Instructions

Chapter 1 Quick Guide RIGOL
DG2000 User Guide 1-7
Sine
Preset
Outputs Sine waveform.
When you select the function, you can set Freq/Period,
Ampl/HiLevel, Offset/LoLevel and Start Phase of sine
waveform.
Restores the instrument to its preset state. At most 10
states can be preset.
Square
Lock
Outputs Square waveform.
When you select the function, you can set Freq/Period,
Ampl/HiLevel, Offset/LoLevel, Duty Cycle and Start
Phase of square waveform.
Locks or unlocks the front-panel keys and the touch
screen. In the unlocked state, press this key to lock the
front-panel keys and the touch screen. At this time,
except the Square/Lock key, all other keys on the
front panel and the touch screen operation are invalid.
Press the Square/Lock key again to unlock the keys
and the touch screen.
Ramp
Trig
Outputs Ramp waveform.
When you select the function, you can set Freq/Period,
Ampl/HiLevel, Offset/LoLevel, Symmetry and Start
Phase of ramp waveform.
Used for manual trigger.
The default trigger setting for the generator is
Internal trigger. In this mode, when you select the
sweep or burst mode, the generator outputs the
waveforms continuously. At this time, press the
Ramp/Trig key, and instrument automatically
switches to the manual trigger mode from auto
trigger.
Each time you press the Ramp/Trig key, one
sweep will be triggered manually or one burst will
be output.
Pulse
Utility
Outputs Pulse waveform.
When you select the function, you can set Freq/Period,
Ampl/HiLevel, Offset/LoLevel, Width/Duty, Leading,
Trailing and Start Phase of pulse waveform.
It is used to set the utility function parameters and
system parameters.
Noise
Outputs the Gaussian noise.
When you select the function, you can set Ampl/HiLevel
and Offset/LoLevel of Noise.