Item Brochure

05
VITAL SIGNS MONITORS
RVS-100/200
RIESTER VITAL SIGNS MONITORS
Technical Data
Monitor
Power Supply: 100 – 240 V AC, 50/60 Hz
Battery: Lithium-Ion battery, rechargeable, 10.8 V DC, 6,600 mAh, Charging time to a
100%: 6 hours, Operating time approx. 11 hours (new and fully charged battery at
25°C ambient temperature with connected SpO2, Temp, and NIBP on AUTO mode
(15 minutes interval))
Low battery indicator
Charging battery indicator
Dimensions Monitor: L x H x W: 303 x 230 x 110 mm (11.93 x 9.05 x 4.33 in)
Weight: 3600 g, (7.94 lb) including temp module, printer,
Dimensions Basket: L x H x W: 270 x 225 x 200 mm (10.63 x 8.86 x 7.87 in)
Alarm: Visible LEDs (Yellow, Red, Blue, Green, Orange)
Audible Speaker
Gives audible alarm, QRS tone
Supports Pitch Tone and multi-level volume
Alarm tones meet the requirement of IEC 60601-1-8.
Alarm Pressure: 45 dB to 85 dB test distance is 1 meter from the tone.
Ports: 1 x AC power inlet
1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
4 x USB ports
1 x Equipotential grounding point
1 x RJ11 connector for nurse call
DC out port 15V/1.2 A
Clinical-grade NIBP (non-invasive blood pressure)
Measuring method: automatic oscillometric
Measurement range: Adult
SYS 30 – 270 mmHg, DIA 10 – 220 mmHg, MAP 20 – 235 mmHg
Pediatric
SYS 30 – 235 mmHg, DIA 10 – 220 mmHg, MAP 20 – 225 mmHg
Neonate
SYS 30-135 mmHg, DIA 10-110 mmHg, MAP 20-125 mmHg
Cuff pressure range: 0 – 280 mmHg
Resolution: 1 mmHg
Pressure accuracy: Static ±3 mmHg
Clinic ±5 mmHg, standard deviation ) 8 mmHg
Pulse rate range: 40 – 240 bpm
Pulse rate accuracy: 1 bpm
Initial inflation pressure: Adult 160 mmHg, Pediatric 130 mmHg, Neonate 75 mmHg
Double overpressure protection (hardware and software)
Adult 297 ± 3 mmHg, Pediatric 252 ± 3 mmHg , Neonate 147 ± 3 mmHg
Alarm range:
SYS 0 – 300 mmHg
DIA 0 – 300 mmHg
MAP 0 – 300 mmHg