Specifications

RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-18
Syntax
:MEASure:PERiod {0|1|2|3|4|MIN|MAX|DEF}
Function
Sets the period measurement range.
Explanation
For meanings in each range, please refer to the
Explanation in :MEASure:VOLTage:AC.
The period measurement ranges from 3.3 us to 0.33 s.
The DEF is 2.
Example
Setting the period measurement range to maximum:
:MEASure:PERiod 4 or
:MEASure:PERiod MAX
Syntax
:MEASure:PERiod:RANGe?
Function
Queries the AC voltage range currently used by period
measurements.
The query returns 0, 1, 2, 3 or 4.
Syntax
:MEASure:PERiod:DIGIt?
:MEASure:PERiod:DIGIt {INC|DEC|5|6|7}
Function
Sets the display digit for the period measurement values.
The query returns 5, 6 or 7.
Explanation
Refer to the Explanation in :MEASure:VOLTage:DC:DIGIt.
Example
Setting the display digit to 7:
:MEASure:PERiod:DIGIt 7
Decreasing the display digit one bit:
:MEASure:PERiod:DIGIt DEC
Syntax
:MEASure:CONTinuity?
Function
The query returns the resistance that connected to the meter under
the continuity measurement in the form of scientific notation such
as +8.888000e+03, the unit is Ω.
Syntax
:MEASure:CONTinuity {
<value>
|MIN|MAX|DEF}