Item Brochure

06
RVS-100/200
RIESTER VITAL SIGNS MONITORS
Technical Data
Riester SpO2 Measurement range: 0 – 100%
Resolution: 1%
Accuracy: ±2%(70 to 100%)
±3% (40% to 69%)
Unspecified (0% to 39%)
Alarm range: 0 – 100%
Average time: 4 sec, 8 sec, 16 sec
Pulse rate
Measurement range: 20 – 250 bpm
Resolution: 1 bpm
Accuracy: ±1% or ±1 bpm whichever is greater
Nellcor SpO2 Measurement range: 0-100%
Resolution: 1%
Accuracy: 70 – 100%: ±2% (adult, pediatric)
70 100%: ±3% (neonate)
0 69%: unspecified
Alarm range: 0 – 100%
Average time: 8 sec, 16 sec
Pulse rate
Measurement range: 20 – 300 bpm
Resolution: 1 bpm
Accuracy: 20 – 250 bpm: ±3 bpm
251 300 bpm: unspecified
Masimo SpO2 Measurement range: 0 – 100%
Resolution: 1%
Accuracy: 70 – 100%: ±2% (adult, pediatric, non-motion conditions)
70 – 100%: ±3% (neonate, non-motion conditions)
70 100%: ±3% (motion conditions)
0 69%: unspecified
Average time: 2 – 4 sec, 4 – 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
Pulse rate
Measurement range: 25 – 240 bpm
Resolution: 1 bpm
Accuracy: ±3 bpm (non-motion conditions)
±5 bpm (motion conditions)
Predictive thermometer
Oral ( Quick Mode ) non-fever temps: 3-5 seconds
fever temps: 8-10 seconds
Oral ( Standard Mode ): 6-10 seconds
Axillary Mode: 8-12 seconds
Rectal Mode: 10-14 seconds
Direct Mode (all sites): 60-120 seconds
Measurement accuracy: 35.5ºC - 42ºC
Direct Mode (all sites): ±0.1ºC (±0.2ºF)
Standard Prediction Mode (all sites): ±0.1ºC (±0.2ºF)
Quick Prediction Mode (oral only): ± 0.3ºC (±0.5º)