User Manual

RIGOL
© 2006 RIGOL Technologies, Inc.
Users Guide for DG3000 Series
4-15
The Cycle upper limit is 500,000Cyc. If the specified Cycle is greater than
500,000Cyc, the system will adjust the Cycle to 500,000Cyc automatically.
# of Cycles lower limit = 1 Cycle
The Cycle lower limit is 1Cyc. If the specified Cycle is less than 1Cyc, the system will
adjust the Cycle to 1Cyc automatically.
Delay upper limit = 85.0s
The Delay upper limit is 85.0s. If the specified Delay is greater than 85.0s, the
system will adjust the Delay to 85.0s automatically.
Delay lower limit = 0.0000s
The Delay lower limit is 0.0000s. If the specified Delay is less than 0.0000s, the
system will adjust the Delay to 0.0000s automatically.
Phase deviation upper limit=360
o
The Phase deviation upper limit is 360
o
. If the specified Phase deviation is greater
than 360
o
, the system will adjust the Phase deviation to 360
o
automatically.
Phase deviation lower limit= 0
o
The Phase deviation lower limit is 0
o
. If the specified Phase deviation is less than 0
o
,
the system will adjust the Phase deviation to 0
o
automatically.
PM Frequency upper limit= 20 kHz
The PM Frequency upper limit is 20 kHz. If the specified PM Frequency is greater
than 20 kHz, the system will adjust the PM Frequency to 20 kHz automatically.
PM Frequency lower limit= 2mHz
The PM Frequency lower limit is 2mHz. If the specified PM Frequency is less than
2mHz, the system will adjust the PM Frequency to 2mHz automatically.
Phase upper limit= +360.0º
The Phase upper limit is +360.0
o
. If the specified Phase is greater than 360.0
o
, the
system will adjust the Phase to 360.0
o
automatically.
Phase lower negative limit
The Phase lower negative limit is 0
o
. If the specified Phase is less than 0
o
, the system
will adjust the Phase to 0
o
automatically.
Trigger Period upper limit = 500.00s
The Trigger Period upper limit is 500.00s. If the specified Trigger Period is greater
than 500.00s, the system will adjust the Trigger Period to 500.00s automatically.
Trigger Period lower limit = 1.0μs
The Trigger Period upper limit is 1.0μs. If the specified Trigger Period is less than
1.0μs, the system will adjust the Trigger Period to 1.0μs automatically.
Phase Deviation Upper Limit= 360
o
Phase Deviation Upper Limit is 360
o
. If the specified Phase Deviation is greater than
360
o
, the system will adjust the Phase Deviation to 360
o
automatically.
Phase Deviation Lower Limit= 0
o
.