Users Guide for DG3000 Series
4-14
the Start Frequency upper limit is 25MHz.
Start Frequency lower limit = 1μHz
The Start Frequency lower limit is 1μHz. If the specified Start Frequency is less than
1μHz, the system will adjust the Start Frequency to 1μHz automatically.
Stop Frequency upper limit = **
Different Source types have different Stop Frequency upper limit. If the Carrier is a
Square or Sine wave, the Stop Frequency upper limit is 120MHz; if the Carrier is a
Ramp wave, the Stop Frequency upper limit is 1MHz; if the Carrier is an Arb wave,
the Stop Frequency upper limit is 25MHz.
Stop Frequency lower limit =1μHz
The Stop Frequency lower limit is 1μHz. If the specified Stop Frequency is less than
1μHz, the system will adjust the Stop Frequency to 1μHz automatically.
Center Frequency upper limit = **
Different Source types have different Center Frequency upper limit. If the Carrier is a
Square or Sine wave, the Center Frequency upper limit is 120MHz; If the Carrier is a
Ramp wave, the Center Frequency upper limit is 1MHz; If the Carrier is an Arb wave,
the Center Frequency upper limit is 25MHz.
Center Frequency lower limit = 1μHz
The Center Frequency lower limit is 1μHz. If the specified Center Frequency is less
than 1μHz, the system will adjust the Center Frequency to 1μHz automatically.
Frequency span upper limit = **
With different Source type modes and center frequencies, the frequency span upper
limits vary.
Frequency span lower limit = **
With different Source type modes and center frequencies, the frequency span lower
limits vary.
Sweep Time upper limit = 500.00s
The Sweep Time upper limit is 500.00s. If the specified Sweep Time is greater than
500.00s, the system will adjust the Sweep Time to 500.00s automatically.
Sweep Time lower limit = 1.0ms
The Sweep Time lower limit is 1.0ms. If the specified Sweep Time is less than 1.0ms,
the system will adjust the Sweep Time to 1.0ms automatically.
Start Phase upper limit = 360.0
o
The Start Phase upper limit is 360.0
o
. If the specified Start Phase is greater than
360.0
o
, the system will adjust the Start Phase to 360.0
o
automatically.
Start Phase lower limit = -360.0
o
The Start Phase lower limit is -360.0
o
. If the specified Start Phase is less than -360.0
o
,
the system will adjust the Start Phase to -360.0
o
automatically.
# of Cycles upper limit = 1,000,000 Cycle