RIGOL User’s Guide DG800 Series Function/Arbitrary Waveform Generator Jul. 2018 RIGOL (SUZHOU) TECHNOLOGIES INC.
RIGOL Guaranty and Declaration Copyright © 2018 RIGOL (SUZHOU) TECHNOLOGIES INC. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL (SUZHOU) TECHNOLOGIES INC. Publication Number UGB11100-1110 Software Version 00.00.01 Software upgrade might change or add product features. Please acquire the latest version of the manual from RIGOL website or contact RIGOL to upgrade the software. Notices RIGOL products are covered by P.R.C.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly. Use the BNC output connector properly. The BNC output connector on the front panel only allows to output the signal but not to input the signal.
RIGOL Use Proper Fuse. Please use the specified fuses. Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered on. Do Not Operate With Suspected Failures. If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel. Provide Adequate Ventilation.
RIGOL Safety Notices and Symbols Safety Notices in this Manual: WARNING Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death. CAUTION Indicates a potentially hazardous situation or practice which, if not avoided, could result in damage to the product or loss of important data. Safety Terms on the Product: DANGER WARNING CAUTION It calls attention to an operation, if not correctly performed, could result in injury or hazard immediately.
RIGOL Care and Cleaning Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning Clean the instrument regularly according to its operating conditions. 1. Disconnect the instrument from all power sources. 2. Clean the external surfaces of the instrument with a soft cloth dampened with mild detergent or water. Avoid having any water or other objects into the chassis via the heat dissipation hole.
RIGOL DG800 Series Function/Arbitrary Waveform Generator Overview As a multi-functional signal generator, DG800 series function/arbitrary waveform generator integrates many instruments into 1, such as function generator, arbitrary waveform generator, noise generator, pulse generator, harmonic generator, analog/digital modulator, and frequency counter.
RIGOL Document Overview Main Topics in this Manual Chapter 1 Quick Start Chapter 2 Front Panel Operations Chapter 3 Remote Control Chapter 4 Troubleshooting Chapter 5 Appendix Introduces the appearance and dimensions of DG800, its front/rear panel, and user interface. Introduces the main functions and operation methods of DG800. Introduces how to control DG800 remotely. Lists the possible failures and solutions in using DG800.
RIGOL 2. DG800 series function/arbitrary waveform generator is equipped with two channels (CH1 and CH2). Unless otherwise specified, this manual takes CH1 as an example to introduce the operation methods of the generator. The operation methods of CH2 is the same as that of CH1. Manuals of this Product The manuals of this product mainly include the quick guide, user’s guide, programming guide, data sheet, and etc. For the latest version of this manual, download it from the official website of RIGOL (www.
Contents RIGOL Contents Guaranty and Declaration .........................................................................I Safety Requirement ................................................................................ II General Safety Summary ........................................................................... II Safety Notices and Symbols ...................................................................... IV Care and Cleaning ..............................................................
RIGOL Contents To Set Fundamental Waveform Parameters ...................................... 2-24 To Select Harmonic Type ................................................................ 2-24 To Set Harmonic Order ................................................................... 2-24 To Select Harmonic Amplitude......................................................... 2-25 To Set Harmonic Phase .................................................................. 2-25 Example: To Output Harmonic ..........
RIGOL Contents File Operation ................................................................................2-75 Seamless Interconnection with Oscilloscope ..................................... 2-79 Channel Setting....................................................................................2-81 Output Setting ...............................................................................2-81 Sync Setting ..................................................................................
Chapter 1 Quick Guide RIGOL Chapter 1 Quick Start This chapter briefly introduces the appearance and dimensions of DG800 series function/arbitrary waveform generator, its front/rear panel, and user interface.
RIGOL Chapter 1 Quick Guide General Inspection 1. Inspect the packaging If the packaging has been damaged, do not dispose the damaged packaging or cushioning materials until the shipment has been checked for completeness and has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument.
RIGOL Chapter 1 Quick Guide Appearance and Dimensions DG800 User's Guide Front View Unit: mm Side View Unit: mm 1-3
RIGOL Chapter 1 Quick Guide Front Panel Overview 13 1 12 11 10 2 3 9 4 5 8 6 7 Figure 1-1 Front Panel 1. Power Key Turns on or off the generator. 2. Align Key Performs the phase alignment operation. For details, refer to descriptions in "Align Phase". 3. CH1 Output Connector BNC connector, with 50 Ω nominal output impedance. When Output1 is enabled (the backlight turns on), this connector outputs waveforms according to the current configuration of CH1. 4.
Chapter 1 Quick Guide RIGOL 5. CH2 Output Connector BNC connector, with 50 Ω nominal output impedance. When Output2 is enabled (the backlight turns on), this connector outputs waveforms according to the current configuration of CH2. 6. Counter Measurement Signal Input Connector BNC connector, with 1 MΩ input impedance. It is used to receive the signal measured by the counter. CAUTION To avoid damages to the instrument, the input signal voltage cannot exceed ±2.5 V. 7.
RIGOL Chapter 1 Quick Guide to select the desired parameter. Then, rotate the knob to modify the parameter, and press it down again to confirm your modification. — It is used to select the desired configuration type in the Preset interface. Press down the knob to confirm your selection. At this time, a dialog box is displayed.
Chapter 1 Quick Guide RIGOL Note: When the instrument is in the remote mode, press this key to return to the local mode. 13. LCD 4.3-inch TFT (480×272) color LCD display. The menu label and parameter settings of the current function, system state, prompt messages, and other information can be clearly displayed on the LCD. For details, refer to descriptions in "User Interface".
RIGOL Chapter 1 Quick Guide Rear Panel Overview 1 2 3 4 5 6 Figure 1-2 Rear Panel 1. [10MHz In/Out] BNC female connector, with 50 Ω nominal impedance. Its function is determined by the clock type used by the instrument. 1) When internal clock source is selected, this connector (as 10MHz Out) outputs the 10 MHz clock signal generated by the internal crystal oscillator inside the generator.
Chapter 1 Quick Guide RIGOL 3) FSK When ASK, FSK, or PSK of CH1 is enabled and external modulation source is selected, this connector receives an external modulation signal whose polarity can be set by users. Its input impedance is 1000 Ω. For details, refer to descriptions in "Modulation". 4) Trig In When Sweep or Burst of CH1 is enabled and external trigger source is selected, this connector receives an external trigger signal whose polarity can be set by users.
RIGOL Chapter 1 Quick Guide screen can also be saved to the USB storage device in the format of an image (*.Bmp). TMC DS: seamlessly interconnects with the RIGOL DS that meets the TMC standard. Reads and stores the waveform data collected by the DS and rebuilds waveforms without distortion. USB-GPIB interface converter (optional accessory): extends the GPIB interface for RIGOL instruments that integrates the USB HOST interface but not the GPIB interface. 5.
Chapter 1 Quick Guide RIGOL To Prepare for Use To Connect to AC Power Please use the power cord provided in the accessories to connect the signal generator to the AC power source, as shown in the figure below. The rated AC power source supported by the signal generator is (100-127 V, 45-440Hz) or (100-240 V, 45-65Hz), and its maximum input power shall not exceed 30 W.
RIGOL Chapter 1 Quick Guide User Interface The DG800 user interface is shown in the following figure. 6 5 4 7 8 9 10 1 1. 2 Figure 1-1 User Interface 3 Channel Output Configuration Status Bar Displays the current output configuration of the channel. Selected Waveform: Sine/Square/Ramp/Pulse/Noise/Prbs/Dualtone/Harm /Rs232/DC/Arb/Sequence Modulation Type: AM/FM/PM/ASK/FSK/PSK/PWM Sweep Type: Linear/Log/Step Burst Type: Ncycle/Infinite/Gated Channel Output State: ON: illuminated in yellow.
Chapter 1 Quick Guide RIGOL 2. Interface Switchover When left point is grayed out and the right point is illuminated, sliding right on the touch screen can switch to the waveform selection interface. When the left point is illuminated and the right point is grayed out, sliding left on the touch screen can switch to the current interface of waveform parameter setting. 3. Information Setting : opens the Store interface. : opens the Utility interface. : performs the channel copy function.
RIGOL Chapter 1 Quick Guide 10. Phase Displays the phase of the current waveform of each channel. Tap the Phase parameter input field to modify the parameter with the pop-up numeric keypad. You can also use the arrow keys and the knob to modify the parameter.
Chapter 1 Quick Guide RIGOL To Use the Built-in Help System DG800 series provides the help information for each front-panel function menu and the current display interface. You can view the help information if you have any questions during the operation process. 1. Obtain the help information of the front panel keys To get the help information about any front-panel key or menu softkey, press the Help/Local key first, and then press the desired key for the help information.
Chapter 2 Front Panel Operations RIGOL Chapter 2 Front Panel Operations Introduces the main functions and operation methods of DG800.
RIGOL Chapter 2 Front Panel Operations To Output Basic Waveform DG800 series can output basic waveforms (including Sine, Square, Ramp, Pulse, and Noise) from one channel or from both two channels at the same time. At start-up, the dual channels are configured to output a sine waveform with 1 kHz frequency and 5 Vpp amplitude by default. Users can configure the instrument to output various basic waveforms.
RIGOL Chapter 2 Front Panel Operations To Select Basic Waveform DG800 can output 5 basic waveforms, including Sine, Square, Ramp, Pulse and Noise. Press Menu on the front panel, then tap the touch screen to select the desired waveform. After selecting the desired waveform, you are automatically directed to the waveform parameter setting interface. To go back to the waveform selection interface, tap to slide right or press the Menu key on the front panel. At start-up, Sine is selected by default.
RIGOL Chapter 2 Front Panel Operations To Set Frequency/Period Frequency is one of the most important parameters of basic waveforms. For different instrument models and waveforms, the setting ranges of frequency are different. The default frequency is 1 kHz.
RIGOL Chapter 2 Front Panel Operations To Set Amplitude/High Level The amplitude setting range is limited by the "Impedance" and "Freq/Period" settings. By default, it is 5 Vpp. The amplitude displayed on the screen is the default value or the amplitude previously set. When the instrument configuration (e.g.
RIGOL Chapter 2 Front Panel Operations According to the figure above, the conversion relation between Vpp and Vrms fulfills the following equation: Vpp = 2 2Vrms For example, if the current amplitude is 5 Vpp, input a decimal point in the numeric keypad, and then select a unit Vrms. Tap "Ok". Then, it can be converted to a value with Vrms as the unit. For sine waveform, the converted value is 1.768 Vrms.
Chapter 2 Front Panel Operations RIGOL To Set Offset/Low Level The DC offset setting range is limited by the "Impedance" and "Amplitude/High Level" settings. The default value is 0 Vdc. The DC offset voltage displayed on the screen is the default value or the offset previously set. When the instrument configuration (e.g.
RIGOL Chapter 2 Front Panel Operations To Set Start Phase The setting range of start phase is from 0° to 360°. The default is 0°. The start phase displayed on the screen is the default value or the phase previously set. When the instrument function is changed, the new function will still use this phase. Tap the Phase parameter input field to input the desired phase value with the numeric keypad, then select the unit "°". Tap "Ok".
Chapter 2 Front Panel Operations RIGOL To Set Duty Cycle (Square) Duty cycle is defined as the percentage that the high level takes up in the whole period (as shown in the figure below). This parameter is only available when Square or Pulse is selected. Duty Cycle=t/T*100% t T The available range of the duty cycle is from 0.01% to 99.99% (limited by the current frequency setting). The default is 50%.
RIGOL Chapter 2 Front Panel Operations To Set Symmetry (Ramp) Symmetry is defined as the percentage that the rising period of the ramp takes up in the whole period (as shown in the figure below). This parameter is only available when Ramp is selected. Symmetry=t/T*100% t T The setting range of symmetry is from 0% to 100%. The default is 50%. Tap the Symm parameter input field to input the desired symmetry value with the numeric keypad, then select the unit "%". Tap "Ok".
RIGOL Chapter 2 Front Panel Operations To Set Pulse Width/Duty Cycle (Pulse) Pulse width is defined as the time from the 50% threshold of a pulse's rising edge to the 50% threshold of the next falling edge (as shown in the figure below). tFall tRise 90% 50% 10% Pulse Pulse Period The settable range of the pulse width is from 16 ns to 999.999982118 ks (limited by "minimum pulse width" and "pulse period"). The default is 500 μs.
RIGOL Chapter 2 Front Panel Operations To Set Rising/Falling Edge (Pulse) The rising edge time is defined as the duration of the pulse amplitude rising from 10% to 90% threshold, while falling edge time is defined as the duration of the pulse amplitude moving down from 90% to 10% threshold (as shown in the figure below). The setting range of rising/falling edge time is limited by the currently specified pulse width limit (as shown in the formula below).
Chapter 2 Front Panel Operations RIGOL To Enable Channel Output After configuring the parameters of the waveform selected, enable the channel to output the waveforms. or and Before enabling channel output, you can also tap then tap the corresponding menu label in the channel setting interface to set the parameters related with the channel output. For details, refer to descriptions in "Channel Setting". Press Output1 or Output2 on the front panel, and then the backlight turns on.
RIGOL Chapter 2 Front Panel Operations Align Phase DG800 series dual-channel function/arbitrary waveform generator enables you to align the phases of the two channels. Press Align on the front panel, then the instrument will re-configure the two channels and enable the generator to output with specified frequency and start phase. For two signals whose frequencies are the same or in multiple relationship, this operation can align their phases.
RIGOL Chapter 2 Front Panel Operations Example: To Output Sine This section mainly introduces how to output Sine waveforms (frequency 20 kHz, amplitude 2.5 Vpp, offset 500 mVdc, start phase 90°) from the [CH1] connector. 1. Select the output channel: Press Output1 on the front panel or tap the channel output configuration status bar to select CH1. 2. Select Sine waveforms: Press Menu on the front panel, and then the waveform selection interface is displayed.
RIGOL Chapter 2 Front Panel Operations Figure 2-3 Sine Waveform 2-16 DG800 User's Guide
Chapter 2 Front Panel Operations RIGOL To Output the Arbitrary Waveform DG800 can output built-in waveforms from a single channel or from two channels at the same time. The 160 kinds of built-in arbitrary waveforms are stored in the internal non-volatile memory. To Enable Arbitrary Waveforms Press Menu Continuous "Arb" to enable arbitrary waveform function and open the arbitrary waveform selection interface, as shown in the figure below.
RIGOL Chapter 2 Front Panel Operations Built-in Waveform DG800 has 160 built-in arbitrary waveforms, as shown in Table 2-2. In the arbitrary waveform selection interface, tap the Engineering, Medical, AutoElec, or Maths menu label to select the corresponding type. Tap or rotate the knob (pressing the right arrow key will locate the cursor to the right of the interface) to select the desired waveform (the selected waveform is highlighted).
RIGOL Chapter 2 Front Panel Operations Medical DG800 User's Guide Chshev2 TV Voice Surge Radar DualTone Ripple Quake Gamma StepResp BandLim CPulse CWPulse GateVibr LFMPulse MCNoise AM FM PFM PM PWM Chebyshev2 filter TV signal Voice signal Surge signal Radar signal Dual-tone signal Power ripple Analog quake waveform Gamma signal Step-response signal Bandwidth-limited signal C-Pulse signal CW pulse signal Gate self-oscillation signal Linear FM pulse signal Mechanical construction noise Sectioned sine AM
RIGOL Chapter 2 Front Panel Operations LFPulse Tens1 Tens2 Tens3 Waveform of the low frequency pulse electrotherapy Waveform 1 of the nerve stimulation electrotherapy Waveform 2 of the nerve stimulation electrotherapy Waveform 3 of the nerve stimulation electrotherapy AutoElec Ignition SP VR TP1 TP2A TP2B TP3A TP3B TP4 TP5A TP5B Maths 2-20 Airy Besseli Bessely Cubic Dirichlet Erf Erfc ErfcInv ErfInv ExpFall ExpRise HavSin Ignition waveform of the automotive motor Automotive starting profile with ring
Chapter 2 Front Panel Operations Laguerre Legend Versiera ARB_X2 Gauss Weibull LogNorm Laplace Maxwell Rayleigh Cauchy CosH CosInt Cot CotHCon CotHPro CscCon CscPro CscHCon CscHPro RecipCon RecipPro SecCon SecPro SecH SinH SinInt Sqrt Tan TanH AbsSine AbsSinH ACos ACosH ACotCon ACotPro ACotHCon ACotHPro ACscCon ACscPro ACscHCon ACscHPro ASecCon DG800 User's Guide RIGOL 4-times Laguerre polynomial 5-times Legend polynomial Versiera Square function Gaussian distribution or normal distribution Weibull distr
RIGOL Chapter 2 Front Panel Operations ASecPro ASecH ASin ASinH ATan ATanH Bartlett BarWin Blkman BlkmanH BohWin Boxcar ChebWin FlatWin Ham Hanning Kaiser NutWin ParWin TayWin Triang TukWin Protuberant arc secant Arc hyperbolic secant Arc Sinc Arc hyperbolic sine Arc tangent Arc hyperbolic tangent Bartlett window Modified Bartlett-Hann window Blackman window Blackman-Harris window Bohman window Rectangular window Chebyshev window Flat Top weighted window Hamming window Hanning window Kaiser window Nuttal
Chapter 2 Front Panel Operations RIGOL To Output Harmonic DG800 can be used as a harmonic generator to output harmonic with specified order, amplitude and phase. It is usually used in the test of harmonic detector device or harmonic filter device. This section introduces how to configure the generator to output harmonic.
RIGOL Chapter 2 Front Panel Operations To Set Fundamental Waveform Parameters DG800 allows users to set various fundamental waveform parameters such as frequency, period, amplitude, DC offset voltage, high level, low level, and start phase. It also supports align phase operation. To set the above fundamental waveform parameters, refer to descriptions in "To Output Basic Waveform". To Select Harmonic Type DG800 can output even harmonic, odd harmonic, both harmonics, and user-defined order of harmonic.
RIGOL Chapter 2 Front Panel Operations frequency, an integer. The maximum is 8. You can also use the arrow keys and the knob to set the parameter value: press down the knob to enter the editing mode, use the arrow keys to move the cursor to select the digit to be edited, and then rotate the knob to modify the value. Besides, you can use the knob to switch between different parameters.
RIGOL Chapter 2 Front Panel Operations 2. Set fundamental waveform parameters: In this example, various parameters of the fundamental waveform take the default values. You can also refer to descriptions in "To Output Basic Waveform" to set various fundamental waveform parameters such as frequency/period, amplitude/high level, offset/low level, and start phase. 3. Enable harmonic function: Press Menu on the front panel, and then the waveform selection interface is displayed.
RIGOL Chapter 2 Front Panel Operations Figure 2-5 Harmonic DG800 User's Guide 2-27
RIGOL Chapter 2 Front Panel Operations DC DG800 can output DC signal with amplitude range from -10 V to 10 V (HighZ) or from -5 V to 5 V (when the load is 50 Ω). The figure below shows a sketch of the DC signal. Press Menu Continuous "DC" to select the DC signal. Figure 2-6 Sketch of DC signal Tap the Offset parameter input field to input the desired offset value with the numeric keypad, then select the unit. Tap "Ok".
Chapter 2 Front Panel Operations RIGOL To Output Dual-tone Waveform Press Menu Continuous "Dualtone" to enter the dual-tone signal parameter setting interface. 1. Freq1/CentFreq: sets the output frequency/period of Signal 1 in the dual-tone signal. 2. Freq2/OffsFreq: sets the output frequency/period of Signal 2 in the dual-tone signal. 3. Ampl/HighL: sets the output amplitude/high level of the dual-tone signal. 4. Offset/LowL: sets the output offset/low level of the dual-tone signal.
RIGOL Chapter 2 Front Panel Operations To Output Advanced Waveform DG800 provides three advanced waveforms: PRBS, RS232, and Sequence. Note: When one channel is set to one of the advanced waveforms, the other channel will automatically switch to the set advanced waveform from other waveform. PRBS A PRBS is generated by a linear-feedback shift register (LFSR), as shown below.
Chapter 2 Front Panel Operations RIGOL RS232 In RS232 serial protocol, a character is transmitted as a frame of data. The frame consists of 1 start bit, 5-8 data bits, 0-1 check bit, and 1-2 stop bits. Its format diagram is as shown in the figure below. RS232 uses "Negative Logic", i.g. high level is logic "0" and low level is logic "1". Press Menu Advanced "RS232" to open the RS232 waveform parameter setting interface. 1. Baud Rate: sets the output baud rate of RS232 waveform. 2.
RIGOL Chapter 2 Front Panel Operations Sequence You can self-define the sequence, and the edited waveforms can be stored in the internal or external memory of the instrument in "*.SEQ" format. Press Menu Advanced "Sequence" to open the Sequence parameter setting interface. Tap the icon at the upper-right corner of the interface to enter the Sequence editing interface, as shown in the figure below. Figure 2-7 Sequence Editing Interface No.: indicates the serial number represented by "n".
Chapter 2 Front Panel Operations 5. Phase: sets the output start phase of Sequence. 6. Edge Time: sets the edge time of Sequence. RIGOL Please refer to "To Output Basic Waveform" to configure the parameters and output for the channel.
RIGOL Chapter 2 Front Panel Operations Modulation DG800 can output modulated waveforms from a single channel or from dual channels at the same time. Modulation is the process of modifying certain parameters (e.g. amplitude, frequency, or phase) of the carrier waveform signal according to the changes of the modulating signal. The carrier waveform can be Sine, Square, Ramp, Arbitrary waveform, or Pulse (only in PWM). The modulating waveform can be from internal or external modulation source.
Chapter 2 Front Panel Operations RIGOL When the carrier waveform currently selected is Sine, Square, or Ramp, please refer to descriptions in "To Output Basic Waveform" to set the parameters of the carrier waveform currently selected. When the carrier waveform currently selected is arbitrary waveform, please refer to descriptions in "To Output the Arbitrary Waveform" to set the parameters of the carrier waveform currently selected.
RIGOL Chapter 2 Front Panel Operations Key Points: How to realize the intermodulation between dual channels? The following example takes the output signal of CH2 as the modulating waveform. 1. Connect the CH2 output terminal to the rear-panel [CH1/Sync/Ext Mod/Trig/FSK] connector by using the dual BNC cable. 2. Select CH1, then press Menu Modulation to select the desired modulation type and set the corresponding parameters. After that, select the external modulation source. 3.
Chapter 2 Front Panel Operations RIGOL Carrier Waveform Suppression DG800 supports two types of amplitude modulation: normal amplitude modulation and double sideband suppressed carrier (DSSC) amplitude modulation. In the normal amplitude modulation, the modulated waveform contains carrier waveform components. Because carrier waveform components carry no information, the modulation is less efficient.
RIGOL Chapter 2 Front Panel Operations Frequency Modulation (FM) For frequency modulation (FM), the frequency of the carrier waveform varies with the instantaneous voltage of the modulating waveform. To Select FM Modulation Press Menu Modulation "FM" to enable the FM function. When Modulation is enabled, Sweep or Burst will be automatically disabled (if enabled currently).
Chapter 2 Front Panel Operations RIGOL To Select Modulation Source DG800 can receive modulating waveform from the internal or external modulation source. Tap the Source parameter selection field to select an "Internal" or "External" modulation source. 1. Internal Source When internal modulation source is selected, tap the Mod.Wave parameter selection field to select Sine, Square, Triangle, UpRamp, DnRamp, Noise, or Arb as modulating waveform. By default, Sine is selected. Square: 50% duty cycle.
RIGOL Chapter 2 Front Panel Operations The frequency deviation must be smaller than or equal to the carrier frequency. The sum of frequency deviation and carrier frequency must be smaller than or equal to the the upper limit of the current carrier frequency plus 1 kHz.
Chapter 2 Front Panel Operations RIGOL Phase Modulation (PM) For phase modulation (PM), the phase of the carrier waveform varies with the instantaneous voltage of the modulating waveform. To Select PM Modulation Press Menu Modulation "PM" to enable the PM function. When Modulation is enabled, Sweep or Burst will be automatically disabled (if enabled currently).
RIGOL Chapter 2 Front Panel Operations To Select Modulation Source DG800 can receive modulating waveform from the internal or external modulation source. Tap the Source parameter selection field to select an "Internal" or "External" modulation source. 1. Internal Source When internal modulation source is selected, tap the Mod.Wave parameter selection field to select Sine, Square, Triangle, UpRamp, DnRamp, Noise, or Arb as modulating waveform. By default, Sine is selected. Square: 50% duty cycle.
Chapter 2 Front Panel Operations RIGOL To Set Phase Deviation Phase deviation indicates the deviation of the modulating waveform phase from the carrier waveform phase. Tap the Phas.Dev parameter input field and use the numeric keypad to set the PM phase deviation. You can also use the arrow keys and the knob to input a desired phase value. The setting range of the phase deviation is from 0° to 360°, and the default is 90°.
RIGOL Chapter 2 Front Panel Operations Amplitude Shift Keying (ASK) When using ASK (Amplitude Shift Keying), you can configure the generator to "shift" its output amplitude between two preset amplitude values ("carrier amplitude" and "modulating amplitude"). To Select ASK Modulation Press Menu Modulation "ASK" to enable the ASK function. When Modulation is enabled, Sweep or Burst will be automatically disabled (if enabled currently).
Chapter 2 Front Panel Operations RIGOL To Select Modulation Source Tap the Source parameter selection field to select an "Internal" or "External" modulation source. 1. Internal Source When the internal source is selected, the modulating waveform is set as a Square with 50% duty cycle. At this time, the rate at which the output amplitude "shifts" between "carrier amplitude" and "modulating amplitude" is determined by the modulation rate. 2.
RIGOL Chapter 2 Front Panel Operations To Set the Modulation Polarity Tap the Polarity parameter selection field to select "Positive" or "Negative" polarity of the modulating waveform to control the output amplitude. In the internal modulation, set the polarity to "Positive", and the generator will output an amplitude whichever is smaller between the carrier amplitude and modulating amplitude when the modulating waveform is a logic low level.
Chapter 2 Front Panel Operations RIGOL Frequency Shift Keying (FSK) When using FSK (Frequency Shift Keying), you can configure the generator to "shift" its output frequency between two preset frequency values ("carrier amplitude" and "hop frequency"). To Select FSK Modulation Press Menu Modulation "FSK" to enable the FSK function. When Modulation is enabled, Sweep or Burst will be automatically disabled (if enabled currently).
RIGOL Chapter 2 Front Panel Operations To Select Modulation Source Tap the Source parameter selection field to select an "Internal" or "External" modulation source. 1. Internal Source When the internal source is selected, the modulating waveform is set as a Square with 50% duty cycle. At this time, the rate at which the output frequency "shifts" between "carrier frequency" and "hop frequency" is determined by the modulation rate. 2.
Chapter 2 Front Panel Operations RIGOL To Set the Modulation Polarity Tap the Polarity parameter selection field to select "Positive" or "Negative" polarity of the modulating waveform to control the output frequency. In the internal modulation, set the polarity to "Positive", and the generator will output the carrier frequency when the modulating waveform amplitude is a logic low level. It will output the hop frequency when the modulating waveform amplitude is a logic high level.
RIGOL Chapter 2 Front Panel Operations Phase Shift Keying (PSK) When using PSK (Phase Shift Keying), you can configure the generator to "shift" its output phase between two preset phase values ("carrier phase" and "modulating phase"). To Select PSK Modulation Press Menu Modulation "PSK" to enable the PSK function. When Modulation is enabled, Sweep or Burst will be automatically disabled (if enabled currently).
Chapter 2 Front Panel Operations RIGOL To Select Modulation Source Tap the Source parameter selection field to select an "Internal" or "External" modulation source. 1. Internal Source When the internal source is selected, the modulating waveform is set as a Square with 50% duty cycle. At this time, the rate at which the output phase "shifts" between "carrier phase" and "modulating phase" is determined by the modulation rate. 2.
RIGOL Chapter 2 Front Panel Operations To Set Modulation Polarity Tap the Polarity parameter selection field to select "Positive" or "Negative" polarity of the modulating waveform to control the output phase. In the internal modulation, set the polarity to "Positive", and the generator will output the carrier phase when the modulating waveform amplitude is a logic low level. It will output the modulating phase when the modulating waveform amplitude is a logic high level.
Chapter 2 Front Panel Operations RIGOL Pulse Width Modulation (PWM) For PWM (Pulse Width Modulation), the pulse width of the carrier waveform varies with the instantaneous voltage of the modulating waveform. To Select PWM Modulation PWM can only be used to modulate pulse. To select PWM modulation, first select Pulse, then press Menu Modulation "PWM" to enable PWM function. Tap the icon at the upper-right corner of the interface to view and set the parameters of the current carrier waveform.
RIGOL Chapter 2 Front Panel Operations To Select Modulation Source Tap the Source parameter selection field to select an "Internal" or "External" modulation source. 1. Internal Source When internal modulation source is selected, tap the Mod.Wave parameter selection field to select Sine, Square, Triangle, UpRamp, DnRamp, Noise, or Arb as modulating waveform. By default, Sine is selected. Square: 50% duty cycle. Triangle: 50% symmetry. UpRamp: 100% symmetry. DnRamp: 0% symmetry.
Chapter 2 Front Panel Operations RIGOL To Set Width/Duty Deviation If "Width" is currently selected in the pulse of the specified channel, "Width Dev" is displayed in the interface when the PWM modulation function is enabled; if the "Duty" is currently selected in the pulse of the specified channel, "Duty Dev" is displayed in the interface when the PWM modulation function is enabled. Tap the Width Dev (or Duty Dev) parameter input field and use the numeric keypad to input a desired value.
RIGOL Chapter 2 Front Panel Operations Sweep DG800 can output sweep waveforms from a single channel or from dual channels at the same time. In sweep mode, the generator outputs a signal variably from the start frequency to stop frequency within the specified sweep time.
RIGOL Chapter 2 Front Panel Operations Log Sweep In Log Sweep type, the output frequency of the instrument varies in a logarithmic fashion, that is, the output frequency changes in the way of "octave per second" or "decade per second". The variation is controlled by "Start Frequency", "Stop Frequency", and "Sweep Time". When Log Sweep is enabled, you can set the following parameter: start frequency Fstart , stop frequency Fstop , and sweep time Tsweep .
RIGOL Chapter 2 Front Panel Operations increase. At this point, tap to slide the screen or use the knob to page up or down. Tap the Step parameter input field and use the numeric keypad to set the step. You can also use the arrow keys and the knob to set the step. By default, it is 2, and its settable range is from 2 to 1024. Note: In the "Linear" and "Log" sweep modes, the Step menu label is grayed out and disabled.
RIGOL Chapter 2 Front Panel Operations Start Frequency and Stop Frequency Start frequency and stop frequency are the upper and lower limits of the frequency for frequency sweep. The generator always sweeps from the start frequency to the stop frequency, and then returns back to the start frequency. Start Frequency < Stop Frequency: the generator sweeps from low frequency to high frequency. Start Frequency > Stop Frequency: the generator sweeps from high frequency to low frequency.
RIGOL Chapter 2 Front Panel Operations Ramp: 1 μHz to 1 MHz Arb: 1 μHz to 10 MHz The range of the frequency span is affected by the center frequency. Center frequency < Fm : the range of the frequency span is ±2× (center frequency – Fmin ); Center frequency ≥ Fm : the range of the frequency span is ±2× ( Fmax - center frequency). Take sine as an example. Fmin is 1 μHz, Fmax is 35 MHz, and Fm is 17.5 MHz. If the center frequency is 550 Hz, the frequency span range is ±2×(550 Hz–1 μHz) = ±1.
Chapter 2 Front Panel Operations 3. RIGOL Manual Trigger In the Manual trigger, when you press Trig on the front panel for one time, one sweep is immediately launched for the corresponding channel. Key Points: When "External" trigger is selected, tap the Trig In to select "RisEdge" or "FallEdge". The rear-panel [CH1/Sync/Ext Mod/Trig/FSK] connector is used as the input terminal of the external trigger signal.
RIGOL Chapter 2 Front Panel Operations Start Hold Start hold is the duration that the output signal outputs with the "Start Frequency" after the sweep starts. After the start hold time expires, the generator will continue outputting with varied frequencies according to the current sweep type. When the sweep function is enabled, tap the StartHold menu label and use the numeric keypad to modify the start hold time. You can also use the arrow keys and the knob to modify the start hold time.
Chapter 2 Front Panel Operations RIGOL Burst DG800 can output the waveforms with a specified number of cycles (Burst) from a single channel or from dual channels at the same time. DG800 supports control of burst output by internal, manual or external trigger source; supports three kinds of burst types, including N cycle, Infinite, and Gated. The generator can generate burst by using Sine, Square, Ramp, Pulse, Noise (only for gated burst), arbitrary waveform, PRBS, RS232, or Sequence.
RIGOL Chapter 2 Front Panel Operations Figure 2-11 N Cycle Burst For N cycle burst, "Internal", "External", or "Manual" trigger source could be selected. Besides, you can also set the parameters of "Period" (internal trigger), "Delay", "Trig In" (external trigger) and "Trig Out" (internal and manual trigger). Infinite Burst In Infinite mode, the cycle number of the waveform is set as an infinite value. The generator outputs a continuous waveform after receiving a trigger signal.
RIGOL Chapter 2 Front Panel Operations Gated Burst In Gated Burst Mode, the generator controls the waveform output according to the external signal level input from the rear-panel [CH1/Sync/Ext Mod/Trig/FSK] connector. Press Menu Burst "Gated" to enable the gated burst. Then tap the Polarity menu label to select "Positive" or "Negative". Only when the gated signal is Positive or Negative, can the waveforms be output.
RIGOL Chapter 2 Front Panel Operations Burst Period Burst period is only available for N cycle burst in the internal trigger. It is defined as the time from the start of a burst to the time when the next burst starts. Burst Period ≥ 1 μs + waveform period × number of bursts. Wherein, the waveform period indicates the period of burst function (e.g. Sine and Square).
Chapter 2 Front Panel Operations RIGOL Tap the User menu label, and input the desired value by using the numeric keypad. You can also use the arrow keys and the knob to input a desired value. Its range is from 0 to 65535. Burst Trigger Source Burst trigger source can be internal, external, or manual. The generator will generate a burst output when a trigger signal is received, and then it waits for the next trigger signal.
RIGOL Chapter 2 Front Panel Operations Gated Polarity Gated polarity is only available in gated burst mode. The instrument outputs the burst when the gated signal on the rear-panel [CH1/Sync/Ext Mod/Trig/FSK] connector is "High Level" or "Low Level". Press Menu Burst "Gated" to enable the gated burst. Tap the Polarity to select "Positive" or "Negative", and the default is "Positive".
Chapter 2 Front Panel Operations RIGOL Frequency Counter DG800 provides a frequency counter which can measure various parameters (e.g. frequency, period, duty cycle, positive pulse width, and negative pulse width) of the external input signal and make statistics of measurement results.
RIGOL Chapter 2 Front Panel Operations To Set the Frequency Counter 1. Gate Time Tap the GateTime parameter selection field to select the gate time for the measurement system. The default is "100ms". When "Auto" is selected, the instrument will select an appropriate gate time automatically according to the characteristics of the signal to be measured. 1 ms 1.048 ms 10 ms 8.389 ms 100 ms 134.218 ms 1s 1.074 s 10 s 8.590 s >10 s >8.590 s 2. To Select the Parameters to be Measured Tap the Meas.
Chapter 2 Front Panel Operations RIGOL (1) To Clear the Statistics Results Tap the "Clear" menu label, and the signal generator clears the current statistics results. (2) Operating Status Tap the Status parameter selection field to select the running status of the frequency counter. When you press Counter on the front panel, the frequency counter enters "Run" state automatically and it will continuously measure the input signal according to the current configuration.
RIGOL 7. Chapter 2 Front Panel Operations High Frequency Rejection High frequency rejection can be used to filter out the high-frequency components and improve the measurement accuracy in low-frequency signal measurement. Tap the High Freq Rejection parameter selection field to enable or disable the high frequency rejection function. Note: Enable the high frequency rejection when low-frequency signal with less than 200 kHz frequency is measured to filter out the high-frequency noise interference.
Chapter 2 Front Panel Operations RIGOL Store and Recall DG800 can store the current instrument state and user-defined Sequence waveforms in the internal or external memory. You can recall them when needed. DG800 can also seamlessly interconnect with RIGOL's oscilloscopes that support USB-TMC; rebuild and output the waveform data collected by the oscilloscope losslessly.
RIGOL Chapter 2 Front Panel Operations File Type The file types supported by DG800 include State File, Arb File, Csv File, and Bmp File. 1. State File Stores the instrument state in the internal or external memory in "*.RSF" format. The state file stored includes the dual-channel selected waveform, frequency, amplitude, offset, duty cycle, symmetry, phase, modulation, sweep, burst parameters, frequency counter parameters, as well as utility function parameters and system parameters under the Utility menu.
Chapter 2 Front Panel Operations RIGOL File Operation In the Disk C or D (when inserted with a USB storage device), you can perform a series of operations on the files, including Save, Read, Copy, Paste, and Delete. Save 1. To Select the Save Path Selects Disk C or D (when inserted with a USB storage device) as the current save path. You can tap Disk C or D to enter the current directory.
RIGOL Chapter 2 Front Panel Operations Tap the desired character, and the selected character will be displayed in the filename input area. Use the same method to input all the characters of the filename. You can also tap the delete the unwanted characters in the filename input area. Input Numbers in the filename input interface to switch to the numbers and Tap symbols selection interface, as shown in the following figure. The input method is similar with that of the English input.
Chapter 2 Front Panel Operations RIGOL the Pinyin input area. After inputting the Pinyin for a Chinese character, tap the desired Chinese character in the Chinese character display area (if no proper Chinese character exists, tap or to page down to search for the desired Chinese character). The selected Chinese character will be displayed in the filename input area. Use the same method to input all the the delete the Chinese characters of the filename.
RIGOL Chapter 2 Front Panel Operations When copying file from Disk C to Disk D, tap or use the knob to select the Disk D directory. When copying file from Disk D to Disk C, tap or use the knob to select the Disk C directory. 4. To Paste a File Tap Paste, the generator will paste the copied file to the currently selected directory. After that, a corresponding prompt message will be displayed, indicating that the paste operation is completed. Delete 1.
RIGOL Chapter 2 Front Panel Operations Seamless Interconnection with Oscilloscope DG800 can also seamlessly interconnect with RIGOL's oscilloscopes that support USB-TMC; rebuild and output the waveform data collected by the oscilloscope losslessly. The connection method between the instruments is as shown in the figure below. USB Cable BNC Cable DG800 DS1 USB DEVICE USB HOST DS2 Output Terminal Input Terminal Operation Procedures: 1.
RIGOL Chapter 2 Front Panel Operations is set to "Ram Data"), and its available choices include "Head Pos", "Trig Pos", and "Tail Pos". Tap or use the knob to select the channel to be read, and then set the parameters of the corresponding channel. Tap Read. At this time, DS1 will enter the "Stop" status automatically, and DG800 will read the arbitrary waveform data (i.g. the waveform data collected by DS1) automatically.
Chapter 2 Front Panel Operations RIGOL Channel Setting The function and setting methods of CH2 are the same as that of CH1. This section takes CH1 as an example to introduce the channel setting method. To set the CH2 channel output parameters, select CH2 (refer to "To Select Output Channel") and then set the channel according to the descriptions in this section. or under Tap the channel output configuration status bar the user interface to enter the channel setting interface.
RIGOL Chapter 2 Front Panel Operations Note: When the waveform is inverted, the sync signal related to the waveform is not inverted. To set the invert of the sync signal, tap SyncSet Polarity to select "Negative". 3. Output Impedance The output impedance setting affects the output amplitude and DC offset. The instrument has a 50 Ω fixed serial output impedance for the front-panel [CH1] connector.
Chapter 2 Front Panel Operations RIGOL Sync Setting DG800 can output the basic waveforms (except Noise and DC), arb waveforms, harmonics, sweep waveforms, burst waveforms, and the sync signal of the modulated waveforms from a single channel or the dual channels at the same time. The signal will be output from the rear-panel [CH1/Sync/Ext Mod/Trig/FSK] connector. 1. Sync Switch Enables or disables the sync signal on the [CH1/Sync/Ext Mod/Trig/FSK] connector.
RIGOL 3) Chapter 2 Front Panel Operations is (the basic waveform frequency÷2n). Wherein, n represents the frequency dividing coefficient and it is 1 when the frequency of the basic waveform is greater than 30 MHz and less than or equal to 60 MHz. Noise: there is no sync signal output. 2. Harmonic For harmonics, the sync signal is a Square with 50% duty cycle, and its frequency is the same as that of the fundamental waveform. When the first waveform point is output, the sync signal is TTL high level.
Chapter 2 Front Panel Operations 2) 6. RIGOL this marker frequency. When external trigger source is selected: The terminal is used to receive the external trigger signal, and there is no sync signal output. Burst Waveform 1) When the internal or manual trigger source is selected: Infinite burst: the sync signal is the same as that of the basic waveform.
RIGOL Chapter 2 Front Panel Operations desired value by using the numeric keypad, the arrow keys, or the knob. Freq Deviation: indicates the frequency deviation between CH1 and CH2. The relations among the parameters are as follows: FCH2=FCH1+FDev (take CH1 as the reference source); FCH1=FCH2+FDev (take CH2 as the reference source). Freq Ratio: indicates the ratio of the frequencies of CH1 to that of CH2.
Chapter 2 Front Panel Operations RIGOL Then, tap the Phase Dev or Phase Ratio parameter input field to input the desired value by using the numeric keypad. Phase Deviation: indicates the phase deviation between CH1 and CH2. The relations among the parameters are as follows: PCH2=PCH1+PDev (take CH1 as the reference source); PCH1=PCH2+PDev (take CH2 as the reference source). Phase Ratio: indicates the ratio of the phases of CH1 to that of CH2.
RIGOL Chapter 2 Front Panel Operations Figure 2-24 Channel Coupling Key Points: Channel coupling is only available when both the waveforms of the two channels are basic waveforms (Sine, Square, Ramp, or Arbitrary waveform). The generator will adjust the frequency, phase, or amplitude of the channel to avoid parameter over-limit if the following conditions occur: 1.
Chapter 2 Front Panel Operations RIGOL outputs the inverted signal of the output signal of CH1 (if channels are currently enabled). Off: disables the track function. This is the default state. Note: When the track function is enabled, the coupling function is disabled. Waveform Combination 1.
RIGOL Chapter 2 Front Panel Operations Common Settings Press Utility on the front panel or tap the icon in the information setting area at the lower-right corner of the user interface to open the following operation interface. The interface displays the system parameters of the currently selected channel. Figure 2-25 Utility Interface System Setting To Set the System Language At present, DG800 supports menus in multiple languages.
Chapter 2 Front Panel Operations RIGOL Clock Source DG800 provides an internal 10MHz clock source and receives the external clock source from the rear-panel [10MHz In/Out] connector. It can also output clock source from the [10MHz In/Out] connector for other devices. Press Utility System Setting Clk Source to select "Internal" or "External". The default is "Internal".
RIGOL Chapter 2 Front Panel Operations decimal point in the number parameter. It can be set to a decimal "." or a comma ",". By default, it is a decimal ".". This setting is stored in non-volatile memory and will not be affected by the "restore to defaults" operation. Delimiter Press Utility System Setting Delimiter to set the display format of the delimiter in the number parameter. This setting is stored in non-volatile memory and will not be affected by the "restore to defaults" operation.
RIGOL Chapter 2 Front Panel Operations To Set the LAN Parameters DG800 only uses the following model of the USB-LAN interface converter to expand the LAN interface. Manufacturer TP-LINK UGREEN Model TL-UF2101.0 177TL-UF210 1.0 177TH2 20254 Feature USB2.0, 100Mbps Press Utility Interface LAN to open the LAN parameters configuration interface as shown in the figure below. You can view the network status and configure the network parameters.
RIGOL Chapter 2 Front Panel Operations Tap the DHCP parameter selection field to select "On" or "Off" to turn DHCP mode on or off. 2. AutoIP In Auto IP mode, the generator acquires IP address within the range from 169.254.0.1 to 169.254.255.254, and gets the subnet mask 255.255.0.0 automatically according to the current network configuration. Tap the AutoIP parameter selection field to select "On" or "Off" to turn the AutoIP mode on or off.
Chapter 2 Front Panel Operations RIGOL VISA descriptor is the resource name and it describes the accurate name and location of the VISA resource. If LAN interface is currently used for communication with the instrument, the VISA descriptor is TCPIP0::192.168.001.102::INSTR. To Set the Subnet Mask In ManualIP mode, the subnet mask can be manually set. The format of the subnet mask is nnn.nnn.nnn.nnn. Wherein, the range of "nnn" is from 0 to 255.
RIGOL Chapter 2 Front Panel Operations Default Settings Tap Default to restore the network parameters to the default settings. By default, DHCP and AutoIP are enabled and ManualIP is disabled. Apply Tap Apply to apply the currently set network parameters. System Info Press Utility System Info, and the model, serial number, and software version number of the current instrument are displayed in the system information interface.
Chapter 2 Front Panel Operations RIGOL interface. Select the desired file in BMP format and tap Read. Note: The internal memory cannot store the file in BMP format. You need to store the contents in the USB storage device in BMP format first. The size of the image cannot exceed 480*272. Print Setting You can store the contents displayed on the screen in the external USB storage device in picture format. 1. Insert a USB storage device into the USB HOST interface on the rear panel. 2.
RIGOL Chapter 2 Front Panel Operations System Utility Function To Restore Preset Press Preset to restore the preset status interface, as shown in the figure below. Figure 2-27 Preset Status Restore Interface Factory: Tap the Def icon, then a dialog box is displayed, tap "Apply" to restore the system to its factory defaults. User save: Tap any one of the items from "User1" to "User10". Then a dialog box is displayed.
RIGOL Chapter 2 Front Panel Operations FreqCplMode Freq Dev Freq Ratio Ampl Cpl AmplCplMode Ampl Dev Ampl Ratio Phase Cpl PhaseCplMode Phase Dev Phase Ratio Trig Coupling Track Mode Combine Waveform Freq Ratio Deviation 0 μHz 1 Off Deviation 0 Vpp 1 Off Deviation 0° 1 Off Off Off Sine 1 kHz 10% Basic Waveforms Frequency Amplitude Amplitude Unit Offset Phase Square Duty Cycle Ramp Waveform Symmetry Pulse Duty Cycle Pulse Width Pulse Rising Edge Pulse Falling Edge Harmonic Type Harmonic Order Harmonic Pha
RIGOL RS232 Baud Rate RS232 Data Bits RS232 Stop Bits RS232 Parity Bit RS232 Data Sequence Filter Sequence Sample Rate Sequence Phase Modulation AM Modulation Source Mod.Wave Mod.Freq Mod.Depth DSSC FM Modulation Source Mod.Wave Mod.Freq Freq.Dev PM Modulation Source Mod.Wave Mod.Freq Phas.Dev ASK Modulation Source Mod.Rate Mod.Ampl Polarity FSK Modulation Source Mod.Rate Hop Freq Polarity PSK Modulation Source Mod.Rate Phase Polarity PWM Modulation Source Mod.
RIGOL Chapter 2 Front Panel Operations Mod.
RIGOL Coupling High Freq Rejection System Parameter System Setting Language* Power-on Clk Source Beeper Decimal Delimiter System Log DispSet Screen Saver Brightness* Print Setting Location Format* Interface GPIB* DHCP* Auto IP* Manual IP* Chapter 2 Front Panel Operations DC Off Determined by Factory Delivery Setting Default Internal On Dot Comma Off Off 100 USB Storage Device PNG 2 On (default setting in LAN) On (default setting in LAN) Off (default setting in LAN) Channel Copy DG800 supports states an
RIGOL Chapter 2 Front Panel Operations To Install the Option DG800 provides various options (for option information, refer to "Accessories and Options") to expand the function of the signal generator. If you need to purchase the option, please contact RIGOL sales representative. If you have ordered this option, you can obtain the corresponding key. Perform the following operations to install the option. 1. Acquire the license of the option Log in to the RIGOL official website (www.rigol.
RIGOL Chapter 2 Front Panel Operations 2) Install the option by sending SCPI Commands Open the remote control window and send the following option installation commands by referring to "Remote Control". :LICense:SET or :LICense:INSTall Wherein, is the option license (note that the hyphens should be omitted).
Chapter 2 Front Panel Operations RIGOL To Lock the Keyboard You can lock any key or all the keys on the front panel by using the keyboard lock command. 1. Keyboard Lock Command Introduction :SYSTem:KLOCk ,{ON|OFF|0|1} /*Locks or unlocks the specified key.*/ :SYSTem:KLOCk? /*Queries whether the specified key is locked.
Chapter 3 Remote Control RIGOL Chapter 3 Remote Control DG800 can communicate with PC through USB, LAN, and GPIB (with the USB to GPIB interface converter provided by RIGOL) interface. The remote control can be realized by using SCPI (Standard Commands for Programmable Instruments) commands. It can be realized through the following two methods: 1.
RIGOL Chapter 3 Remote Control Remote Control via USB 1. Connect the device Use the USB cable to connect the rear-panel USB DEVICE interface of the signal generator to the USB HOST interface of the PC. 2. Search for the device resource Start up the Ultra Sigma and the software will automatically search for the generator resources currently connected to the PC. You can also click to search the resources. 3.
RIGOL Chapter 3 Remote Control (a) (b) 4. View the device resource The resources found are shown under the "RIGOL Online Resource" directory. For example, DG832 (TCPIP::172.16.3.82::INSTR). 5. Communication test Right-click the resource name "DG832 (TCPIP::172.16.3.82::INSTR)" and select "SCPI Panel Control" to turn on the remote command control panel, through which you can send commands and read data.
RIGOL Chapter 3 Remote Control Remote Control via GPIB 1. Connect the device Connect the generator to your computer by using a USB-GPIB interface converter. Note: Please make sure that a GPIB card has been installed to your PC. Connect the USB terminal of the USB-GPIB interface converter to the USB HOST interface on the rear panel of the generator. Connect its GPIB terminal to the GPIB card terminal of the PC. 2.
Chapter 3 Remote Control RIGOL (b) If the resource cannot be searched automatically: Select the GPIB card address of the PC from the drop-down box of "GPIB0::" and select the GPIB address set in the signal generator from the drop-down box of "0::INSTR". Click "Test" to check whether the GPIB communication works normally. If not, please follow the corresponding prompt messages to solve the problem. 5. View the device resource to go back to the main interface of Ultra Sigma.
Chapter 4 Troubleshooting RIGOL Chapter 4 Troubleshooting The commonly encountered failures of DG800 and their solutions are listed below. If the following problems occur, locate and resolve the problems according to the following steps. If the problems still persist, contact RIGOL and provide your instrument information to us. (For instrument information, press Utility System Info.) 1. When I power on the instrument, the instrument stays black and does not display anything.
RIGOL 3) 4) Chapter 4 Troubleshooting After restarting the instrument, insert the USB storage device again to check whether it can work normally. If the USB storage device still cannot work normally, please contact RIGOL. 6. How do you set the amplitude of the waveform in dBm? 1) Select the desired channel. 2) In the channel setting interface, check whether OutputSet HighZ is set to "On". If yes, you cannot set the amplitude of the waveform in dBm at this time.
RIGOL Chapter 5 Appendix Chapter 5 Appendix Appendix A: Accessories and Options Model Standard Accessories Option Optional Accessories Description DG812 (10 MHz, Dual-channel) DG822 (25MHz, Dual-channel) DG832 (35MHz, Dual-channel) DG811 (10MHz, Single-channel) DG821 (25MHz, Single-channel) DG831 (35MHz, Single-channel) 1 Power Cord conforming to the standard of the destination country 1 BNC Cable (only provided by DG832/DG831/DG822/DG821) 1 Quick Guide 1 Product Warranty Card Single-dual CH Upgrade O
RIGOL Chapter 5 Appendix Appendix B: Warranty RIGOL (SUZHOU) TECHNOLOGIES INC. (hereinafter referred to as RIGOL) warrants that the product will be free from defects in materials and workmanship within the warranty period. If a product proves defective within the warranty period, RIGOL guarantees free replacement or repair for the defective product. To get repair service, please contact with your nearest RIGOL sales or service office.
Index RIGOL Index *.RAF ...................................... 2-74 *.RSF ...................................... 2-74 Align Phase .............................. 2-14 AM .......................................... 2-34 Amplitude .................................. 2-5 Amplitude Coupling ................... 2-86 Arb File .................................... 2-74 ASK ......................................... 2-44 AutoIP ..................................... 2-94 Bmp File ..................................
RIGOL VISA Address ........................... 2-94 Waveform Combination ............. 2-89 2 Index Width Deviation ........................