Instructions

RIGOL Chapter 2 Performance Verification Test
DG2000 Performance Verification Guide
2-14
results.
Note
[2]
: 2
nd
order harmonic distortion = 2
nd
order harmonic measurement value base waveform
measurement value
For example, when the output waveform frequency of the channel is 10MHz, if the base
waveform measurement value is 0.8dBm and the 2
nd
order harmonic measurement value
is -56.2dBm, the 2
nd
order harmonic distortion = (-56.2) -0.8=-57dBc<-55dBc and the
test result fulfills the specification requirement.
Test Record Form
Frequency
Setting
Value
Setting
Measurement
Value
Calculation
Result
[3]
Specification Pass/Fail
10MHz
Waveform:
Sine
Amplitude:
0dBm
Offset: 0V
dc
Base waveform:
<-55dBc
2
nd
order
harmonic:
20MHz
Base waveform:
<-50dBc
2
nd
order
harmonic:
40MHz
Base waveform:
<-40dBc
2
nd
order
harmonic:
60MHz
Base waveform:
<-35dBc
2
nd
order
harmonic:
Note
[3]
: Calculation result = 2th order harmonic measurement value - base waveform measurement value.