Specifications

RIGOL Chapter 2 Command System
Programming Guide for DM3000
2-48
3
1/m
1/MIN
4
1/10 s
1/10 SEC
5
1/s
1/SEC
6
10/s
10/SEC
7
50/s
50/SEC
8
100/s
100/SEC
9
833/s
833/SEC
10
1 k/s
1000/SEC
11
5 k/s
5000/SEC
12
10 k/s
10000/SEC
13
50 k/s
50000/SEC
Syntax
:DATAlog:RUN
Function
Executes the configured data acquisitions.
Syntax
:DATAlog:RUN?
Function
Queries if the meter is running under the data acquisition.
The query returns RUN or STOP.
Syntax
:DATAlog:STOP
Function
Stops the data acquisition.
Syntax
:DATAlog:DATA?
<value1>,<value2>
Function
The query returns the acquired data with a specified number and
quantity in the memory.
Explanation
The acquired data are saved from number 1 after the start of
acquisition.
<value1>
defines the start number of returned data.
<value2>
defines the data quantity (within 1 and 100) to be
returned.
Example
The query returns three data from number 2:
:DATAlog:DATA? 2,3
The query returns