Chapter 2 Front Panel Operations RIGOL
DG2000 User Guide 2-79
Seamless Interconnection with Oscilloscope
DG2000 can also seamlessly interconnect with RIGOL's oscilloscopes that support
USB-TMC; rebuild and output the waveform data collected by the oscilloscope
losslessly. The connection method between the instruments is as shown in the figure
below.
Operation Procedures:
1. Make sure the signal under test is displayed stably on the screen of DS1.
2. Connect the USB HOST interface of DG2000 with the USB DEVICE interface of
DS1 by using the USB cable. Now, DS1 will enter the remote mode
automatically.
3. Use DG2000 to read the signal under test that is displayed currently on the
screen of DS1: When the Shift key is illuminated, press Noise/Store on the
front panel of DG2000 to enter the store and recall interface. At this time, the
Oscilloscope menu label is displayed below the Categories menu label, and
the model identifier of DS1 will be displayed under Oscilloscope. Tap or use
the knob of DG2000 to select the DS1 model identifier to enter the seamless
interconnection operation interface.
Figure 2-22 Seamless Interconnection Operation Interface
All the channels of DS1 and their on/off status are displayed at the left section of
the interface. The parameter settings are displayed at the right section of the
interface. Tap the first parameter selection field to set the data type of the
current channel to be read to "Screen Data" or "Ram Data". Tap the second
parameter selection field to set the data position (only available when data type
DG2000
DS1
DS2
USB DEVICE USB HOST Output Terminal Input Terminal
USB Cable BNC Cable