeat the relevant previous cleaning steps, so that a visible soiled device is
not used again.
6. Shake the component to remove excess water, and allow the component to air dry
rather out of direct sunlight.
Note: Failure to clean the component as indicated may result in inadequate
disinfection.
2 Disinfection
WARNING: The disinfection process must be carried out after cleaning process.
1. After the cleaning procedure, immerse the component in a water bath, and make
a solution of a 3% hydrogen peroxide, Soak the components in the solution at
room temperature for at least 30 minuets. Rinse and agitate the component in the
solution for 1 minutes
2. Soaking all the components in the solution,and rinse and agitate the component in
a fresh water thoroughly for at least 3 times, 10 minutes per time and hang up to
allow to air dry but out of direct sunlight or heat.
3. Perform a visual inspection of the device components. If any visible deterioration
is apparent (cracking, crazing etc) discontinue use and contact your care provider
or Resvent.
12.3 Cleaning Schedule
To avoid prolonged exposure to dusty and humid environment, resulting in impaired
performance and reliability, the user must clean the device regularly. The clean interval
of the device and accessories, please refer the below table:
Interval
Action