Users Guide for DG3000 Series
4-11
The Square Wave Period upper limit is 1Ms. If the specified period is greater than
1Ms, the system will adjust the period to 1Ms automatically.
Square Wave period lower limit= 8.33ns
The Square Wave Period lower limit is 8.33ns. If the specified period is less than
8.33ns, the system will adjust the period to 8.33ns automatically.
Ramp Wave period upper limit= 1Ms
The Ramp Wave Period upper limit is 1Ms. If the specified period is greater than 1Ms,
the system will adjust the period to 1Ms automatically.
Ramp Wave period lower limit= 1μs
The Ramp Wave Period lower limit is 1μs. If the specified period is less than 1μs, the
system will adjust the period to 1μs automatically.
Pulse Wave period upper limit= 2ks
The Pulse Wave Period upper limit is 2Ks. If the specified period is greater than 2Ks,
the system will adjust the period to 2Ks automatically.
Pulse Wave period lower limit= 28ns
The Pulse Wave Period lower limit is 16ns. If the specified period is less than 16ns,
the system will adjust the period to 16ns automatically.
Arb Wave period upper limit= 1Ms
The Arb Wave Period upper limit is 1Ms. If the specified period is greater than 1Ms,
the system will adjust the period to 1Ms automatically.
Arb Wave period lower limit= 40ns
The Arb Wave Period lower limit is 40 ns. If the specified period is less than 40 μs,
the system will adjust the period to 40 ns automatically.
Amplitude upper limit= **
Different mode has different amplitude upper limit; amplitude upper limit may vary
in the same mode if the load is different.
Amplitude lower limit= **
Different mode has different amplitude lower limit; amplitude lower limit may vary in
the same mode if the load is different.
High level upper limit = **
Different mode has different High Level upper limit; High Level upper limit may vary
in the same mode if the load is different.
Low level lower negative limit=**
Different mode has different High Level lower limit; High Level lower limit may vary
in the same mode if the load is different.
High level must be greater than Low level
High level must always be greater than Low level. If the high level is set less than the
low level, the generator will set the low level 1 mV less than the high level
automatically.