Manual
Calibration 3
3.0 Calibration
To enter calibration mode, remove the screw on the bottom of the scale as shown 
in Figure 3-1. Use a small tool to press the span setup switch located under the 
screw. Setup will time out to verify the switch is set to the 
ON position. Then, 
perform a span switch state check (see Section 3.1).
Remove Screw
Figure 3-1. Setup Switch
3.1 Software Version and Span State Check
1.With the unit in weigh mode, press and hold the ZERO key.
2.With the 
ZERO key held, press and release the POWER key three times.
3.The display will show the software version 
for two seconds, then show the 
current span switch state (S-ON or S-OFF).
After three seconds, the unit will return to weigh mode.
3.2 Performing a Calibration
The Versa-portion RLP-S/SM must be calibrated in grams or kilograms, and 
SPEC 01 and 02 must be set for the gram value (or kg in the SM version). See 
Section 3.4. 
1.With the unit in weigh mode, pre
ss and hold the ZERO key.
2.With the 
ZERO key held, press and release the POWER key twice, then press 
and release the TARE key.
The display will show CAL00.