Marine Instruments User Manual
Retrotec Energy Innovations, Ltd.
9
DM-2A Mark II 2008-02-22
• Velocity  Displays the air velocity. This mode requires that a pitot tube be 
installed.
• Velocity Flow  Displays the flow calculated from the velocity and cross-
sectional area of the measured duct.
 Range COnfig  Press the Range COnfig keypad button to cycle through the 
available flow ranges for the current test fan.
 Setup  Press the Setup keypad button to access the main Setup menu. The Setup 
menu allows you to control and customize the gauge to meet your specific 
testing needs. From the Setup menu you can choose to enable or disable devices 
and/or modes, set the ‘n’ (@-mode extrapolation) value, set power down options, 
and enter area and volume units and many other options (see Enabling and 
Disabling Modes in Section 2, Working with the Setup Screen). Any changes you 
make in the Setup menu are permanently saved. 
 BaSeline  The Baseline feature enables you to measure the background (baseline) 
pressure in the room or enclosure before you begin testing. Once measured, 
your gauge will automatically subtract this baseline pressure from your house-
pressure (PrA) measurements. Press the BaSeline keypad button once to initiate 
the baseline pressure reading test. Your gauge will display ‘Acquiring’ and will 
begin to sample the background pressure. While acquiring the pressure values, 
the gauge will continuously display the updated average pressure and the 
duration of sampling on the screen. Pressing the button a second time to enable 
the Baseline mode with the average pressure that was observed. From this point 
on and until the feature is disabled, the baseline pressure displayed in the top left 
corner of the screen will automatically be subtracted from the PrA readings. Press 
the exit keypad button to cancel out of Baseline mode.
 tiMe avg  Press the tiMe avg button to cycle through the available time 
averaging settings or turn time averaging off. Time averaging causes the gauge 
to display the average readings over the time interval. The current time averaging 
value is displayed below the Time label on the gauge. Time averaging may be set 
to 1, 2, 4, 8, 10, or 20 seconds, or 1 or 2 minutes.
DM-2A Mark II Dual-Channel Pressure Gauge User Manual
1: Getting Started