User's Manual

RXiBreeze PAP System User Manual
3
Do not operate the device in direct sunlight or near a heating appliance because
these conditions can increase the temperature of the air coming out of the device.
Contact your health care professional if symptoms of sleep apnea recur.
If you notice any unexplained changes in the performance of this device, if it is
making unusual or harsh sounds, if it has been dropped or mishandled, if water is
spilled into the enclosure, or if the enclosure is broken, disconnect the power cord
and discontinue use. Contact your home care provider.
Repairs and adjustments must be performed by Resvent-authorized service
personnel only. Unauthorized service could cause injury, invalidate the warranty, or
result in costly damage.
Do not use any accessories, detachable parts, and materials not recommended by
Resvent. Incompatible parts or accessories can result in degraded performance.
The Health Industry Manufacturers Association recommends that a minimum
separation of 16cm be maintained between a wireless phone and a pacemaker to
avoid potential interference with the pacemaker.
Use only power cords supplied by Resvent for this device. Use of power cords not
supplied by Resvent may cause overheating or damage to the device and may
result in increased emissions or decreased immunity of the equipment or system.
The device should not be used while stacked or in close approximation to other
non-approved devices.
Use only approved cables and accessories. Misuse may affect EMC performance
and should be avoided.
Inspect the tubing for damage or wear. Discard and replace the tubing as necessary.
Periodically inspect electrical cords and cables for damage or signs of wear.
Discontinue use and replace if damaged.
To avoid electrical shock, always unplug the power cord from the wall outlet before
cleaning the device. DO NOT immerse the device in any fluids.
Be sure to route the power cord to the outlet in a way that will prevent the cord from
being tripped over or interfered with by chairs or other furniture.
This device is activated when the power cord is connected.
For safe operation when using a humidifier, the humidifier must always be
positioned below the breathing circuit connection at the mask. The humidifier must
be level for proper operation.
Nebulization or humidification can increase the resistance of breathing system filters
and the operator must monitor the breathing system filter frequently for increased
resistance and blockage to ensure the delivery of the therapeutic pressure.
Please check whether there is water in the device before use. The maximum fill level
is 290 mL.
Failure to use a mask or accessory that minimizes re-breathing of carbon dioxide or
permits spontaneous breathing can cause asphyxiation.
Do not connect breathing tubes or accessories with any humidifier and ventilator that
are not specified for use with these breathing tubes or accessories.
Do not cover or heat the breathing tube with anything influent the patient end
temperature.
Do not use this device outside the specified ambient temperature range or humidity
range. The humidity performance of the device can be compromised when used
outside the specified ambient temperature range or humidity range.
No modification of this equipment is allowed.
Parts of the equipment are not serviced or maintained while in use with the patient.
The PATIENT is an intended OPERATOR.
The Patient can safely use therapy functions of the equipment and this equipment
shall not be serviced or maintained while in use with the patient.