Chapter 1 Test Overview
Test Preparations
Before performing the test, make sure that the instrument is within the calibration
period (the recommended calibration period is 1 year) and has been warmed up for
at least 30 minutes under the specified operation temperature (18℃ to 28℃).
Recommended Test Devices
It is recommended that you use the test devices listed in the table below or other test
devices whose performance specifications satisfy the "Performance Requirement"
listed in the table below to test the performance specifications of the DG2000 series.
Table 1-1 Recommended test devices
Device Performance Requirement
Recommended
Instrument
Frequency Counter
>10MHz
Accuracy: 0.1ppm
Keysight 53131A
Digital Multimeter
6
1
/
2
digits
RIGOL DM3068
Power Meter
-30dBm to +20dBm
Accuracy: ±0.02dB
Resolution: 0.01dB
Keysight N1913A
Power Sensor
-30dBm to +20dBm
Keysight N8482A
Spectrum Analyzer
Minimum resolution bandwidth is
100Hz
RIGOL DSA815
Oscilloscope
Bandwidth: 500MHz
Rise/Fall time measurement
function
Overshoot measurement
function
RIGOL DS4052/DS4054
Power Sensor
Connecting Cable
Used to connect the power meter
and power sensor
--
Connecting Cable
BNC (m)-BNC (m)
--
Connecting Cable
BNC (m)-Dual banana plug (m)
--
50Ω Load
50Ω/1W
--
Adaptor
N (f)-BNC (m)
--
Adaptor
BNC (f)-N (m)
--