User`s guide

Chapter 2 Front Panel Operations RIGOL
DG1000Z Users Guide 2-39
Frequency Modulation (FM)
For frequency modulation (FM), the frequency of the carrier waveform varies with
the instantaneous voltage of the modulating waveform.
To Select FM Modulation
Press Mod Type FM to enable FM function. When Mod is enabled, Sweep or
Burst will be disabled automatically (if enabled currently).
To Select Carrier Waveform Shape
FM carrier waveform could be Sine, Square, Ramp or Arbitrary waveform (except
DC). The default is Sine.
Press the front panel key Sine, Square, Ramp or Arb (you need to press
Select Wform in the arbitrary waveform setting interface to select the desired
arbitrary waveform) to select the desired carrier waveform shape.
Pulse, Noise and DC could not be used as carrier waveform.
To Set Carrier Waveform Parameters
The different settings of various parameters (frequency, amplitude, DC offset, start
phase and etc.) of the carrier waveform will influence the FM modulated waveform.
For different carrier waveform shape, the ranges of the various parameters are
different (the ranges are related to the model of the instrument used and the carrier
waveform currently selected. See Specificationsfor details). For all carrier
waveform shapes, the default values are 1kHz frequency, 5Vpp amplitude, 0V
DC
offset and 0° start phase.
When the carrier waveform currently selected is Sine, Square or Ramp, please
refer to the introduction in To Output Basic Waveform to set the
parameters of the carrier waveform currently selected.
When the carrier waveform currently selected is arbitrary waveform, please
refer to the introduction in To Output Arbitrary Waveform to set the
parameters of the carrier waveform currently selected.