Safety Specifications
Class II
Type BF
Ingress Protection: IP22
Air filter
Air filter: Filter Efficiency: >75% (7 micron dust)
Pressure
Setting Range: 4-20 cmH
2
O
Max Single Fault Steady Pressure: 40 cmH
2
O
Static Pressure Accuracy: ±0.5 cmH
2
O (Expanded uncertainty:1.87%)
Dynamic Pressure Accuracy: ±1cmH
2
O (Expanded uncertainty:2.55%)
Pressure monitoring Accuracy: ± (2%*full scale reading + 4%*actual reading)
Flow
Test pressure (cmH
2
O)
4
8
12
16
20
19mm
breathing
tubing
Measured pressure at the
patient connection port
(cmH
2
O)
2.9
7.1
11.0
15.1
19.1
Average flow at the patient
connection port (L/min)
154.3
146.4
135.7
125.3
114.1
15mm
breathing
tubing
Measured pressure at the
patient connection port
(cmH
2
O)
2.9
6.9
10.8
14.8
18.7
Average flow at the patient
connection port (L/min)
141.4
132.7
124.2
114.3
85.4
Humidifier
Water capacity: 290 ml (MAX Water Level)
Humidity: >10 mg/L BTPS (Within the set pressure range)