To Set the Edge Time
1. Press Pulse Edge Time , to set the Edge Time.
The Edge Time shown on the screen when the instrument is powered is the
default value or the set value beforehand. When changing the function, if the
current value is valid for the new waveform, it will be used sequentially.
2. Input the desired Edge Time
Use the keypad or the knob to input the desired value, choose the unit, and press
the corresponding button. The Generator will change the waveform immediately.
Figure 2-15 Setting the Edge Time
Instruction:
The default setting is the same Edge Time for Rising and Falling Edges.
Current
Parameter:
Pulse Edge