User Manual

Operating the machine
34
Fig. 23: Using the opening aid
Pos: 6.20 /00004 Warnhinweis e/H0056 HINW EIS Cryomill PTFE Mahlbecher n ur mit Zimmert emperatur einsetze n - Nicht vorkühlen @ 4\ mod_1314082 879240_9.doc @ 25509 @ @ 1
NOTE
1.H0017
Damage to the PTFE grinding jar
The threading on the grinding jar can become damaged.
When it is being inserted into the device, the PTFE grinding jar must
be at room temperature and must not be pre-cooled.
Pos: 6.21 /00005 Ü berschriften/ 1.1.1 Überschrift en/111 Kühl mantel Abdichtung einsetzten @ 2\mod_12663 14975785_9.doc @ 18787 @ @ 1
5.5.2 Inserting the cooling jacket sealing plug
Pos: 6.22 /00010 Bedienungsa nleitungen Kapitels ammlungen/Cr yoMill II/0007 B edienung/0011 Cr yoMill II Mod ul Kühlmantela bdichtung einsetzen @ 5\ mod_13438292 44950_9.doc @ 33257 @ @ 1
NOTICE
Blocked pipe
Ice formation
Ice or impurities can block pipes and thereby cause malfunctions.
Insert the cooling jacket seal (KA) as soon as you are no longer using
the device.