Specifications

RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-8
Syntax
:MEASure?
Function
Queries whether the current measurement has been completed. If
completed, the query returns TRUE, or else returns FALSE.
Syntax
:MEASure {AUTO|MANU}
Function
Sets the measurement mode to Auto or Manual.
Default
AUTO
Syntax
:MEASure:VOLTage:DC?
Function
The query returns the current DC voltage in the form of scientific
notation such as +2.53021747E-04, the unit is V.
Syntax
:MEASure:VOLTage:DC {0|1|2|3|4|MIN|MAX|DEF}
Function
Sets the DC voltage measurement range.
Explanation
The measurement mode will change to Manual while you set
the range.
Different parameters have different ranges:
Parameter
Range
0
200 mV
1
2 V
2
20 V
3
200 V
4
1000 V
MIN
200 mV
MAX
1000 V
DEF
20 V
Example
Setting the range to minimum:
:MEASure:VOLTage:DC 0 or
:MEASure:VOLTage:DC MIN
Syntax
:MEASure:VOLTage:DC:RANGe?
Function
Queries the current DC voltage range.