User Manual
Table Of Contents
- Safety Notices
 - The Introduction of DG3000 Series
 - Getting Started
 - Operating Your Generator
 - Application & Examples
- Example 1: To Generate a Sine Wave
 - Example 2: To Generate a Square Wave
 - Example 3: To Generate a Ramp Wave
 - Example 4: To Generate a Pulse Wave
 - Example 5: To Generate a Noise Wave
 - Example 6: To Generate an Arbitrary Waveform
 - Example 7: To Create an Arbitrary Waveform
 - Example 8: To Generate an AM Waveform
 - Example 9: To Generate an FSK Waveform
 - Example 10: To Generate a PWM waveform
 - Example 11: To Generate Linear Sweep
 - Example 12: To Generate a Burst Waveform
 
 - Prompt messages & troubleshooting
 - Specifications
 - Appendix
 - Index
 
RIGOL 
 © 2006 RIGOL Technologies, Inc. 
  User’s Guide for DG3000 Series 
4-6 
Conflict Setting 
Turned off infinite burst. 
Infinite burst is allowed only when the source is external or manual. The upper limit 
of the burst cycle is 1,000,000.   
Trigger source changes to Ext/Manual. 
Only when the External or Manual Trigger Source is selected, infinite burst is allowed.   
Burst period increased to fit entire burst. 
The PRI of Period is higher than that of the Burst Period. As long as the burst period 
does not reach its maximum, the generator will increase the burst period to cater to 
the need of the specific burst cycle or waveform frequency.   
Burst count reduced. 
If the burst period has reached its maximum, burst count will decrease to cater to 
the need of the specified waveform frequency.   
Trigger delay reduced to fit entire burst. 
Trigger delay will be reduced to fit the current burst period and burst count. Trigger 
delay is  the  time between the  time when  trigger is  received and  the start  of  the 
burst.   
Sorry, unable to N-cycle burst with noise. 
Noise can not be used in the trigger burst mode; it can only be used in the gated 
burst. 
Triggered burst not available for noise. 
Noise can not be used in the trigger burst mode; it can only be used in the gated 
burst. 
Trigger ignored -OUTPUT is off. 
When the OUTPUT is disabled, the Trigger Output will be disabled automatically. 
Trigger output connector used by FSK. 
If the FSK is activated and the Source is external, no “Trigger Output” is allowed.   
Trigger output connector used by burst gate. 
If the gated burst mode is activated and burst is also activated, no “Trigger Output” 
is allowed.   
Frequency is set maximum for Arb. 
The maximum frequency for arbitrary waveform is  25MHz. when the waveform is 
turned from one that allows higher frequency to the arbitrary waveform, the system 
will adjust the frequency to 25MHz automatically. 
Frequency is set maximum for Pulse. 
The maximum frequency for pulse is 30MHz. when the waveform is turned from one 
that  allows  higher  frequency  to  the  pulse  waveform,  the  system  will  adjust  the 