Item Brochure
27
10
Guidanceandmanufacturer'sdeclarationelectromagneticemission
TheriscopeLinstrumentisintendedforuseintheelectromagneticenvironmentspecifiedbelow.The
customerortheuserofthee
scope®shouldensurethatitisusedinsuchanenvironment
Emissiontest Compliance Electromagneticenvironment–guidance
RFemissions
HF
emissionspursuantto
CISPR11
Group1
TheriscopeLusesRFenergy
exclusivelyforaninternalfunction.Therefore
hisRFtransmissionisverylowanditis
unlikelytobeadjacent
electronicdevicesaredisturbed.
RFemissions
HF
emissionspursuantto
CISPR11
ClassB
The
riscopeLisintendedforuseinall
establishments,includingresidentialareasand
thosedirectlyconnectedtoapublicsupplynetwork
thatalsosuppliesbuildingsusedforresidential
purposes.
Emissionsofharmonics
IEC61000
32
Notapplicable
Emissionsofvoltage
fluctuations,flicker
IEC61000
33
Notapplicable
11
Guidanceandmanufacture’sdeclaration–electromagneticimmunity
TheriscopeLinstrumentisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerorthe
userofthe
riscopeLshouldensurethatitisusedinsuchanenvironment
Immunitytesting
IEC60601testlevel
Compliance
Electromagneticenvironment
Instructions
Electrostaticdischarge
(ESD)
IEC61000
42
Con:±8kV
Air:±2,4,8,15kV
Con:±8kV
Air:±2,4,8,15kV
Floorsshouldbewood,concreteor
ceramictile.Iffloorarecoveredwith
syntheticmaterial,therelative
humidityshouldbeatleast30%.
Fasttransientelectrical
disturbances/bursts
IEC
61000
44
5/50ns,100kHz,±2kV
Notapplicable
Thequalityofthesupplyvoltage
shouldbethatofatypicalbus
iness
orhospitalenvironment
.
Surgevoltage
IEC61000
45
±0.5kVvoltage
Phase
tophaseconductor
±2kVvoltage
Line
toearth
Notapplicable
Thequalityofthesupplyvoltage
shouldbethatofatypicalbusiness
orhospitalenvironment
.
Voltagedips,shortterm
interruptionsand
fluctuationsinthesupply
voltageacc.to
IEC61000
411
<0%UT0.5periodat0.45,90,135,
180,225,270and315degrees
0%UT1periodand70%UT25/30
periodssingle
phase:at0degrees
(50/60Hz)
Notapplicable
Thequalityofthesupplyvoltage
shouldbethatofatypicalbusiness
orhospitalenvironment
.
Magneticfieldwith
efficiency
rated
frequencies
IEC61000
48
30A/m
50/60Hz
30A/m
50/60Hz
Mainsfrequencymagneticfields
shouldbeatalevelcharacteristicof
atypicallocationinatypical
commercialhospitalenvironment.
NOTEUTistheACsource.Mainsvoltagebeforetheapplicationofthetestlevel.