User`s manual
Section 5  Alignment and Maintenance 
  Balance Test Switch 
 VR1 VR2  VR3 VR4 
Meter Zero Cathode Bal. Plate Bal. Amp. Bal. 
All adjustments should be performed by a qualified technician. 
5.1 Gain Reduction Meter Zero 
Use a small Xcelite Greenie or precision adjustment tool to make internal adjustments. 
The Gain Reduction Meter can be adjusted to zero to compensate for normal aging of 
components over time. With no input applied and the release control set to 50, adjust 
VR1 so that the meter sits exactly at the 0 dB mark. The Release control can have a 
slight effect on the zeroing. The asymmetry switch should have no effect on the meter 
zero. If it does, the 6AL5 tube may be at fault. 
5.2 Internal Balance Test Switch 
The Internal Balance Test Switch has three settings. For normal operation, it must be in 
the center position. The Top position (static test) is used to set VR2 and VR4. The 
Bottom position (dynamic test) is used to set VR3. The Meter Zero VR1 may have to be 
trimmed after balancing. 
5.3 Balancing the 6BC8 Gain Reduction Stage 
Poor balance will be apparent on compression attack, causing a ripple sound or a thump. 
Set the Meter Selector to Output. Set Interstage Switch to IN. Set Release to 100 (fast). 
Set Output Level to 100 (full). Set Balance Test Switch to Up (Static Test). 
Adjust VR2 for minimum indication on the VU meter. 
Set Balance Test Switch to the Bottom position. 










