Specifications

Chapter 3 Commands Compatibility RIGOL
Programming Guide for DM3000
3-9
[,{<resolution>|MIN|MAX|DEF}]]
configures the meter for DC current measurements
and immediately triggers a measurement. The
results are sent directly to the instrument output
buffer.
[SENSe:]MEASure:DIODe?
Resets all diode measurement parameters and
trigger parameters to their defaults, and then
configures the meter for diode measurements and
immediately triggers a measurement. The results
are sent directly to the instrument output buffer.
[SENSe:]MEASure:FREQuency?
[{<range>|MIN|MAX|DEF}
[,{<resolution>|MIN|MAX|DEF}]]
Resets all frequency measurement parameters and
trigger parameters to their defaults, and then
configures the meter for frequency measurements
and immediately triggers a measurement. The
results are sent directly to the instrument output
buffer.
[SENSe:]MEASure:FRESistance?
[{<range>|AUTO|MIN|MAX|DEF}
[,{<resolution>|MIN|MAX|DEF}]]
Resets all 4-wire resistance measurement
parameters and trigger parameters to their defaults,
and then configures the meter for 4-wire resistance
measurements and immediately triggers a
measurement. The results are sent directly to the
instrument output buffer.
[SENSe:]MEASure:PERiod?
[{<range>|MIN|MAX|DEF}
[,{<resolution>|MIN|MAX|DEF}]]
Resets all period measurement parameters and
trigger parameters to their defaults, and then
configures the meter for period measurements and
immediately triggers a measurement. The results
are sent directly to the instrument output buffer.
[SENSe:]MEASure:RESistance?
[{<range>|AUTO|MIN|MAX|DEF}
[,{<resolution>|MIN|MAX|DEF}]]
Resets all 2-wire resistance measurement
parameters and trigger parameters to their defaults,
and then configures the meter for 2-wire resistance
measurements and immediately triggers a
measurement. The results are sent directly to the
instrument output buffer.
[SENSe:]MEASure[:VOLTage]:AC?
[{<range>|AUTO|MIN|MAX|DEF}
[,{<resolution>|MIN|MAX|DEF}]]
Resets all AC voltage measurement parameters and
trigger parameters to their defaults, and then
configures the meter for AC voltage measurements
and immediately triggers a measurement. The
results are sent directly to the instrument output