Instructions

Chapter 2 Performance Verification Test RIGOL
DG2000 Performance Verification Guide
2-19
b) Set the horizontal time base to 1us.
c) Adjust the trigger level to a proper value.
d) Set the input impedance to 50Ω.
e) Turn on the rise time and fall time measurement functions.
5. Set the edge type of the oscilloscope to rising edge, record the measurement
result of the rise time and judge whether it is within the specification range.
6. Set the edge type of the oscilloscope to falling edge, record the measurement
result of the fall time and judge whether it is within the specification range.
7. Repeat Step 1-6 to test the rise/fall time of CH2 and record the measurement
results.
Test Record Form
Waveform Setting
Measurement
Value
Specification Pass/Fail
Square
Frequency: 1kHz
Amplitude: 1Vpp
Offset: 0V
dc
Rise Time
Typical (1 Vpp, 1
kHz)
9 ns
Fall Time