1. Self-Test
Press Utility Test/Cal Self-Test , to perform self-testing.
When the instrument is powered, the system will perform self-testing
automatically.This limited test can test if the instrument could work ordinarily.
To perform general test, 15 second is need. If all the tests have been passed, you
can use the generator freely.
If general test has been passed, the screen will show “Self-Test Passed”; if failed,
it will tell you self-test failed and display err message.
2. Cal Info
Info mainly includes the calibration times and the edition.
Please note that the generator has been calibrated before it is shipped. When
you turn on a new instrument, the default number of calibration is 0.
The edition version is the updating information of the edition; the latest version
can be reached at the company’s website: www.rigol.com.
3. Secure Code
This function is used to prevent unauthorized or accidental calibration to the
generator. When the generator is firstly shipped to the user, it is secured. You
have to unlock it before calibrating the instrument.
4. Perform the Calibration
Press Utility Test/Cal Secure Code , and input the right password. The
menu in figure 2-101 will show. Choose Perform Cal , and enter the menu
shown in table 2-51.