RIGOL Service Guide DM3058 Digital Multimeter Jan. 2009 RIGOL Technologies, Inc.
RIGOL Copyright © 2008 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is registered trademark of RIGOL Technologies, Inc. Notices RIGOL products are protected by patent law in and outside of P.R. China. RIGOL Technologies, Inc. reserves the right to modify or change part of or all the specifications and pricing policies at company‟s sole decision. Information in this publication replaces all previously corresponding material.
RIGOL Safety Notices Review the following safety precautions carefully before operating the instrument to avoid any personal injuries or damages to the instrument and any products connected to it. To avoid potential hazards use the instrument as specified by this user‟s guide only. Use Proper Power Cord. Only the power cord designed for the instrument and authorized by local country could be used. Ground The Instrument. The instrument is grounded through the grounding conductor of the power cord.
RIGOL Safety Terms and Symbols Terms in This Guide. These terms may appear in this guide: ! ! WARNING Warning statements indicate the conditions or practices that could result in injury or loss of life. CAUTION Caution statements indicate the conditions or practices that could result in damage to this product or other property. ! CAT I (1000 V) ! CAT II (600 V) IEC Measurement Category II. The maximum voltage can be measured by HI-LO terminal is 1000 Vpk. IEC Measurement Category II.
RIGOL DM3058 Digital Multimeter Overview RIGOL DM3058 is an instrument especially designed for the needs of high-precision, multifunction, and automation measurements. It realized a combination of basic measurement function and multifold math functions as well as Random Sensor measurement function. The DM3058 holds a high-resolution monochrome LCD display system with clear keyboard layout and operation hints to make it easier and agility to use.
RIGOL Structure of this Document Chapter 1 Performance & Specifications List the Performance specifications of DM3058 series. Chapter 2 Quick Start Help users to be familiar with the operating skills of DM3058. Chapter 3 Performance Test Introduce how to test the performance so as to know about its current state of DM3058 well. Chapter 4 Calibration Guide you how to calibrate DM3058.
RIGOL Contents Safety Notices ............................................................................................. II DM3058 Digital Multimeter Overview ........................................................... III Chapter 1 Performance & Specifications ................................................. 1-1 General Technical Specifications ................................................................. 1-2 Electric Technical Specifications .......................................................
RIGOL The Disassembly and Assembly Notices ...................................................... 5-2 The 3D View of DM3058 ............................................................................ 5-3 To Disassemble and Assemble Handle ........................................................ 5-5 To Disassemble and Assemble Rear Panel ................................................... 5-6 To Disassemble and Assemble Metallic Shell................................................
RIGOL Figure Contents Figure 2-1 The Method of Adjusting Handle ....................................................... 2-3 Figure 2-2 Adjustable Positions for Handle ......................................................... 2-3 Figure 2-3 Sketch map of Front Panel ................................................................ 2-4 Figure 2-4 Sketch map of Rear Panel ................................................................. 2-5 Figure 2-5 Explanatory drawing of Double display interface ...........
RIGOL Table Contents Table Table Table Table Table Table Table Table Table Table Table Table XII 3-1 3-2 3-3 3-4 3-5 3-6 3-7 4-1 4-2 4-3 6-1 6-2 List of Recommended Equipments ...................................................... 3-2 Zero Offset of the Standard Equipment ............................................... 3-5 DC Gain Errors of the Standard Equipment ......................................... 3-6 AC Voltage Gain Error of the standard equipment ................................
Chapter 1 Performance & Specifications RIGOL Chapter 1 Performance & Specifications This chapter covers the following topics: General Technical Specifications Electric Technical Specifications DC Characteristics AC Characteristics Frequency/Period Characteristics Capacitance Characteristics Other Measurement Characteristics Application of Analog Filter Service Guide for DM3058 1-1
RIGOL Chapter 1 Performance & Specifications General Technical Specifications Power Supply: AC 100 V ~ 120 V, 45 Hz ~ 440 Hz AC 200 V ~ 240 V, 45 Hz ~ 66 Hz Power consumption: 20 VA peak value Dimension: Height: 107.0 mm Width: 231.6 mm Depth: 290.5 mm Weight: 2.5 kg Display: 256× 64 LCD display, support dual display, menu, Chinese and English languages, operation help. Working Environment: Precision 0 ℃ ~ 50 ℃, 80% R.H., 40 ℃, No coagulation. Storage Temperate: -20 ℃ ~ 70 ℃.
Chapter 1 Performance & Specifications RIGOL Electric Technical Specifications DC Characteristics Accuracy Specifications (% of reading + % of range)[1] Function DC Voltage DC Current[8] Resistance[3] [2] Range 200.000 2.00000 20.0000 200.000 1000.00 200.000 2.00000 20.0000 200.000 2.00000 10.0000 200.000 2.00000 20.0000 200.000 2.00000 10.0000 100.000 mV V V V V[4] μA mA mA mA A A[5] Ω kΩ kΩ kΩ MΩ MΩ MΩ [6] Test Current or Burden Voltage <8 mV <80 mV <0.05 V <0.5 V <0.1 V <0.
Chapter 1 Performance & Specifications RIGOL Input Protection: 1000 V on all ranges CMRR (common mode rejection ratio): 120 dB (For the 1 kΩ unbalanced resistance in LO lead, maximum ± 500 VDC). NMRR (normal mode rejection ratio): 80 dB at “slow” measurement rate Add 20 dB when open the “AC filter” (Settling time adds 0.35 s (source impedance near zero)) Resistance Measurement Method: Open-circuit Voltage: Max.
Chapter 1 Performance & Specifications RIGOL AC Characteristics Function True RMS AC Voltage[[3] True RMS AC Current[5] Accuracy Specifications (% of reading + % of range)[1] Temperature Frequency 1 Year Coefficient [2] Range Range 23 ℃ ± 5 ℃ 0 ℃ to 18 ℃ 28 ℃ to 55 ℃ 200.000 mV 20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005 45 Hz – 20 kHz 0.2 + 0.05 0.01 + 0.005 20 kHz – 50 kHz 1.0 + 0.05 0.01 + 0.005 50 kHz – 100 kHz 3.0 + 0.05 0.05 + 0.010 2.00000 V 20 Hz – 4 5Hz 1.5 + 0.10 0.01 + 0.005 45 Hz – 20 kHz 0.
RIGOL [4] [5] [6] Chapter 1 Performance & Specifications Specifications are for sine wave input >5% of range. Add 0.1% of the range for the sine wave input is 1% ~ 5% of the range. 30 seconds OFF after 30 seconds ON is recommend for the continuous current higher than DC 7 A or AC RMS 7 A. For frequency<100 Hz. True RMS AC Voltage Measurement Method: AC coupled true RMS measure - up to 1000 V DC bias are permitted on every range.
Chapter 1 Performance & Specifications RIGOL Frequency/Period Characteristics Function Range Frequency Period 200 mV to 750 V[2] 20 mA to 10 A[3] Frequency Range 20 Hz – 2 kHz 2 kHz – 20 kHz 20 kHz – 200 kHz 200 kHz – 1 MHz 20 Hz – 2 kHz 2 kHz – 10 kHz Accuracy Specifications (% of reading)[1] Temperature Coefficient 1 Year 0 ℃ to 18 ℃ 23 ℃ ± 5 ℃ 28 ℃ to 55 ℃ 0.01 + 0.003 0.002 + 0.001 0.01 + 0.003 0.002 + 0.001 0.01 + 0.003 0.002 + 0.001 0.01 + 0.006 0.002 + 0.002 0.01 + 0.003 0.002 + 0.001 0.
Chapter 1 Performance & Specifications RIGOL Capacitance Characteristics Function Range[2] Capacitance 2.000 nF 20.00 nF 200.0 nF 2.000 μF 200 μF 10000 μF Accuracy Specifications (% of reading + % of range)[1,2] Temperature 1 Year Coefficient Test Current 23 ℃ ± 5 ℃ 0 ℃ to 18 ℃ 28 ℃ to 55 ℃ 200 nA 3 + 1.0 0.08 + 0.002 200 nA 1 + 0.5 0.02 + 0.001 2 μA 1 + 0.5 0.02 + 0.001 10 μA 1 + 0.5 0.02 + 0.001 100 μA 1 + 0.5 0.02 + 0.001 1 mA 2 + 0.5 0.02 + 0.001 Remarks: [1] Specifications are for 0.
Chapter 1 Performance & Specifications RIGOL Other Measurement Characteristics Triggering and Memory Samples per Trigger: Trigger Delay: External trigger input: Input Level: Trigger Condition: Input Impendence: Min Pulse Width: VMC Output: Electric Level: Output polarity: Output impedance: 1 to 2,000 8 ms to 2000 ms TTL compatible (High level when left trigger input open) Selectable Rising, Falling, Low-level, High-level.
Chapter 1 Performance & Specifications RIGOL Application of Analog Filter The analog filter of DM3058 can be used to reduce the influence to measured results from AC component in DC mode. For the most measures, the filter may not require, but sometimes it can improve the DC measurement. For instance, if the DC electrical source to be measured has a big AC ripple, it can be reduced by the analog filter.
Chapter 2 Quick Start RIGOL Chapter 2 Quick Start The chapter mainly covers the following topics: General Inspection Handle Adjustment Front Panel Rear Panel User Interface Inspect the Instrument Check the List of Accessories Double Display Single Display Service Guide for DM3058 2-1
RIGOL Chapter 2 Quick Start General Inspection Inspect the Instrument Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. In case of any damage, or defect, or failure, notify the RIGOL Sales Representative. If the shipping container is damaged, or the protective material shows signs of stress, notify the carrier as well as your RIGOL sales office.
Chapter 2 Quick Start RIGOL Handle Adjustment To adjust the handle position of DM3058 Digital Multimeter, please grip the handle by the sides and pull it outward. Then, rotate the handle to the desired position as shown in figure 2-1, figure 2-2.
Chapter 2 Quick Start RIGOL Front Panel Yellow District: Trigger Control Keys LCD Display USB Host On/Off Switch Purple District: Measurement Function Keys Directive keys Blue District: Assistant Function Signal Input Terminal Keys Menu Operation Keys Figure 2-3 Sketch map of Front Panel Gray Function Keys: Measure DC Voltage Measure DC Current Measure Resistance Measure Continuity Measure Frequency Preset Mode Measure AC Voltage Measure AC Current Measure Capacitance Measure Diode Measure Sensor Se
Chapter 2 Quick Start RIGOL Rear Panel 10/100 Ethernet Current Input Fuse P FU U S H GPIB (IEEE--488) USB Device Power Socket 40 S E Fuse: AC 250V F300mA RS232 RXD RS232 TXD RS232 GND LXI Class C Non Auto-MDIX SEL Line (20W Max) 115 AC 100-120V 45-440Hz 230 AC 200-240V 45-66Hz 1 2 3 4 5 6 7 8 9 EXT TRIG VM COMP +5 V OUT USB Device RS-232 Power Fuse AC Voltage Selector Figure 2-4 Sketch map of Rear Panel Service Guide for DM3058 2-5
Chapter 2 Quick Start RIGOL User Interface Double Display S Slow M Middle F Fast REF No icon REL Open REL Close P/F P/FOpen No icon P/FClose Main Display dB dB dBm dBm Stat. Open Max. Max. Min. Min Avg.
Chapter 3 Performance Test RIGOL Chapter 3 Performance Test The chapter mainly covers the following topics: Calibration Interval Calibration Instructions Testing Equipments Performance Verification Test Input Connections Testing Conditions Verification Testing Zero Offset Test Gain Test Service Guide for DM3058 3-1
Chapter 3 Performance Test RIGOL Calibration Interval In order to keep the accuracy while measuring, the meter must be calibrated periodically, and the calibration interval depends on the requirements for accuracy of measurement, As for the applications needs higher accuracy, 90 day‟s intervals are recommended; besides, 1 year‟s interval is acceptable in most cases. No matter in what kind of applications, RIGOL does not recommend the calibration interval is more than 1 year.
Chapter 3 Performance Test RIGOL Performance Verification Test The test is used to verify the measuring performance of the instrument. The specifications listed in user’s manual of DM3058 are used while verifying through two kinds of testing model: Quick test or Common test. Quick Test Quick test is an eclectic project which can either speed up test or ensure a high confidence level while testing.
RIGOL Chapter 3 Performance Test Input Connections 4-teriinal shortcircuiter which is made of copper or copper alloy with low thermoemf must be used for zero offset test. Coaxial-cable is adopted when calibrating capacitance whose shielding layer is connected with LO terminal.
Chapter 3 Performance Test RIGOL Verification Testing Zero Offset Test This test is used to inspect the performance of zero offset of the device, and which is required only when regular offsets happened to function or range. Verification procedure: 1. 2. 3. 4. Read “Testing Conditions” carefully before start. Short-circuit the Input HI-LO and Sense HI-LO using 4-terminal shortcircuiter and open circuit of input terminal of current, see as Figure 3-1.
Chapter 3 Performance Test RIGOL Zero Offset of the Standard Equipment (Continued) Function [1] Resistance[2] Range 200.000 2.00000 20.0000 200.000 2.00000 10.0000 100.000 Ω kΩ kΩ kΩ MΩ MΩ MΩ Input signal Quick Test short short short short short short short Q Q Test limits (1 year) Upper Lower limit limit 10 mΩ -10 mΩ 60 mΩ -60 mΩ 600 mΩ -600 mΩ 6Ω -6 Ω 80 Ω -80Ω 300 Ω -300 Ω 4 kΩ -4 kΩ Remarks: [1] The rate of reading is “Slow”.
Chapter 3 Performance Test RIGOL DC Gain Errors of the Standard Equipment (Continued) Test limits (1 year) Input Quick Function[1] Range Upper limit signal Test Lower limit DC Voltage 200.000 V -200 V -199.964 V -200.036 V 1000.00 V 1000 V 1000.19 V 999.81 V 1000.00 V -1000 V -999.81 V -1000.19 V DC Current 200.000 μA 200 μA Q 200.120 μA 199.880 μA 200.000 μA -200 μA Q -199.880 μA -200.120 μA 2.00000 mA 2 mA 2.00120 mA 1.99880 mA 2.00000 mA -2 mA -1.99880 mA -2.00120 mA 20.0000 mA 20 mA Q 20.0230 mA 19.
Chapter 3 Performance Test RIGOL Table 3-4 AC Voltage Gain Error of the standard equipment Function[1] 200.000 mV 2.00000 V 20.0000 V 200.000 V 750.00 V Range Input signal Quick Test 10 mV 200 mV 200 mV 200 mV 200 mV 200 mV 100 mV 2V 2V 2V 2V 2V 1V 20 V 20 V 20 V 20 V 20 V 10 V 200 V 200 V 200 V 200 V 200 V 37.
Chapter 3 Performance Test RIGOL (3) Test the ranges listed in Table 3-5 item by item. And then set the rate of reading as “Slow” and disable all the math operations. (4) Input signals listed in Table 3-5 using calibration instrument and compare the test results with test limits listed. (Make sure that the calibration instrument is in a state of output) Table 3-5 AC Current Gain Error of the Standard Equipment Function[1] Range 20.
Chapter 3 Performance Test RIGOL Table 3-6 Frequency Gain Error of the Standard Equipment Virtual value 200 mV 200 mV Frequency 20 Hz 1 MHz Input signal Quick Test 200 mV 200 mV Q Test limits (1 year) Upper limit 20.0620Hz 1.00016MHz Upper limit 19.9380Hz 0.99984MHz Remarks: [1] Q denotes optional quick test point. 5. Verification procedure of Capacitance Gain: (1) Read “Testing Conditions” carefully before start. (2) Switch to Capacitance measure and select the ranges listed in Table 3-7.
Chapter 4 Calibration RIGOL Chapter 4 Calibration The chapter mainly covers the following topics: Calibration Protection Calibration Notice DC Voltage, DC Current, Resistance Calibration AC Voltage and AC Current Calibration Frequency Calibration Capacitance Calibration Service Guide for DM3058 4-1
RIGOL Chapter 4 Calibration Calibration Protection The protection is used to avoid unexpected or non-authoritative calibration. DM3058 is in a state of calibration protection before delivery. Before calibrating, correct security code must be input to release protection. The calibration code could be made of 10 characters at most which is from number “0~9” and letter “A~Z”. To modify the code, correct former code must be input at first.
Chapter 4 Calibration RIGOL default . Select On/Off as “On” and the instrument will amend the measuring result according to user calibration parameters. Conversely, factory parameters will be rested on. DC Voltage, DC Current, Resistance Calibration The calibration procedures of DC voltage, DC current and resistance are all divided into two steps: Zero calibration and Gain calibration. Zero calibration must be done firstly while you calibrate.
Chapter 4 Calibration RIGOL 6. After finishing above steps, press Save to store the parameters have been calibrated and exit. NOTE You must be very cautious about calibrating in order to avoid detrimental effection for the performance of instrument owing to misoperation. Before calibrate, please read “Testing Conditions” in Chapter3. Table 4-1 DC Zero and DC Gain Calibration Input Value Function DC Voltage DC Current Resistance Range 200.000 2.00000 20.0000 200.000 1000.00 200.000 2.00000 20.
Chapter 4 Calibration RIGOL 1. Read “Testing Equipments” and “Testing Conditions”. 2. Press and enable AC voltage measure, and then separately set the range as 20kΩ and the reading resolution as “Slow”. 3. Press , and select T/C PSW to input password and set SecrOff , then select Cal Enter ; 4. Connect the voltage output of calibration instrument with HI-LO of the meter. Then, set the output of calibration instrument as Sine, 100 mV, 1 KHz.
Chapter 4 Calibration RIGOL Then, set the output of calibration instrument as Sine, 2 V, 100 kHz. Next, enable the output and press Gain to start calibrating until it is stable. The results will display in the table of UnCal G . 5. After finishing above steps, select Save to store the parameters have been calibrated and exit. Capacitance Calibration As for capacitance, both Zero and Gin calibration are required to be executed. Take Capacitance in 2 nF as reference to show you the procedure: 1.
Chapter 5 Disassembly & Assembly RIGOL Chapter 5 Disassembly & Assembly The chapter mainly covers the following topics: The Disassembly and Assembly Notice The 3D View of DM3058 To Disassemble and Assemble Handle To Disassemble and Assemble Rear Panel To Disassemble and Assemble Metallic Shell To Disassemble and Assemble Fuse Socket To Disassemble and Assemble GPIB PCB To Disassemble and Assemble Filter Board & To Disassemble and Assemble Front Panel & LCD To Disasse
Chapter 5 Disassembly & Assembly RIGOL The Disassembly and Assembly Notices Notices: Please don‟t disassemble the instrument except the work need Please don‟t disassemble the instrument except the professionals Please cut off the multimeter power before disassembly Please wear the anti-static hand-ring or take other anti-static measures when disassembling Please use proper tools and disassemble in order Prevent metallic parts from transfiguration and avoid being scratched when disassembling
Chapter 5 Disassembly & Assembly RIGOL The 3D View of DM3058 In order to help you have a primary understanding of the structure of DM3058, the following parts show you its interior and exterior skeleton drawing. Besides, the disassembly and assembly procedures are also provided to reference.
Chapter 5 Disassembly & Assembly RIGOL Figure 5-2 The Interior 3D View of DM3058 Please acquaint the main parts before starting. In progress of disassembling or assembling, operate gently and step by step. Note that please do not scratch the surface of the device and damage PCB and so on, for the details please refer to Disassembly and Assembly Notices.
Chapter 5 Disassembly & Assembly RIGOL To Disassemble and Assemble Handle Grip the handle in both sides and pull it outward, then rotate it to the desired position to take it away. For the position of the handle please refer to following figure.
RIGOL Chapter 5 Disassembly & Assembly To Disassemble and Assemble Rear Panel Figure 5-4 The Schematic of Disassembling and Assembling Rear Panel 5-6 Service Guide for DM3058
Chapter 5 Disassembly & Assembly RIGOL To Disassemble and Assemble Metallic Shell Figure 5-5 The Schematic of Disassembling and Assembling Metallic Shell Service Guide for DM3058 5-7
RIGOL Chapter 5 Disassembly & Assembly To Disassemble and Assemble Fuse Socket Figure 5-6 The Schematic of Disassembling and Assembling Fuse Socket 5-8 Service Guide for DM3058
Chapter 5 Disassembly & Assembly RIGOL To Disassemble and Assemble GPIB PCB Figure 5-7 The Schematic of Disassembling and Assembling GPIB PCB Service Guide for DM3058 5-9
RIGOL Chapter 5 Disassembly & Assembly To Disassemble and Assemble Filter Board & Transformer Figure 5-8 The Schematic of Disassembling and Assembling Filter Board and Transformer 5-10 Service Guide for DM3058
Chapter 5 Disassembly & Assembly RIGOL To Disassemble and Assemble Front Panel & LCD Figure 5-9 The Schematic of Disassembling and Assembling Front Panel and LCD Service Guide for DM3058 5-11
RIGOL Chapter 5 Disassembly & Assembly To Disassemble and Assemble Motherboard Figure 5-10 The Schematic of Disassembling and Assembling Motherboard 5-12 Service Guide for DM3058
Chapter 5 Disassembly & Assembly RIGOL To Disassemble and Assemble Key Board PCB Figure 5-11 The Schematic of Disassembling and Assembling Key Board PCB We amicably suggest you obey above orders or methods as possible as you can while disassemble or assemble in order to avoid damages to equipment and save your time.
Chapter 6 Troubleshooting & Maintenance RIGOL Chapter 6 Troubleshooting & Maintenance The chapter mainly covers the following topics: DM3058 Principle Troubleshooting Maintenance and Cleaning Common Troubleshooting Components Inspection Replaceable Part list Service Guide for DM3058 6-1
Chapter 6 Troubleshooting & Maintenance RIGOL DM3058 Principle Summary The circuit of DM3058 can be divided into floating circuit, grounding circuits and power input. 1. The floating circuit includes current fuse, signal input end of front panel (banana socket) and measurement analog front-end which contains all the circuits of measurement function such as input protection, function switch, signal transform conditioning, A/D conversion and control interface.
Chapter 6 Troubleshooting & Maintenance RIGOL 2. Ground circuit contains most circuit such as keyboard, LCD, USB Host, BNC board, LAN/GPIB board and the host CPU. 3. The AC supply inpour the meter via power socket and connect to transformer through filter board, from which, it will be stepped down or divided into multi low voltage and then respectively supply for floating circuit and grounding circuits.
Chapter 6 Troubleshooting & Maintenance RIGOL Troubleshooting Common Troubleshooting 1. After press the power switch, the multimeter is blank screen with nothing display: (1) Check if the power is correctly connected. (2) Check if the power switch on Rear panel has been turned on. (3) Check if the power light on Front panel is bright. (4) Check if the safety fuse is blown; replace a new one if necessary. (5) Check if the power selector is in correct position.
Chapter 6 Troubleshooting & Maintenance ! RIGOL CAUTION Make sure the power is cut off before replacing fuse, and the specifications of fuse should be in accordance with the requirements in this manual. 2. When the reading of small capacitance is not stable or the offset is a little bigger: (1) Shielded cable must be used to connect the capacitance while measuring small capacitance so as to reduce the influence to result from exterior interference.
Chapter 6 Troubleshooting & Maintenance RIGOL 8. When fail to connect with the meter via LAN interface: (1) Check whether the type of netting twine is correct and still connected. (2) Check whether the indicator light of network connections of the meter is in normal state; if not, that denotes there may be something wrong about LAN. (3) Check whether the address of both DHCP and IP and the settings of DNS are in accordance with the configuration of all LANs.
Chapter 6 Troubleshooting & Maintenance RIGOL Components Inspection From this part, you can get more information about malfunction in order to process advanced troubleshooting by yourself. 1.
Chapter 6 Troubleshooting & Maintenance RIGOL Red Blue AC 8.5 V Blue Brown AC 110 V AC 17 V Yellow Yellow AC 17 V White Black AC 110 V AC 9 V White Black Figure 6-3 Transformer Sketch Map Compare the measured voltages of each winding with the ones in Table 6-1, we can see that: Table 6-1 Check Table of Voltages among Windings of Transformer windings Original side Vice side White-Yellow White-Red Black-Black White-Yellow Yellow-Brown Blue-Blue Max(Vrms) Min(Vrms) 132 265 10.8 21 21 10.
Chapter 6 Troubleshooting & Maintenance RIGOL Inspection method: Press the second menu operation key in the left side, and turn on the meter at the same time. Keep pressing for about 5 seconds until you hear beep. After that, multimeter starts testing display and a tip message on the screen “Press ·Help· Key to Switch, Hold ·Help· Key to Exit” will be shown. Press Help button, the screen would switch between Full white (all pixel are bright) and Full black (all pixel are put out).
Chapter 6 Troubleshooting & Maintenance RIGOL Press Utility System Setup Factory . It will be difficult to read the information on the interface if the display configuration is abnormal; here we can recall the factory setup by another way to solve this problem: Keep pressing the second menu operation key and turn on the multimeter. Keep pressing for about 5 seconds until you hear beep. After that, the multimeter will load the factory configuration automatically.
Chapter 6 Troubleshooting & Maintenance RIGOL Replaceable Part list RIGOL provides some replaceable parts for you to maintain or update. Please see following table.
Chapter 6 Troubleshooting & Maintenance RIGOL Table 6-2 Replaceable Part List Part No.
Chapter 6 Troubleshooting & Maintenance RIGOL For the above parts, please contact RIGOL technical support department or RIGOL authorized distributors to order and refer to “Contact Us” to get the contact information.
Chapter 6 Troubleshooting & Maintenance RIGOL Maintenance and Cleaning System Maintenance In order to ensure the running performance of instrument and extend useful time, please abide by the following advices: 1. Before operating please ensure that all the performance and particular using method. Have been taken into you memory and brain. For any puzzles please refer to Chapter 2. 2.
Chapter 7 Service & Support RIGOL Chapter 7 Service & Support The chapter mainly covers the following topics: Warranty Contact Us Service Guide for DM3058 7-1
RIGOL Chapter 7 Service & Support Warranty RIGOL warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment from an authorized RIGOL distributor. If a product proves defective within the respective period, RIGOL will provide repair or replacement as described in the complete warranty statement.
Chapter 7 Service & Support RIGOL Contact Us If you have any problem or requirement during using our products, please contact RIGOL Technologies, Inc. or the local distributors. Domestic: Please call Tel: (86-10) 8070 6688 Fax: (86-10) 8070 5070 Service & Support Hotline: 800 810 0002 9:00 am –5: 00 pm from Monday to Friday Or by e-mail: Service@rigol.com Or mail to: RIGOL Technologies, Inc.
Index RIGOL Index AC Characteristics ........................ 1-5 AC Current Calibration .................. 4-5 AC Current Gain Verification ......... 3-8 AC Voltage Calibration ........... 4-1, 4-5 AC Voltage Gain Verification .......... 3-7 Analog Filter .............................. 1-10 Arbitrary sensor measurement ...... 1-9 Calibration Instructions................. 3-2 Calibration Interval....................... 3-2 Calibration Notices ....................... 4-3 Calibration Protection ............
Chapter 7 Service & Support RIGOL Testing Conditions ........................ 3-4 Testing Equipments ...................... 3-2 The Disassembly and Assembly Notices ........................................ 5-2 The Exterior 3D View .................... 5-3 The Interior 3D View .................... 5-4 Triggering and Memory ................. 1-9 2 True RMS AC Current ................... 1-6 True RMS AC Voltage ................... 1-6 User Interface ............................. 2-6 Warm-up Time ..........