1. To Set the Output Load
For the [Output] Connector on the Front panel, the Generator has a built-in 50Ω
series impendence. If the actual load does not match the set one, the displayed
amplitude and offset are incorrect. This function is used to match the displayed
voltage with the expected one.
Steps for setting the Load:
(1) Press Utility Output Setup Load , to enter the following interface.
Please note that the Load Parameter shown on the right bottom is the
default setting when the power is on or the pre-set load value. If the current
value is valid for the output, then is current value will be used.
(2) Input the desired Load Value.
Use the Keypad or the Knob to enter the desired value and choose the unit,
Ω or KΩ, press the corresponding button.
Figure 2-77 Set the Output Load
Instruction
DG3000 Series has a fixed 50Ω Series Impendence. No matter what Value the set
parameter is, if the real load is different from the set one, the displayed voltage
will not equal the real voltage.