Note
4 Versa-portion RLP-S/SM Operation Manual
To exit calibration at any time, press the TARE key.
3.Ensure there is no weight on the platter, then press the ZERO key to 
calibrate the zero point.
The display will show dashes, then 
CALSP.
4.Place the scale’s full capacity weight 
on the platter and press the ZERO key to 
calibrate span.
The display will show dashes, then the weight value.
Calibration weight is full grams capacity. Full capacity values are as 
follows: RLP-6S = 3000 g (3 kg); 
RLP-15S = 6000 g (6 kg); RLP-30S = 15000 g (15 kg); RLP-60S = 30000 g 
(30 kg).
5.The scale will return to weigh mode automatically. Power down the unit.
6.Calibration is complete.
3.3 Enclosure Security
After an NTEP inspector has examined the unit, he/she will install security cables 
through three bolts on the underside of the scale. These cables prevent the unit 
from being tampered with by an unauthorized individual. If the cable is removed, 
NTEP certification will become void.
Figure 3-2. Legal for trade seal; cable passes through three bolts
3.4 Setting the SPEC Value
The SPEC value is used to specify decimal point position, minimum display, net/
gross multi-interval, resolution, weight single or multi-interval, decimal point 
figure, negative weight display mask, IR mode protection (SPAN switch must be 
turned on to allow calibration), start range (amount of weight that can be on the 
scale when it is turned on), re-zero range, manual tare cancellation, tare 
subtraction, tare accumulation, auto tare, weight reset, tare auto clear, zero 
tracking, re-zero, and the 
UNIT key function. See Table 3-1 on page 5.
1.Press the setup switch (see Figure 3-1).
2.With the unit in weigh mode, pre
ss and hold the ZERO key.