Specifications
Chapter 2 Command System RIGOL
Programming Guide for DM3000
2-23
The query returns 0, 1 or 2.
Explanation
The AC voltage measurement function must be enabled before
using this command.
The “DEF” is 1.
Each parameter has its own reading resolution:
Range
Reading resolution
0
3 ½ digits
1
4 ½ digits
2
5 ½ digits
MAX
5 ½ digits
MIN
3 ½ digits
DEF
4 ½ digits
Example
Setting the reading resolution of AC voltage measurement to 5 ½ :
:RESOlution:VOLTage:AC 2
4. :RESOlution:CURRent:DC
Syntax
:RESOlution:CURRent:DC?
:RESOlution:CURRent:DC {0|1|2|MIN|MAX|DEF}
Function
Sets the reading resolution for DC current measurements.
The query returns 0, 1 or 2.
Explanation
The DC current measurement function must be enabled before
using this command.
For the reading resolution of each parameter, please refer to
the “Explanation” in :RESOlution:VOLTage:DC.
The “DEF” is 1.
Example
Setting the reading resolution of DC current measurement to 5 ½ :
:RESOlution:CURRent:DC 1
5. :RESOlution:CURRent:AC
Syntax
:RESOlution:CURRent:AC?
:RESOlution:CURRent:AC {0|1|2|MIN|MAX|DEF}
Function
Sets the reading resolution for AC voltage measurements.
The query returns 0, 1 or 2.
Explanation
The AC voltage measurement function must be enabled before
using this command.
For the reading resolution of each parameter, please refer to
“Explanation” in :RESOlution:VOLTage:AC.