Instructions

Chapter 2 Performance Verification Test RIGOL
DG2000 Performance Verification Guide
2-3
readings of the frequency counter respectively and judge whether the readings
are between 0.999 998 999 999 999 99MHz and 1.000 001 000 000 000 01MHz.
7. Repeat steps 1 to 6 to test the frequency accuracy of CH2 and record the test
results.
Test Record Form
Waveform
Setting
Value
Measurement
Value
Specification Pass/Fail
Sine
Frequency:
1MHz
Amplitude:
1Vpp
0.999 998 999 999 999
99MHz to 1.000 001 000 000
000 01MHz
Square
Ramp
Pulse