Instructions
Appendix RIGOL
DG2000 Performance Verification Guide
5
Channel: CH2
Frequency Accuracy Test
Waveform
Setting
Value
Measurement
Value
Specification Pass/Fail
Sine
Frequency:
1MHz
Amplitude:
1Vpp
0.999 998 999 999 999
99MHz to 1.000 001 000 000
000 01MHz
Square
Ramp
Pulse
AC Amplitude Accuracy Test
Amplitude
Setting
Value
Setting
Measurement
Value
Specification Pass/Fail
20mVpp
Frequency:
1kHz
Offset: 0V
dc
Impedance:
50Ω
5.2mVrms to 8.9mVrms
100mVpp 33.2mVrms to 37.5mVrms
500mVpp 173.3mVrms to 180.3mVrms
1Vpp 348.3mVrms to 358.9mVrms
5Vpp 1.75Vrms to 1.7875Vrms
10Vpp 3.5Vrms to 3.5732Vrms
DC Offset Accuracy Test
Offset
Setting
Value
Setting
Measurement
Value
Specification Pass/Fail
-2.5V
dc
Frequency:
1kHz
Amplitude:
5Vpp
Impedance:
50Ω
-2.580V
dc
to -2.420V
dc
-1V
dc
-1.065V
dc
to -0.935V
dc
-500mV
dc
-0.560V
dc
to -0.440V
dc
0V
dc
-0.055V
dc
to 0.055V
dc
500mV
dc
0.440V
dc
to 0.560V
dc
1V
dc
0.935V
dc
to 1.065V
dc
2.5V
dc
2.420V
dc
to 2.580V
dc