Summit SB-1000 ® Bariatric Wheelchair Scale Installation & Operation Manual 106992
Contents 1.0 Introduction and System Overview.............................................................................................. 1 2.0 Scale Installation......................................................................................................................... 2 2.1 2.2 2.3 2.4 2.5 2.6 3.0 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Site Preparation . . . . . . . . . . . . .
ii Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
1.0 Introduction and System Overview The Summit® SB-1000 Bariatric Wheelchair Scale is a heavy-duty scale that easily accommodates wheelchairs with its large platform size and easy-access ramp. It provides exceptional performance in applications where typical wheelchair scales are not large enough to meet the needs of the patient. The Summit SB-1000 Bariatric Wheelchair Scale is a fully electronic, low profile floor scale that measures 36 in x 36 in (.91 m x .
2.0 Scale Installation The following sections describe the correct installation procedures when installing the Summit SB-1000 Bariatric Wheelchair Scale. 2.1 Overview Standard installation of the Summit SB-1000 Bariatric Wheelchair Scale consists of the following steps: 1. Select a site 2. Check levelness and smoothness of site 3. Unpack the scale, ramp and indicator 4. Adjust the four feet on the scale 5. Set the ramp next to the scale so that two of the scale feet set in the ramp foot keeper plates.
2.4 Bariatric Wheelchair Scale Feet Assembly and Adjustment The scale feet are shipped attached to the scale. If the feet have to be adjusted, use the following instructions. Screw one foot into each load cell and turn all the way in until the foot touches either the load cell or the underside of the deck. Then unscrew each foot three complete turns. Place a spirit level on the deck.
servicing or cleaning. SCALE FLOOR LOAD CELL CABLE Figure 2-3. Load Cell Protection 8. When the interface cable is protected and in its final position, complete connections to the indicator. If necessary, trim corners as described inSection 3.2 on page 5. 9. Check all strain relief fittings for tightness. 10. Put the cover back onto the junction box assembly and slide the junction box back into the floor scale cutout. 11. Secure the Summit SB-1000 side plate. 2.
3.0 Adjustments and Calibration The following sections describe adjustments that need to be made to the Summit SB-1000 Bariatric Wheelchair Scale. 3.1 Mechanical Adjustments To accommodate minor floor unevenness, the scale feet can be used to adjust scale height up or down a fraction of an inch. Adjust the feet by hand (lift the scale corner slightly with a pry bar) until all feet are contacting the floor equally.
With all potentiometers at full signal, place the test weight over one corner and record the indicated weight. Repeat the process for each of the other three corners. The load cell with the lowest corner reading is used as a reference point and will not be trimmed. Next, place the test weight over one of the other three corners and use that cell’s potentiometer to adjust the cell output down to the reference cell output. Repeat this procedure with the other two high corners.
4.0 4.1 Indicator Installation and Configuration Indicator Installation The Summit SB-1000 Bariatric Wheelchair Scale has several connections on the back of the indicator. Connection points include: • AC power connection • RS-232 port connection • Scale connection port • Adaptor storage connection AC Power Connection The Summit SB-1000 Scale has a 120 VAC adaptor or 230 VAC adaptor to use when power is readily available. The AC power adaptor plugs into the back of the indicator as shown in Figure 4-1.
Load Cell Connections To gain access to the load cell connection point, remove the four back retaining screws as shown in the left hand side of Figure 4-3. Figure 4-2 illustrates where the load cell connection point is. Scale connection (shown in Figure 4-2 with the plastic cover removed) Figure 4-2. Load Cell Connection Point 4.
4.3.1 Configuration Mode To get into the configuration mode, turn the scale off and remove the configuration jumper as shown in Figure 4-3. Turn the scale on. While Start is displayed, press and hold the Kg-Lb key until IDENT appears on the display. To change from one parameter to the next, press the REWEIGH key once. To change the value of the parameter, use the Kg-Lb key. From the SAVE phase: to save the configuration data, press the REWEIGH key.
The following table lists the various display messages and sequence when setting up the scale. NOTE: <-> means that you can toggle between two values. Step Function 1 With the scale turned off, remove the confiiguration jumper as shown in Figure 4-3. 2 With the scale turned off, simultaneously press the On-Off/Zero and Kg-Lb key. Display Available Parameters Allows the scale to enter into either configuration or calibration mode.
Step Function Display Available Parameters 12 Message style on weight algorithm MESS <-> WEIGH This displays the message that will show on the indicator display. To change the message, press the Kg-Lb key. WEIGH, LIVE, -----To advance to the next step, press the REWEIGH key. 13 This allows for setting the baud rate of the RS-232 connection BAUd <-> Default - 9600 To change the baud rate, use the numeric data entry (See “Enabling Configuration or Calibration Modes” on page 8.
Step Function Display Option 5 allows for live or dynamic weighing OP5 = 0 Hold the Kg-Lb key for browsing through the options. Double click on the Kg-Lb key to change the options value or to advance to the next step, press the REWEIGH key. 0=Disable 1=Enable Option 6 allows you to either enable or disable the hold function on the scale OP6 = 1 Hold the Kg-Lb key for browsing through the options.
Step 15 Function Set Option defaults Display Available Parameters Set < - > 1 To change, press the Kg-Lb key. 0 - default Options = (OPx) 1 = US default (US defaults are OP1-1, OP2-0, OP3-0, OP4-0, OP5-0, OP6-1, OP-7-0, OP8-1, OP9-1, OP10-1). 2 = European 1 default 3 = European 2 default 9 = Used if OP1 through OP10 have been changed to something other than the factory default settings. Note: Always set to 1 or 9 for US indicators. To advance to the next step, press the REWEIGH key.
5.0 Scale Calibration Before you can calibrate the scale, verify and set all scale parameters which are noted in “Indicator Installation and Configuration” on page 7. Remove the configuration/calibration jumper (see page 8). Turn the scale on. While StArt is displayed, press and hold both the Kg-Lb key and the REWEIGH keys until CAL is displayed. To do a calibration on the scale, the scale must be turned off. NOTE: The calibration weight must be no less than 60 lb (28 kg) and no more than 300 lb (135 kg).
6.0 Scale Operation The scales have the capability of performing different operations beyond just calculating weight. The various operating instructions are described below. 6.1 Weighing Use the following steps to weigh a person. 1. Press the On-Off/Zero key to turn on the scale and 0.0 will appear on the display. 2. Ask the patient or person to step onto the scale. The display shows WEIGH, then the person’s weight, and beeps to indicate the end of the weighing process. 3.
increments. Press the BMI key again to accept the inches value. The final height value will be displayed ie: 5-07.5 = 5’ 7.5". 2. If you are weighing in Kgs, the default will be 170.0 cm. Use the up or down arrows to increase or decrease by 0.5 cm increments. 3. To see the patient’s calculated BMI, press the BMI key again. The BMI appears. 4. To cancel the BMI display, press the BMI key. 6.
7.0 RS-232 Communications The scale comes with an RS-232 port which enables weight data to be transmitted to other equipment, such as a computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing System. Figure 4-1 on page 7 shows where the RS-232 connection is. The RS-232 parameters are 9600 baud (selectable in the programming mode), 8 data bits, 1 stop bit, no parity and no handshaking.
An Escape Protocol is where the escape is used to indicate that there is a command following and not just data. Table 7-1 lists a complete list of ESC commands that are used with the scale. Command/ Response ESC Character ESC value with Parameters Description Reading R R This value tells the PC that the scale is sending a reading. Immediately following this will be the value that is sent Example: RE RW0200.5NmE Weight W Wnnn.
3. Double click on the Rswin.exe file and the following screen appears. Click on FILES and in dropdown menu, select LOAD CONFIGURATION as noted in Step 4 Figure 7-1. Rswin Main Screen 4. Click on FILES and in the dropdown menu, select LOAD CONFIGURATION. At this time double click on the file, Inbar.ini. 5. Click on STRINGS and the following screen appears. Figure 7-2.
Examples of what you would see in the HEX screen are shown in Figure 7-3. Figure 7-3. HEX Screen Example Examples of what you would see in the ASCII screen are shown in Figure 7-4 Figure 7-4. ASCII Screen Example 7.4 Sample and Explanation of ESC Protocol When the scale measures weight and sends this over the communications line to the PC, the string will look like this.
DIRECTION OF COMMUNICATION 3#!,% 0# &4$ 3 &4$ 8 &4$ /N &4$ & 3FBEJOH $PNNBOE 6OJU PG .FBTVSF .FUSJD &01 8FJHIU When the user wants to diagnose any problems with the scale, the operator will have to ask the scale to send the error data (if any exists).
8.0 Service Information The following sections describe basic service and maintenance procedures. 8.1 Troubleshooting Guide The following table lists some of the common problems and their suggested solutions. Problem Scale does not turn on Questionable weight or the scale does not zero Symptom Description Solution Dead battery Connect the scale to a power source. Faulty electrical outlet Use a different electrical outlet. Bad power supply Replace adaptor.
Problem Symptom Description Solution The display shows a STOP message The load on the scale exceeds the capacity of the scale Remove the excess weight and use the scale according to manufacture’s specs. The display shows LO Bat message The battery is low Recharge the battery. Err 2 Low saturation state (low A/D) The load cell is not connected properly. Check the cables and mechanical connections. If the problem persists, replace the set of load cells.
Table 8-2 lists the various display messages when testing the scale. NOTE: <-> means the display toggles between the two values. Step Function Available Parameters This enters into the test mode of the scale. StArt The scale automatically advances to Step 2. 2 Identifies the software ID IdEnt <-> 11007 Press the REWEIGH key to advance to the next step. 3 Shows the current weight value tESt <-> 0.0 To zero the display, press the Kg-Lb key. To advance to the next step, press the REWEIGH key.
8.3 Load Cell Replacement Replacement load cells can be ordered from Rice Lake Weighing Systems. To replace a load cell, lift the scale with a chain and remove foot, then remove the defective load cell. Disconnect load cell cable from the junction box and cut cable ties. When the cable is freed, pull cable out of the scale frame channels. Figure 8-2. Load Cell Assembly Follow the directions given below to install new load cells.
8.3.1 , o , NPEFMT Figure 8-4. Foot Pad - Side View Route the load cell cables near each corner so that the cable is free from possible contact with each foot. Hold the cable in position with the supplied adhesive-backed cable ties. Do not cut load cell cables. Coil extra cable before it enters the junction box, tie with cable ties, and insert the coils into the channel near the junction box.
9.0 Bariatric Wheelchair Scale Specifications Electrical Grounding For systems where the scale is connected to a 115 VAC circuit, the indicator must be directly connected to an earth ground with a ground interface cable of no more than 3W resistance throughout its length. Load Cell Excitation Rated Excitation: 10 VDC Maximum Excitation: 15 VDC Grade Level Requirements The supporting surface for the four feet of the scale must be level within 1/4 inch of horizontal.
Summit SB-1000 Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two years. RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized by RLWS.