Manual
Calibration 5
3.With the ZERO key held, press and release the TARE key, then press and 
release the POWER key twice.
The SPEC number is displayed.
4.To increment the SPEC setting value, press the POWER key. 
Refer to Table 3-1.
5.Press the ZERO key to save data and increase the SPEC count.
6.Press the ZERO key to advance to the next SPEC.
7.Press the TARE key to store SPECs and return to weigh mode.
SPEC 
Number
Bit 2 Bit 1 Bit 0
00 Decimal point position
000 - None
001 - 2nd digit (0.0)
010 - 3rd digit (0.00)
011 - 4th digit (0.000)
100 - 5th digit (0.0000)
101 ~ 111 - Not used
01 Minimum display
00 - 1
01 - 2
10 - 5
11 - 10
Net/gross multi-interval
0 - Gross
1 - Net
02 Resolution Selection
00 - 1/2000
01 - 1/3000
10 - 1/6000
11 - 1/7500
Weight single interval or 
multi-interval
0 - Single interval
03 Decimal point figure for 
numeral
0 - Period
1 - Comma
Negative weight display 
mask
0 - Minus gross > 9e
1 - Minus gross weight
IR (inter resolution) mode 
protected by SPAN SW 
(SPAN switch must be 
turned on for calibration)
0 - No
1 - Yes
04 Start range (amount of weight that can be on the 
scale when it is turned on; if there is more weight 
than allowed, the display will show all 8s)
00 - ±10% F.S.   10 - ±3%F.S.
01 - ±5% F.S.   11 - ±2%F.S.
Re-zero range
0 - ±2% F.S.
1 - ±3% F.S.
05
Not NTEP 
Approved
Manual tare cancellation
0 - Allow
1 - Prohibit
Tare  s ub t ra c ti o n
0 - Allow
1 - Prohibit
Tare accumulation
0 - Allow
1 - Prohibit
Table 3-1. SPEC Number And Bit Explanation (Defaults Are Bold) 