Item Brochure

27
10
Guidanceandmanufacturer'sdeclarationelectromagneticemission
TheriscopeLinstrumentisintendedforuseintheelectromagneticenvironmentspecifiedbelow.The
customerortheuserofthee
scope®shouldensurethatitisusedinsuchanenvironment
Emissiontest Compliance Electromagneticenvironment–guidance
RFemissions
HF
emissionspursuantto
CISPR11
Group1
TheriscopeLusesRFenergy
exclusivelyforaninternalfunction.Therefore
hisRFtransmissionisverylowanditis
unlikelytobeadjacent
electronicdevicesaredisturbed.
RFemissions
HF
emissionspursuantto
CISPR11
ClassB
The
riscopeLisintendedforuseinall
establishments,includingresidentialareasand
thosedirectlyconnectedtoapublicsupplynetwork
thatalsosuppliesbuildingsusedforresidential
purposes.
Emissionsofharmonics
IEC61000
32
Notapplicable
Emissionsofvoltage
fluctuations,flicker

IEC61000
33
Notapplicable
11
Guidanceandmanufacture’sdeclaration–electromagneticimmunity
TheriscopeLinstrumentisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerorthe
userofthe
riscopeLshouldensurethatitisusedinsuchanenvironment
Immunitytesting
IEC60601testlevel
Compliance
Electromagneticenvironment
Instructions
Electrostaticdischarge
(ESD)
IEC61000
42
Con:±8kV
Air:±2,4,8,15kV
Con:±8kV
Air:±2,4,8,15kV
Floorsshouldbewood,concreteor
ceramictile.Iffloorarecoveredwith
syntheticmaterial,therelative
humidityshouldbeatleast30%.
Fasttransientelectrical
disturbances/bursts
IEC
61000
44
5/50ns,100kHz,±2kV
Notapplicable
Thequalityofthesupplyvoltage
shouldbethatofatypicalbus
iness
orhospitalenvironment
.
Surgevoltage
IEC61000
45
±0.5kVvoltage
Phase
tophaseconductor
±2kVvoltage
Line
toearth
Notapplicable
Thequalityofthesupplyvoltage
shouldbethatofatypicalbusiness
orhospitalenvironment
.
Voltagedips,shortterm
interruptionsand
fluctuationsinthesupply
voltageacc.to

IEC61000
411
<0%UT0.5periodat0.45,90,135,
180,225,270and315degrees
0%UT1periodand70%UT25/30
periodssingle
phase:at0degrees
(50/60Hz)
Notapplicable
Thequalityofthesupplyvoltage
shouldbethatofatypicalbusiness
orhospitalenvironment
.
Magneticfieldwith
efficiency
rated
frequencies
IEC61000
48
30A/m
50/60Hz
30A/m
50/60Hz
Mainsfrequencymagneticfields
shouldbeatalevelcharacteristicof
atypicallocationinatypical
commercialhospitalenvironment.
NOTEUTistheACsource.Mainsvoltagebeforetheapplicationofthetestlevel.