Chapter 2 Performance Verification Test RIGOL
DG2000 Performance Verification Guide
2-5
c) Press Output1 or tap the channel output configuration status bar
to turn on the output of CH1 in the channel setting interface.
5. Record the reading of the multimeter and judge whether it is within the
specification ("Amplitude Output Value (Vrms)" in Table 2-1) range.
6. Keep the output impedance of CH1 of DG2000 at 50Ω and the output waveform
of CH1 as a sine waveform with 1kHz frequency and 0V
dc
offset. Set the output
amplitude of CH1 to 100mVpp, 500mVpp, 1Vpp, 5Vpp and 10Vpp respectively.
Record the readings of the multimeter respectively and judge whether the
readings are within the specification (“Amplitude Output Value (Vrms)” in Table
2-1) range.
Table 2-1 Amplitude output values (Vrms) of AC amplitude accuracy test
Amplitude
Setting
Value
(Vpp)
Allowed
Error (Vpp)
[1]
Amplitude Output
Value (Vpp)
Amplitude Output Value
(Vrms)
[2]
20mVpp ±5.2mVpp 14.8mVpp to 25.2mVpp 5.2mVrms to 8.9mVrms
100mVpp ±6mVpp 94mVpp to 106mVpp 33.2mVrms to 37.5mVrms
500mVpp ±10mVpp 490mVpp to 510mVpp 173.3mVrms to 180.3mVrms
1Vpp ±15mVpp 0.985Vpp to 1.015Vpp 348.3mVrms to 358.9mVrms
5Vpp ±55mVpp 4.945Vpp to 5.055Vpp 1.75Vrms to 1.7875Vrms
10Vpp ±105mVpp 9.895Vpp to 10.105Vpp 3.5Vrms to 3.5732Vrms
Note
[1]
: "Allowed Error" is calculated from the specification "±(1% of setting value) ±5mVpp".
Note
[2]
: "Amplitude Output Value (Vrms)" is calculated from "Amplitude Output Value (Vpp)".
The conversion relation between Vrms and Vpp is
rms22pp VV =
.
7. Repeat Step 1-6 to test the AC amplitude accuracy of CH2 and record the test
results.
Test Record Form
Amplitude
Setting
Value
Setting
Measurement
Value
Specification Pass/Fail
20mVpp
Frequency:
1kHz
Offset: 0V
dc
Impedance:
50Ω
5.2mVrms to 8.9mVrms
100mVpp 33.2mVrms to 37.5mVrms
500mVpp 173.3mVrms to 180.3mVrms
1Vpp 348.3mVrms to 358.9mVrms
5Vpp 1.75Vrms to 1.7875Vrms
10Vpp 3.5Vrms to 3.5732Vrms