RIGOL Chapter 1 Quick Start
1-14 DL3000 User’s Guide
Fuse Replacement
The fuse specification is related to the actual input voltage, as shown in Table 1-2.
Table 1-2 Fuse Rating
Input Voltage
Fuse Rating
115 Vac
T0.315 A/250 V
230 Vac
T0.20 A/250 V
When leaving the factory, the instrument has installed a fuse that conforms to the
local standard. If the fuse is required to be replaced, select a fuse that matches the
actual input voltage and perform the following steps.
1. Turn off the instrument, and then disconnect the instrument from the power
source. Remove the power cord.
2. Insert a slotted screwdriver into the slot of the fuse holder to pry it out, as
shown in Figure 1-7.
Figure 1-7 Fuse Replacement
3. Take out the fuse and replace it according to the fuse rating (refer to the "Fuse
Rating" on the rear panel of the instrument or Table 1-2).
4. Mount the fuse holder (pay attention to its direction when mounting it).
WARNING
To avoid personal injuries, turn off the instrument and cut off the power
supply before replacing the fuse; to avoid electric shock or fire disaster,
before connecting to the AC power, select the voltage specification that
matches the actual input voltage and replace a proper fuse that
conforms to the voltage specification.
Fuse Holder
Fuse