Instructions
RIGOL Appendix
DG2000 Performance Verification Guide
14
Type
4.3-inch TFT LCD touch screen
Resolution
480 horizontal × RGB × 272 vertical resolution
Color
16 M
Environment
Temperature Range
Operating: 0℃ to 45℃
Non-operating: -40℃ to 60℃
Cooling Method
Natural air cooling
Humidity Range
Below 30℃: ≤95%RH
30℃ to 40℃: ≤75%RH
40℃ to 50℃: ≤45%RH
Altitude
Operating: below 3,000 meters
Non-operating: below 15,000 meters
Mechanical Characteristics
Dimensions
(W×H×D)
261.5 mm × 112 mm × 318.4 mm
Weight
Package excluded: 3.2 kg
Package included: 4.5 kg
Interface
USB Host, USB Device, and USB-GPIB
IP Protection
IP2X
Calibration
Interval
1 year (recommended)
Certification Information
EMC
Compliant with EN
61326-1:2006
IEC 61000
-3-2:2000
±4.0 kV (Contact Discharge)
±4.0 kV (Air Discharge)
IEC 61000
-4-3:2002
3 V/m (80 MHz to 1 GHz); 3 V/m (1.4 GHz
to 2 GHz); 1 V/m (2.0 GHz to 2.7 GHz)
IEC 61000
-4-4:2004 1kV power line
IEC 61000
-4-5:2001
0.5 kV (phase-to-neutral voltage);
0.5 kV (phase-to-earth voltage);
1 kV (neutral-to-earth voltage)
IEC 61000
-4-6:2003 3 V, 0.15 MHz to 80 MHz
IEC 61000
-4-11:2004
Voltage dip:
0% UT during half cycle
0% UT during 1 cycle
70% UT during 25 cycles
Short interruption:
0% UT during 1 cycle
Electrical Safety
complies with
USA: UL 61010
-1:2012,
Canada:
CAN/CSA-C22.2
No. 61010
-1-2012
EN 61010-1:2010,
Note
[1]
: 0 dBm output, DC offset 0, impedance 50 Ω.