RIGOL User’s Guide Publication number DS1-071026 October 2007 DS1000 Series Digital Oscilloscopes DS1000CD, DS1000C, DS1000MD, DS1000M © Copyright RIGOL Technologies, Inc.
RIGOL z z z z Copyright © RIGOL TECHNOLOGIES, INC. 2006 All Rights Reserved. RIGOL products are protected by patent law in and outside of P.R. China. Information in this publication replaces that in all previously corresponding material. RIGOL Technologies, Inc. reserves the right to modify or change pat of or all the specifications and pricing policies at company’s sole decision. NOTE: RIGOL is registered trademark of RIGOL TECHNOLOGIES, INC. © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Safety Notices Review the following safety precautions carefully before operate the instrument to avoid any personal injury or to damage the instrument and any products connected to it. To avoid potential hazards use the instrument in a manner only as specified by this user’s guide. The instrument should be serviced by qualified personnel only. To Avoid Fire or Personal Injury Use proper power cord Use only the power cord designed for your oscilloscope and authorized in your country.
RIGOL Use Proper Fuse. Use only the fuse type and rating specified for this product. Avoid Circuit or Wire Exposure. Do not touch exposed connections and components when power is on. Do not operate with suspected failures. If you suspect damage with this product, have it inspected by qualified service personnel before further operations. Provide Proper Ventilation. Refer to the manual’s installation instructions for details as to the oscilloscope has proper ventilation.
RIGOL Safety Terms and Symbols Terms in This Guide. These terms may appear in this guide: ! WARNING: Warning statements identify conditions or practices that could result in injury or loss of life. ! CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property. Terms on the Product: These terms may appear on the product: DANGER indicates an injury hazard may be immediately accessible.
RIGOL General-Purpose Oscilloscopes RIGOL DS1000-Series Digital Oscilloscopes offer exceptional waveform viewing and measurements in a compact, lightweight package. The DS1000 series are ideal for production test, field service, research and design and all of the applications involving analog/digital circuits test and troubleshooting, as well as education and training.
RIGOL z z Multiple Language User Interface Pop-up menu makes it easy to read and easy to use z z Built-in Chinese and English help system Easy-to-use file system supports Chinese & English characters key-in DS1000 Series Digital Oscilloscope accessories: z z z z z z VI Two 1.
RIGOL Content Safety Notices ..................................................................................................II General-Purpose Oscilloscopes .......................................................................... V CHAPTER 1: GETTING STARTED ..................................................................... 1-1 The Front Panel and the User Interface ....................................................... 1-2 To Inspect the Instrument ................................................
RIGOL Example 9: Pass/Fail Test .......................................................................... 3-13 Example 10: Triggering on a Digital Signal…………………………………………….….3-14 CHAPTER 4: PROMPT MESSAGES & TROUBLESHOOTING .................................4-1 Prompting Message ....................................................................................4-1 Troubleshooting..........................................................................................4-3 CHAPTER 5: SUPPORT & SERVICE ......
RIGOL Chapter 1 : Getting Started This chapter covers the following topics: The front panel and user interface To inspect the instrument To perform a functional check To compensate probes To use digital leads (Mixed signal oscilloscope only) To display a signal automatically To set up the vertical system To set up the horizontal system To trigger the oscilloscope © Copyright RIGOL Technologies, Inc. 2007.
RIGOL The Front Panel and the User’s Interface One of the first things you will want to do with your new oscilloscope is to become acquainted with its front panel. This chapter will help you be familiar with the layout of the knobs and keys and how to use them. Read the chapter carefully before further operations. The front panel has knobs and buttons. The knobs are used most often and are similar to the knobs on other oscilloscopes.
RIGOL Vertical buttons: Associated with CH1, CH2, MATH, REF and LA menus, the OFF button can set waveform or menu which currently active off. Horizontal buttons: Associated with horizontal MENU. Trigger buttons: Associated with trigger MENU, instant action to set 50% trigger level and FORCE trigger. Action buttons: Include run control buttons for AUTO and RUN/STOP.
RIGOL Figure 1-2 Front Panel Controls Character definitions in this User’s Guide: Throughout this book, character figures of buttons and knobs are the same to those on front-panel. A box around the name of the key denotes MENU function buttons on front-panel, such as Measure. ( ) denotes the multi-function knob . denotes the two POSITION knobs. denotes the two SCALE knobs. denotes the LEVEL knob. The name with a drop shadow denotes the menu operating key, such as WAVEFORM soft key in STORAGE menu.
RIGOL Running status Location of waveform window Trigger point in in memory memory Trigger point in waveform window Channel 1 Menu Channel 2 Waveform display window Figure1-3 Display screen (Analog channels only) Running status Digital channel turned off Digital channels turned on Channel 1 Digital channels Channel coupling and vertical div. Horizontal time base div. Trigger offset Figure 1-4 User Interface (Analog and Digital channels) © Copyright RIGOL Technologies, Inc. 2007.
RIGOL To Inspect the Instrument After you get a new DS1000 series oscilloscope, please inspect the instrument according to the following steps: 1. Inspect the shipping container for damage. Keep a damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. 2. Check the accessories. Accessories supplied with the instrument are listed in "Accessories" in this guide.
RIGOL To Perform a Functional Check Perform this quick functional check to verify that your instrument is operating correctly. 1. Turn on the instrument. Use only power cords designed for your oscilloscope. Use a power source that delivers 100 to 240 VACRMS, 45Hz to 440Hz. Turn on the instruments, and wait until the display shows the waveform window. Push the Storage button, select Storage in the top menu box and push the Factory menu box.
RIGOL Probe compensator Figure 1-6 ② Set the probe attenuation to 10X. To do this, push CH1→Probe→10X. Probe scale Figure 1-7 Figure 1-8 ③ Push the AUTO button. Within a few seconds, you could see a square wave in the display (approximately 1 kHz 3 V peak- to- peak). ④ Push the OFF button or push the CH1 button again to turn off channel 1, push the CH2 button to turn on channel 2, repeat steps 2 and 3. 1-8 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL To Compensate Probes Perform this adjustment to match your probe to the input channel. This should be done whenever you attach a probe for the first time to any input channel. 1. Set the Probe menu attenuation to 10X. Set the switch to 10X on the probe and connect the probe to channel 1 on the oscilloscope. If you use the probe hook-tip, ensure a proper connection by firmly inserting the tip onto the probe.
RIGOL To Use Digital Leads (Mixed Signal Oscilloscope) 1. Switch off power supply of the Device Under Test if necessary to avoid short circuit. Since no voltage is applied to the leads at this step, you may keep the oscilloscope on. 2. Connect one end of the flat cable FC1868 to the Logic Analyzer Input, connect the other end to Logic Head LH1116. An identifier is located on each end of the flat cable, it can only be connected in one way.
RIGOL 3. Connect a test clip to one lead wire; make sure it’s connected well. Test clip Figure 1-11 4. Test your device with the clip. Figure 1-12 5. Remember to connect Ground Channel to the DUT’s ground terminal. GND Figure 1-13 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL To Display a Signal Automatically The oscilloscope has an Auto feature that sets up the oscilloscope automatically to display the input signal in a best fit. This Auto feature requires the input signal with a frequency of 50 Hz or higher and a duty cycle greater than 1%. When you press the AUTO button, the oscilloscope automatically sets up vertical, horizontal and Trigger controls to display the input signal in the screen for view.
RIGOL To Set Up the Vertical Window Figure 1-14 shows the Channels, MATH, REF, Logic Analyzer and OFF buttons and , knobs. The following exercise guides you vertical through the vertical buttons, knobs, and status bar. It will help you be familiar with the setting of the vertical parameters. Figure 1-14 1. Center the signal on the display with the knob. The knob moves the signal vertically, and it is calibrated.
RIGOL Measurement hints If the channel is DC coupled, you can quickly measure the DC component of the signal by simply noting its distance from the ground symbol. If the channel is AC coupled, the DC component of the signal is blocked, allowing you to use greater sensitivity to display the AC component of the signal.
RIGOL To Set Up the Horizontal System Figure 1-15 shows the MENU button, and knobs of horizontal system. The following exercise guides you through these buttons, knobs, and status bar. Figure 1-15 1. Turn the bar. knob and notice the change it makes to the status knob changes the sweep speed in a 1-2-5 step sequence, The horizontal and the value is displayed in the status bar. The time base ranges of the DS1000 series are listed as follows. The horizontal scan speed is from 5ns/div* to 50s/div.
RIGOL 2. The horizontal on waveform window knob moves displayed signal horizontally Horizontal offset back to 0 shortcut key Press the knob to set the horizontal offset to 0 as a shortcut key, this is especially helpful when the trigger point is far out of the screen and want it to get back to the screen center immediately. 3. Press the MENU key to display the TIME menu.
RIGOL To Trigger the Oscilloscope Figure 1-16 shows the trigger control area on the front panel, it has a trigger level knob and three soft buttons. The following exercise guides you through these trigger buttons, knobs, and status bar. Figure 1-16 1. Turn the trigger Level knob and notice the changes it makes to the display. knob or press the On the DS1000 series oscilloscopes, as you turn the 50% menu button, two things happen on the display for a short time.
RIGOL Trigger Level to 0 Shortcut key Turn the knob to change trigger level value and press the knob to set trigger level back to 0 as a shortcut key. 2. Change the trigger setup and notice these changes affect the status bar differently. · Press MENU button in the trigger control area. A soft button menu appears on the display showing the trigger setting choices. Figure 1-17 displays this trigger menu. · Press the trigger Mode button and choose Edge. · Press the trigger Source button to select CH1.
RIGOL 4. Press FORCE Press this button starts an acquisition regardless of an adequate trigger signal, usually used in “Normal” or ”Single” trigger mode. This button has no effect if the acquisition is already stopped. Key point: Holdoff: A time interval before the oscilloscope response to next trigger signal. During this holdoff period, the trigger system becomes “blind” to trigger signals. This function helps to view complex signals such as an AM waveform.
RIGOL Chapter 2 : Operating Your Oscilloscope By now you have got a brief understanding of DS1000 series with the VERTICAL, HORIZONTAL, and TRIGGER groups of the front-panel buttons. You should also know how to determine the setup of the oscilloscope by viewing the status bar. This chapter takes you through all groups of front-panel buttons, knobs, and menus. You will also further your knowledge of the operation hints by reading this guide.
RIGOL Understand the vertical system Ⅰ. Settings of the channels Each channel of DS1000 has an operation menu and it will pop up after pressing CH1 or CH2 button. The settings of all items in the menu are shown in the table below.
RIGOL Figure 2-2 Table2-2 Menu Settings 2/2 Comments Back to the previous menu page (The followings are the same, no more explanation) Selects Coarse Volts/Div Fine Invert ON OFF the resolution of the knob Defines a 1-2-5 sequence. To change the resolution to small steps between the coarse settings. Turn on the invert function. Restore to original display of the waveform. © Copyright RIGOL Technologies, Inc. 2007.
RIGOL 1. Channel coupling To use CH1 as an example, input a sine wave signal with DC shift Press CH1→Coupling→AC to set coupling as “AC” of CH1. In this setting, it blocks the DC component of the input signal. The waveform is displayed as Figure 2-3: AC coupling setup AC coupling status symbol Figure 2-3 Press CH1→Coupling→DC, to set coupling as “DC” of CH1. In this setting, it passes both AC and DC components of the input signal. 2-4 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL The waveform is displayed as Figure 2-4: DC coupling setup DC coupling status symbol Figure 2-4 Press CH1→Coupling→GND, to set coupling as “GND” of CH1. In this setting, it disconnects the input signal. The screen displays as Figure 2-5: GND coupling setup GND coupling status symbol Figure 2-5 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL 2. Set up the channel bandwidth limit To use CH1 as an example, input a signal that contains high frequency component. Press CH1→BW Limit→OFF, to set up bandwidth limit to “OFF” status. The oscilloscope is set to full bandwidth and the high frequency component in the signal will pass. The waveform is displayed as Figure 2-6: Turn off the BW limit Figure 2-6 Press CH1→BW Limit→ON, to set up bandwidth limit to “ON” status. It will reject the frequency component higher than 20MHz.
RIGOL 3. Probe Attenuation Setting When using a probe, the oscilloscope allows you to select the attenuation factor for the probe. The attenuation factor changes the vertical scaling of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip. To change (or check) the probe attenuation setting, press the CH1 or CH2 button (according to which channel you are using), toggle the Probe soft button to change the attenuation factor to match the probe you are using.
RIGOL 4. Volts/Div settings The Volts/Div control has Coarse or Fine configuration. The Vertical Sensitivity is 2mv-5V/div. Coarse: It is the default setting of Volts/Div, and it makes the vertical scaling in a 1-2-5-step sequence from 2mV/div, 5mV/div, 10mV, to 5 V/div. Fine: This setting changes the vertical scale to small steps between the coarse settings.
RIGOL 5. To invert a waveform Invert turns the displayed waveform 180 degrees, as respect to the ground level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted. Figure 2-10 and 2-11 show the changes after inversion. Invert OFF The waveform before inversion Figure 2-10 Invert ON The waveform after inversion Figure 2-11 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Digital Filter: Press CH1→Digital filter, display the digital filter menu. Turn ( and low limit of frequency. ) knob to set high Turn off digital filter Figure 2-12 Turn on digital filter Figure 2-13 Mark of digital filter 2-10 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Figure 2-14 Table 2-4 Menu Digital Filter Settings On Off Lower limit Turn on the digital filter Turn off the digital filter Setup Setup Setup Setup Filter Type Upper limit Comments as as as as LPF (Low Pass Filter) HPF(High Pass Filter) BPF (Band Pass Filter) BRF(Band Reject Filter) Turn ( ) knob to set high limit Turn ( ) knob to set low limit Back to higher level menu (The followings are the same, no more explanation) © Copyright RIGOL Technologies, Inc.
RIGOL Ⅱ. Math functions The mathematic functions include “add”, “subtract”, “multiply” and “FFT” for CH1 and CH2. The mathematic result can also be measured by grid and cursor.
RIGOL I. Using the FFT The FFT (Fast Fourier Transform) process mathematically converts a time-domain signal into its frequency components.
RIGOL Key points for FFT 1 Signals that have a DC component or offset can cause incorrect FFT waveform component magnitude values. To minimize the DC component, choose AC Coupling on the source signal. 2 To reduce random noise and aliases components in repetitive or single-shot events, set the oscilloscope acquisition mode to average. 3 To display FFT waveforms with a large dynamic range, use the dBVrms scale. The dBVrms scale displays component magnitudes using a log scale.
RIGOL Selecting an FFT Window DS1000 series oscilloscopes provide four FFT windows. Each window is a trade-off between frequency resolution and amplitude accuracy. What you want to measure and your source signals characteristics help determine which window to use. Use the following guidelines to select the best window. Table2-7 Window Rectangle Hanning Hamming Blackman Features Best for measuring Best frequency Resolution and worst magnitude resolution. This is essentially the same as no window.
RIGOL II. Using REF Reference Waveforms are saved waveforms to be selected for display. The reference function will be available after saving the selected waveform to non-volatile memory. Press REF button to display reference waveform menu.
RIGOL Import and Export Press REF→Imp./Exp. and go to the following menu. Figure 2-20 Table 2-10 Menu Settings Comments Explorer Path Directory File Switch to Path, directory or file Export Export the REF file from internal memory to export memory (see table 2-11) Import Import the REF file to internal memory Delete File Delete file The figure of import and export as following Figure 2-21 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Export Press REF→Imp./Exp.→Export and go to the following menu. Figure 2-22 Table 2-11 Menu Settings Comments Move the cursor up Move the cursor down To delete chosen letter Save Execute the operation The figure of export as following. Figure 2-23 2-18 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Save to External Memory Press REF→Save and go to the following menu. Figure 2-24 Table 2-12 Menu Explorer Settings Path Directory File Comments Switch among Path, Directory and File New File (Folder) Set up new file in Path and File. Set up new folder in directory. Delete File(Folder) Delete file(Folder) The figure of Save as following: Figure 2-25 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL New File (or New Folder) Press REF→Save→New File (or New Folder) and go to the following menu. Figure 2-26 Table 2-13 Menu Settings Comments Move the cursor up Move the cursor down To delete chosen letter Save Execute the operation The figure of key in as following File name key in Spell key in Character window Switch Chinese/English Switch Capital on/off Figure 2-27 2-20 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Import Press REF→Import and go to the following menu. Figure 2-28 Table 2-14 Menu Explorer Settings Path Directory File Import Comments Switch among Path, Directory and File Import the REF file into internal memory The figure of import as following. Figure 2-29 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Displaying a Reference Waveform. Figure 2-30 1. Push REF button to show the reference waveform menu. 2. Press soft button 1 to select CH1, CH2, MATH, FFT or LA (Mixed Signal Oscilloscope) to choose the REF channel you want. 3. Turn vertical and vertical waveform to a suitable position. 4. Select the save location of REF waveform by press soft button No. 2. 5. Save the waveform being displayed on the screen as REF by press soft button No.3.
RIGOL III. Set up LA channel (mixed signal oscilloscope) Single channel or group channels can be chosen ON or OFF, and also can set the size of waveform. Change display location of digital channel on screen and select threshold style. Press LA function button and go to the following menu.
RIGOL Select running channel Running channel Figure 2-32 2. Set threshold mode of digital channels Press LA→Threshold, select logic standard or User to define your own threshold voltage. The figure of menu shows as follow. Setup threshold Setup data of threshold by user Figure 2-33 2-24 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Threshold explanation LOGIC STANDARD THRESHOULD VLOTAGE TTL 1.4V CMOS 2.5V ECL -1.3V User -8V to +8V © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Set up Cannel Group Press LA→D7-D0 or D15-D8; turn on/off the channel single, or in a group. Also you can change the size of waveforms in 8 bits as a group.
RIGOL 1. Turn on or off a single logic channel Press LA→D7-D0→Channel, and choose the wanted channel by turning ( ) knob. Press No. 1 soft button or push down ( ) knob to turn on /off the channel. When the channel is on, we can see the mark ( ). When the channel is turned off, the mark will be shown as ( ). As figure 2-36 shows. On-off channels Channel list and status of on-off Figure 2-36 2.
RIGOL IV. Turn on/off channels The CH1, CH2, Ext. Trigger and LA (Mixed Signal Oscilloscope) channels on DS1000 series are input channels. All functionalities applied will be based on operating the instrument with channels. So MATH and REF can be regarded as relatively isolated channels. To turn on/off any one of the channels, press the corresponding button on the front panel. The key backlight indicates the channel is currently active. Press the button again to turn the channel off.
RIGOL V. Using Vertical and You can use the vertical controls to display waveforms, adjust vertical , and set input parameters. 1. Using vertical and knob. can change the position of waveforms in all channels The vertical (including MATH, REF and LA). The resolution of the knob changes according to the change of the vertical level (The digital channels of Mixed Signal Oscilloscope changes according to the changes of the waveforms displayed). Pressing this knob will clear the channel offset to zero.
RIGOL Understand the Horizontal System The oscilloscope shows the time per division in the scale readout. Since all active waveforms use the same time base, the oscilloscope only displays one value for all the active channels, except when you use Delayed Scan, or Alternative Trigger. The horizontal controls can change the horizontal scale and position of waveforms. The horizontal center of the screen is the time reference for waveforms.
RIGOL Figure 2-37 Table 2-19 Menu Delayed Time Base Settings Comments ON OFF Turn on Delayed Scan mode Turn off the Delayed Scan mode Y-T Show the relative relation between vertical voltage and horizontal time. Show CH1 value at X axis; CH2 value at Y axis. In Roll Mode, the waveform display updates from right to left. X-Y Roll Trig-offset Reset Adjust to the center ① ② ③ ④ ⑤ Figure 2-38: Status bar and mark for Horizontal control © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Marks Indicator ①The mark represents the current waveform window’s position in the memory. ②This mark displays the trigger position in the memory. ③This mark displays the trigger position in the current waveform windows. ④ Status bar displays the horizontal time base (main time base). ⑤ Status bar displays the trigger’s horizontal offset according to the center of the window. Key Points Y-T: The conventional oscilloscope display format.
RIGOL Delayed Scan: The Delayed Scan is a magnified portion of the main waveform window. You can use Delayed Scan to locate and horizontally expand part of the main waveform window for a more detailed (higher horizontal resolution) analysis of signal. Use Delayed Scan to expand a segment of a waveform to see more details. The Delayed Scan time base setting cannot be set slower than the Main time base setting.
RIGOL and position of the Delayed Scan. The symbol at bottom of the screen means the main time base and the symbol on the center bottom means the Delayed Scan time. z Use the horizontal portion. z Turn the horizontal z To change the main time base, you must turn off the Delayed Scan mode. z Since both the main and Delayed Scan are displayed; there are half as many vertical divisions so the vertical scaling is doubled. Notice the changes in the status bar.
RIGOL X-Y Format This format is useful for studying phase relationships between two signals. Channel 1 in the horizontal axis(X) and channel 2 in the vertical axis(Y), the oscilloscope uses a none-trigger acquisition mode, data is displayed as dots. . Figure 2-40: X-Y display format The following modes or functions will not work in X-Y format.
RIGOL Understand the Trigger System The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms. When the oscilloscope starts to acquire a waveform, it collects enough data so that it can draw the waveform to the left of the trigger point. The oscilloscope continues to acquire data while waiting for the trigger condition to occur.
RIGOL Trigger Modes The oscilloscope provides seven trigger modes: Edge, Pulse, Slope, Video, Alternative, Pattern (only for mixed signal oscilloscopes) and Duration trigger (only for mixed signal oscilloscopes). Edge: An edge trigger occurs when the trigger input passes through a specified voltage level in the specified slope direction. Pulse: Use this trigger type to catch pulses with certain pulse width. Video: Use video trigger on fields or lines for standard video signals.
RIGOL Settings for Edge Trigger An edge trigger determines whether the oscilloscope finds the trigger point on the rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising edge, falling edge or rising & falling edge.
RIGOL Settings for Pulse Width Trigger Pulse trigger occurs according to the width of pulse. The abnormal signals can be detected through setting up the pulse width condition.
RIGOL Figure 2-44 Table 2-22 Menu Settings Auto Sweep Normal Single Set Up Comments Acquire waveform even no trigger occurred Acquire waveform when trigger occurred. When trigger occurs, acquire one waveform and then stop To go to Set Up menu, see table 2-38 Note: The Pulse width adjust range is 20ns ~ 10s. When the condition is met, it will trigger and acquire the waveform. 2-40 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Settings for Video Trigger Choose video trigger to trigger on fields or lines of NTSC, PAL, or SECAM standard video signals. Trigger coupling preset to DC.
RIGOL Figure 2-47 Table 2-25 (when the Sync is set as All lines, Odd field and Even field) Menu Settings Standard PAL/SECAM NTSC Set Up Comments Select Video standard To go to set up menu, see table 2-39 Key points Sync Pulses: When Normal Polarity is selected, the trigger always occurs on negative-going sync pulses. If the video signal has positive-going sync pulses, use the inverted Polarity selection. Figure 2-48: Line Synchronization 2-42 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Figure 2-49: Video Trigger: Field Synchronization © Copyright RIGOL Technologies, Inc. 2007.
RIGOL SLOPE TRIGGER Slope trigger sets the oscilloscope as the positive/negative slope trigger within the specified time Figure 2-50 Table 2-26 ( Page One) Menu Settings CH1 Source CH2 EXT EXT/5 Set Set Set Set channel 1 as trigger source channel 2 as trigger source EXT.
RIGOL Note: Slope time can be set from 20ns to 10s. When a signal meets the trigger condition, scope will execute the acquisition. You can adjust LEVEL A/ LEVEL B or both simultaneous by turning the knob. © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Alternative Trigger When alternative trigger is on, the trigger sources are from two vertical channels. This mode can be used to observe two non-related signals. You can choose two different trigger modes for the two vertical channels. The options are as follows: Edge, Pulse, Slope and video. The info of the trigger level of the two channels will be displayed on the upper-right of the screen.
RIGOL Figure 2-53 Table 2-29 (Trigger Type: Pulse Menu Select Type Page One ) Settings Comments Set trigger mode for Channel 1 Set trigger mode for Channel 2 CH1 CH2 Set Pulse channel Pulse Trigger for the (+Pulse width When Figure 2-54 Table 2-30 Menu Setting Set Up less than) (+Pulse width more than) (+Pulse width equal to) (-Pulse width less than) (-Pulse width more than) (-Pulse width equal to) (Trigger Type: Pulse To select pulse condition Page Two) Settings Comments
RIGOL Figure 2-55 Table 2-31 (Trigger Type: Slope Menu Select Type Settings CH1 CH2 Slope When Figure 2-56 Time Vertical Set Up 2-48 Comments Set trigger mode for Channel 1 Set trigger mode for Channel 2 Set Slope Trigger for the vertical channel Set trigger condition Table 2-32 (Trigger Type: Slope Menu Page One) Settings
RIGOL Figure 2-57 Table 2-33 (Trigger Type: Video Menu Select Type Settings Comments CH2 Set trigger mode for Channel 1 Set trigger mode for Channel 2 Video Video Trigger for the channel CH1 Normal polarity Polarity Inverted polarity Figure 2-58 Page One) Triggers on negative going sync pulses Triggers on positive going sync pulses Table 2-34 (Trigger Mode: Video Menu Sync Line Num Standard Set Up Page two) Settings Comments ALL lines Line Num Trigger on all lines Trigger on an specifie
RIGOL Pattern Trigger (Mixed Signal Oscilloscope) Pattern trigger identifies trigger terms by checking appointed code. The code is logical relationship of all channels, with high (H), low (L) and ignore(X).
RIGOL Duration Trigger (Mixed Signal Oscilloscope) Trigger in appointed time when code terms are satisfied.
RIGOL Key Points: H (High): Logic high: voltage is higher than threshold setting. L (Low): Logic low: voltage is lower than threshold setting. X (Ignore): Don’t care. If all the channels are ignored, the oscilloscope won’t be triggered. Qualifier: A timer begins when code terms are satisfied. Duration trigger occurred in the time set by the qualifier. 2-52 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Trigger Settings In trigger set up menu, you can set up different trigger settings according to different trigger modes. When in the mode of Edge and Pulse, the source is D15-D0 (Mixed Signal Oscilloscope), only Holdoff is adjustable. When source is non-digital channel and in slope trigger, only trigger coupling, trigger sensitivity and Holdoff can be set. For video trigger, Sensitivity and Holdoff can be set.
RIGOL Figure 2-64 Table 2-40 (Settings only for holdoff) Menu Holdoff Holdoff Reset 2-54 Settings Comments Set time slot before another trigger event Reset Holdoff time to 100ns © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Trigger Holdoff Trigger Holdoff can stabilize complex waveform, such as the pulse range. Holdoff time is the oscilloscope’s waiting period before starting a new trigger. During Holdoff, oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first pulse on a group of them, users can set the holdoff time to Pulse cluster width. Trigger site Trigger holdoff Time of holdoff Figure 2-65: Trigger Holdoff To use trigger Holdoff: 1.
RIGOL Trigger Key points 1. Trigger Source: Trigger can occur from several sources: Input channels (CH1, CH2 and D15-D0 on mixed signal oscilloscopes), AC Line, Ext, Ext/5. z CH1 or CH2: It is the most commonly used trigger source. The channel will work when selected as a trigger source whatever displayed or not. z Ext Trig: The instrument can trigger from a third source while acquiring data from CH1 and CH2.
RIGOL Trigger, such as the output of Power supply. NOTE: When horizontal control is set under 50 ms/div, Auto mode allows the oscilloscope don’t capture trigger signal. z Normal: The Normal mode allows the oscilloscope to acquire a waveform only when it is triggered. If no trigger occurs, the oscilloscope keeps waiting, and the previous waveform, if any, will remain on the display. z Single: In Single mode, after pressing the RUN/STOP key, the oscilloscope waits for trigger.
RIGOL Adjustable trigger sensitivity z To avoid the influence of noise from the physical world, and get the stable trigger, the trigger circuit has adopted Stickiness. In DS1000 series, the stickiness is adjustable from 0.1div-1.0div, which means when it sets to 1.0div, the trigger circuit will not affect any signal with peak-peak amplitude less than 1.0div, so as to avoid the influence of the noise. 2-58 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL To Set Up the Sampling System As Figure 2-66 displays, the Acquire button for sampling system is at the MENU Acquire setup button Figure 2-66 Use Acquire button to pop up the menu as follows: Figure 2-67 Table 2-41 Menu Settings Comments Acquisition Normal Average Peak Detect Normal Acquisition mode Average Acquisition mode Peak Detect Acquisition mode Averages 2 to 256 Step by multiple of two.
RIGOL The waveform displayed on the screen will change in conjunction with the setting of Acquire menu. Figure 2-68: Signal that contains noise, and without average sampling Figure 2-69: Display signal after average sampling Note: z Select Real-time acquisition to observe the single-shot or pulse signals. z Select Equ-Time to observe high frequency repetitive signals. 2-60 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL z To reduce the displayed random noise, select the Average Acquisition. And this mode would make the screen refresh slower. z To Avoid signal aliasing, select Peak Detect Acquisition. Figure 2-70: Signal with Peak Detect Acquisition The Peak Detect effect is shown as the figure above.
RIGOL Key Points Real-time Sampling: DS1000 has Real-time sampling rate up to 400MSa/s. At the time base 50ns or faster, the oscilloscopes use the sine(x)/x interpolation to expand the horizontal time base. Equivalent sampling: Also known as Repetitive sampling. In this mode, you can get up to 40 ps of horizontal resolution (equivalent 25Gsa/s). This mode is good for observing repetitive signals, and it is not recommended for single-shot or pulse.
RIGOL To Set Up the Display System Figure 2-71 shows the menu button for the display system on the front panel. Figure 2-71 Display setup button Press the Display button to pop up the menu for the settings of the display system.
RIGOL Figure 2-73 Table 2-43 Menu Settings Display grids and coordinates on the screen Turn off the grids Turn off the grids and coordinates Grid Brightness Menu Display Screen Comments < percentage > Set up grid brightness 1s 2s 5s 10s 20s Infinite Set the time before menu fades away. The menu will be hidden after the set time of last button pressing. Normal Invert Set to normal mode Set to invert color display mode Key points: Display type: Display type includes Vector and Dot.
RIGOL To Store and Recall Waveforms or Setups Figure 2-74 shows the menu button for the storage system on the front panel. Storage setup button Figure 2-74 Press the Storage button to pop up the menu for the settings of the storage system. The waveforms and setups in both internal memory and external memory can be stored and recalled. The waveform file, setup file, bitmap and CSV file in external memory can be created and deleted. System supports English/Chinese key in.
RIGOL For factory default setups, the menu as following: Figure 2-76 Table 2-45 Menu Storage Settings Waveform Setups Bit map CSV Factory Comments Store or recall waveform Store or recall instrument setups Create or delete bit map files Create or delete CSV files Recall factory setups Load Recall factory setups or files Disk Mana. Go to disk manage table 2-50) menu (see For CSV, the menu as following: Figure 2-77 Table 2-46 Menu Storage Data Depth Para Save External Disk Mana.
RIGOL For bitmap, the menu is as follows. Figure 2-78 Table 2-47 Menu Storage Para Save External Disk Mana.
RIGOL Internal Memory Press Storage→Internal to go to the following menu. Figure 2-79 2-68 Table 2-48 Menu Settings Comments Internal Int_00 (N) . . . Int_09 (N) Set up the location of files in internal memory Load Recall waveform files and setup files from the internal memory location Save Save waveform files and setup files to the internal memory location © Copyright RIGOL Technologies, Inc. 2007.
RIGOL External Memory Press Storage→External to go to the following menu. Figure 2-80 Table 2-49 Menu Explorer Settings Comments Path Directory File Switch among Path, Directory and File New File (Folder) To create new file or folder. Delete File(Folder) Delete file(Folder) Load Recall waveform and setup from USB storage device. File system as follows: Path Back to previous folder Current directory Files File info. Figure 2-81 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Disk Manage Press Storage→Disk Mana. to go to the following menu.
RIGOL Rename Press Storage→Disk Mana.→Rename to go to the following menu. Figure 2-84 Table 2-52 Menu Settings Comments To move the cursor up To move the cursor down To delete chosen letter OK Rename the file Rename system as following: Figure 2-85 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Factory The oscilloscope has default settings and can be recalled at anytime by user. Memory location Specify the memory location to save/recall the waveform and setup. Load Recall saved waveform, setup and default settings. Save Save waveform and setups. NOTE: 1. Select Save the oscilloscope stores not only the waveforms, but also the current settings of 2. Only after the settings are changed for more than 5 seconds can the user turn off the instrument.
RIGOL To Set Up the Utility Figure 2-86 shows the menu button for the Utility on the front panel. Figure 2-86 System function setup button Press the Utility button to pop up the menu for the settings of the Utility system.
RIGOL Figure 2-88 Table 2-54 Menu Figure 2-89 Settings Comments Pass/Fail Setup Pass/Fail test Record Setup Waveform Recorder Print set Setup printing Table 2-55 Menu Settings ON OFF Fast-Cal Comments Turn on fast-calibration Turn off fast-calibration Self-Cal Execute Self-calibration Service Go to service menu Preferenc Go to preference menu Note: Fast-Cal: When turn on the fast-cal function, oscilloscope will calibrate vertical offset several times according to the running time.
RIGOL The I/O setup: Press Utility→I/O setting to go to the following menu. Figure 2-90 Table2-56 Menu RS-232 Baud Display 300 . . Comments Set RS-232 baud rate as 300, 2400, 4800, 9600, 19200 or 38400. 38400 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Preference Press Utility→Preference to go to the following menu Figure 2-91 Table 2-57 Menu Display Comments Screen saver 1 min . . 5 hour OFF Set up screen saver timer Expand Refer. Ground Center Set up waveform expand reference Set up sticky state of CH1, CH2, Math, Ref, Trig Level and Trig Offset Sticky key Skin vertical Classical Modern Tradition Succinct Set up skin style NOTE: Screen saver: This function can extend the life of LCD backlight tube.
RIGOL Self-Calibration The Self-Calibration adjusts the oscilloscope’s internal circuitry to get the best accuracy. Use these functions to calibrate the oscilloscope’s vertical and horizontal systems. For maximum accuracy at any time, run this calibration if the ambient temperature changes by 5° C or more. Before running this procedure, do these steps: 1. Disconnect any probes or cables from all channel inputs, otherwise failure or damage to the oscilloscope may occur. 2.
RIGOL Pass/Fail The pass/fail function can monitor changes of signals and output pass or fail signals by judging the input signal is within the pre-defined mask or not. Press Utility→Pass/Fail to go to the following menu.
RIGOL Mask setting Press Utility→Pass/Fail→Mask Setting to go to the following menu. Figure 2-95 Table 2-60 Menu X Mask Y Mask Settings < x div> < y div> Create Mask Location Figure 2-96 Comments Set horizontal clearance waveform (0.04div-4.00div) to the Set vertical clearance to the waveform (0.04div-4.00div) Create a test mask according to the above clearance Internal External Set the memory location of the mask files.
RIGOL Figure 2-97 Table 2-62 when the save as external memory Menu 2-80 Settings Comments Save Go to save menu (same as REF save menu. See table 2-12) Load Go to load menu see table 2-63 Import Go to import menu. (same as REF import menu see table 2-14) © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Load Press Utility→Pass/Fail→Mask Setting→Load to go to the following menu. Figure 2-98 Table 2-63 Menu Explorer Load Settings Path Directory File Comments Switch among Path, Directory and File Recall the specified file NOTE: Pass/Fail function is unavailable in X-Y mode. © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Pass/Fail connection The Pass/Fail connection has a optically isolated output. Users need to connect to another circuit to fulfill the function. Before connecting external circuit, make sure the maximum voltage/current do not exceed the internal photoMOS relay’s rating, 400V/100mA. DS1000 series adopts the technique of Optical Isolation, output device has no polarity and can be used in any circuit within the ratings.
RIGOL Print Set DS1000 series oscilloscopes support external printers. Press Utility→Print set to go to the following menu. Figure 2-100 Table 2-64 Menu Settings Print Comments Execute print function Inverted ON OFF invert the color for print print original color Palette Gray scale Color Set up print color © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Waveform recorder Waveform recorder can record input waveform from CH1 and CH2, with a maximum record length of 1000 frames. This record performance can also be activated by the Pass/Fail test output, which makes this function especially useful to capture abnormal signals in long term without keeping an eye watching it. Press Utility→Record→Mode→Record to go to the following menu Waveform recorder: Record the waveforms with specified interval.
RIGOL Play back: Play back the recorded waveforms. Figure 2-102 Table 2-66 Menu Settings (Run) Operate (Stop) Figure 2-103 Play stopped, press Start playback Playing, press to stop to Set repeat play mode Set single time play mode Play mode Interval Comments <1.
RIGOL Storage: Store recorded waveforms to non-volatile memory according to setup frames. Figure 2-104 Table 2-68 Menu Start Frame End Frame <1-1000> Comments Set first frame to be saved Set last frame to be saved Set up Store location Table 2-69 when the save to Internal memory Menu 2-86 <1-1000> Internal external Location Figure 2-105 Settings Settings Comments Save Save recorded waveform to internal memory location Load Recall recorded waveform internal memory location Imp./Exp.
RIGOL Figure 2-106 Table 2-70 when the save as external memory Menu Settings Comments Save Go to save menu (same as REF save menu. See table 2-12) Load Go to load menu see table 2-63 Import Go to import menu. (same as REF import menu see table 2-14) © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Service Press Utility→Service to go to the following menu. Figure 2-107 Table 2-71 Menu Settings System Info Press to display system information. Screen Test Press to run screen test program. Color Test Press to run color test Key Test Press to run keypad test program. 1. System Info: Press this soft button to display the information of oscilloscope. It contains Model, Power up times, Serial No., Software version and Installed module of the oscilloscope.
RIGOL NOTE: 1. T he corresponding block on the screen will temporary turn into green (Color models) or white (Mono models) when pressing a key. 2. The blocks will reverse its color to red (Color models) or black (Mono models) when the corresponding key or knob get pressed, and the color will keep reversed during the test. 3. Key backlight lamps will be lit one by one to determine if every backlight LED works well. 3.
RIGOL Language: The DS1000 series oscilloscopes have multi-language user menu, choose as your desire. Press Utility→Language to select the language. 2-90 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL To Measure Automatically The Measure button in menu area activates the automatic measurement function. The instruction below will guide you using the DS1000 series powerful measurement function. Auto measurement Figure 2-108 Menu explanation: Press the Measure button to display the menu for the settings of the Automatic Measurements.
RIGOL Page 1 for Voltage Measurements Figure 2-110 Table 2-73 Menu Settings Comments Vmax Measure maximum voltage of a waveform Vmin Measure minimum voltage of a waveform Vpp Measure Peak-to-Peak Voltage Vtop Measure a flat top voltage of a square waveform Page 2 for voltage measurements Figure 2-111 Table 2-74 Menu 2-92 Settings Comments Vbase Measure a flat base voltage of a square waveform Vamp Measure voltage between Vtop and Vbase Vavg Measure average voltage of a waveform Vrms
RIGOL Page 3 for Voltage Measurements Figure 2-112 Table 2-75 Menu Settings Comments Overshoot Measure overshoot in percentage of an edge Preshoot Measure preshoot in percentage of an edge Page 1 for Time Measurements Figure 2-113 Table 2-76 Menu Settings Comments Period Measure Period of a waveform Freq Measure Frequency of a waveform Rise time Measure Rise Time of a rising edge Fall time Measure Fall Time of a falling edge Page 2 for Time Measurements Figure 2-114 Table 2-77 Menu Se
RIGOL Page 3 for Time Measurements Figure 2-115 Table 2-78 Menu Settings Comments Delay1→2 Measure the signals delay between two channels at the rising edge Delay1→2 Measure the signals delay between two channels at the falling edge NOTE: The results of the automatic measurements will be displayed on the bottom of the screen. Maximum 3 results could be displayed at the same time. When there is no room, the next new measurement result will make the previous results moving left out of screen.
RIGOL Using Automatic Measurement as the following steps demonstrate: 1. Select the signal channel for measuring. You could select CH1 or CH2 according to the signal you are interested in. Press soft buttons as follows: Measure→Source→CH1 or CH2. 2. To see all measurement value, set the Display All to ON. 18 measurement parameters will be displayed on the screen. 3. Select parameters page for measuring.
RIGOL The automatic measurement of voltage parameters The DS1000 series oscilloscopes provide automatic voltage measurements including Vpp, Vmax, Vmin, Vavg, Vamp, Vrms, Vtop, Vbase, Overshoot and Preshoot. Figure 2-117 below shows a pulse with some of the voltage measurement points. Overshoot Vmax Vtop Vpp Vamp Vbase Vmin Preshoot Figure 2-117 Vpp: Peak-to-Peak Voltage. Vmax: The maximum amplitude. The most positive peak voltage measured over the entire waveform. Vmin: The minimum amplitude.
RIGOL The automatic measurement of time parameters The DS1000 series oscilloscopes provide time parameters auto-measurements include Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay 1→2 , Delay 1→2 , +Duty and -Duty. Figure 2-118 shows a pulse with some of the time measurement points. Rise Time Fall Time 90% 50% 10% +Width -Width Figure 2-118 Rise Time: Time that the leading edge of the first pulse in the waveform takes to rise from 10% to 90% of its amplitude.
RIGOL To measure with cursors Figure 2-119 displays the Cursor button on the front-panel. Figure 2-119 Cursor measurement The cursor measurement has three modes: Manual, Track and Auto Measure. 1. Manual: In this mode, the screen displays two parallel cursors. You can move the cursors to make custom voltage or time measurements on the signal. The values are displayed on the boxes below the menu.
RIGOL Menus and operations of Cursor Measurements 1. Manual Mode Figure 2-120 Table 2-79 Menu Mode Settings Comments Manual Adjust the cursor to measure X/Y parameters manually X Type Y Source CH1 CH2 MATH FFT LA Shown as vertical line to measure the horizontal parameters Shown as horizontal line to measure the vertical parameters Select the measurement signal source.
RIGOL Table 2-80 Cursor Increment X Cursor A Y X Cursor B Y Operation Turn the multifunctional knob ( cursor A horizontally Turn the multifunctional knob ( cursor A vertically ) to move Turn the multifunctional knob ( cursor B horizontally Turn the multifunctional knob ( cursor B vertically ) to move ) to move ) to move NOTE: Cursor could be moved only when the curse function menu is displayed. 5.
RIGOL Key Points Cursor Y: Cursors Y appear as horizontal lines on the display and measure the vertical parameters. Usually it is used to measure the Volts. When the source is set as function, the units are assigned to the function. Cursor X: Cursors X appear as vertical lines on the display and measure the horizontal parameters. Usually is indicates the time of trigger excursion. When the source is set as FFT, X means frequency © Copyright RIGOL Technologies, Inc. 2007.
RIGOL 2.
RIGOL NOTE: Horizontally moving cursor is not possible when other(not tracking cursor) menu is activated. (4) To get measurement value: Position of Cursor 1 (Time cursor centered on the midpoint of screen; Voltage cursor centered on channel ground level). Position of Cursor 2 (Time cursor centered on the midpoint of screen; Voltage cursor centered on channel ground level). Horizontal space between cursor 1 and 2 (△X): Time between cursors, units in Second. (1/△X), units in Hz, kHz, MHz, GHz.
RIGOL 3. Auto mode: Figure 2-122 Table 2-83 Menu Mode Settings Comments Auto Display the cursors for the current automatic measuring. (See the following figure) Figure 2-123: Auto Measure Mode of Cursor Measurement There will be no cursor display if no parameter were chosen in Measure menu. The oscilloscope could move cursor automatically to measure 20 parameters in Measure menu. 2-104 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL To Use Run Control Buttons The run control buttons include AUTO (auto setting) and RUN/STOP. Auto: The AUTO features automatic adjustments to produce a usable display of the input signal. When AUTO button is pressed, the following menu will appear.
RIGOL Auto-set functions After the AUTO is pressed the oscilloscope is configured to the following defaults: Table 2-85 Menu Settings Display format Y-T Acquire mode Normal Vertical coupling Adjust to AC or DC according to the signal. Vertical “V/div” Adjusted Volts/Div Coarse Bandwidth limit Full Signal Invert OFF Horizontal position Center Horizontal “S/div” Adjust to right position Trigger type Edge Trigger source Find the channel with input signal automatically.
RIGOL Chapter 3 : Application & Examples Example 1: Taking Simple Measurements The function is used to observe the unknown signal and display and measure the frequency and peak-to-peak amplitude. To quickly display a signal, please do the steps as follow: 1. Set the probe and the channel attenuations to 10X 2. Connect signal to CH1 with the probe 3. Press the AUTO button The oscilloscopes set the vertical, horizontal, and trigger controls at the best status automatically.
RIGOL Example 2: View a Signal Delay Caused by a Circuit This example is to test the input and output signals of a circuit and observe the signal delay. First, set the probe and the channel attenuation to 10X and connect CH1 probe to the input, CH2 to the output of the circuit. Do these steps as follow: 1. Display the signals (CH1 and CH2): 1) Press the AUTO button 2) Adjust the vertical and the horizontal scale by turning the appropriate ranges for display.
RIGOL Example 3: Capture a Single-Shot Signal To capture a single event, you need to gather some pre-test knowledge of the signal in order to correctly set up the trigger level and slope. For example, if the event is derived from TTL logic, a trigger level of 2 volts should work on a rising edge. The following steps show you how to use the oscilloscope to capture a single event. 1. Set the probe and the channel attenuations to 10X. 2. Set up the trigger.
RIGOL Example 4: To Reduce the Random Noise on a Signal If the signal applied to the oscilloscope is noisy (Figure 3-2), you can set up the oscilloscope to reduce the noise on the waveform and avoid its interference to the signal itself. Figure 3-2 1. Set the probe and the channel attenuations to 10X. 2. Connect a signal to the oscilloscope and obtain a stable display. 3. Improve the trigger by setting the Coupling (1) Press the MENU in the Trigger control area.
RIGOL (1) If there is noise with the signal and the waveform looks too wide, in this case, you can choose average acquisition. In this mode the waveform will be thin and easy to observe and measure. To use average follow these steps. z Press soft button as Acquire→Acquisition→Average z Toggle the Averages soft button to select the number of averages that best eliminates the noise from the displayed waveform. It can be adjusted from 2256.
RIGOL Example 5: Making Cursor Measurements There are 20 most desired parameters that could be automatically measured with the oscilloscope. These parameters can also be measured using cursors. You could use the cursors to quickly make time and voltage measurements on a waveform. Measuring the Frequency of the First Sinc Waveform Peak To measure the ring frequency at the rising edge of a signal, do these steps: 1. Press Cursor key to see the Cursor menu. 2. Press Mode to set Manual mode. 3.
RIGOL Measure the Amplitude of the First Waveform Peak of the Sinc. Now let’s measure the amplitude of the Sinc. Please follow these steps: 1. Press Cursor key to see the Cursor menu. 2. Press Mode to set Manual mode 3. Press Type to select Y. 4. Turn ( ) knob to place cursor A on the first peak of the wave. 5. Turn ( ) knob to place cursor B on the second peak of the wave.
RIGOL Example 6: The application of the X-Y operation Viewing Phase Changes through a Network Theme: Connect the oscilloscope to monitor the input and output of the circuit and capture the phase changes through the circuit. To view the input and output of the circuit in an X-Y display, do these steps: 1. Set the probe menu attenuation to 10X. Set the switch to 10X on the probes. 2. Connect the channel 1 probe to the input of the network, and connect the channel 2 probe to the output. 3.
RIGOL Signal must be centered in X D A B C Figure 3-6 Ellipse method to observe the phase difference Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals. From the formula above, you could get: θ = ±arcsine (A/B) or ±arcsine (C/D) If the main axis of the ellipse is at I and III quadrant, θ must be in the range of (0~π/2) or (3π/2~2π). If the main axis is at II and IV quadrant, θ must be in the range of (π/2~π) or (π~3π/2). © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Example 7: Triggering on a Video Signal Test a video circuit in the DVD set. Use video trigger to obtain a stable display. Triggering on Video Fields To trigger on the video fields, please do as the following steps: 1 Press the MENU key in the Trigger control area to see the Trigger menu. 2 Press Mode to select Video mode 3 Press Source to select CH1 as trigger source 4 Press Polarity to select 5 Press Sync as Odd Field or Even Field 6 Adjust the trigger.
RIGOL Triggering on the Video Lines 1. Press the MENU key in the triggering control area to see the trigger menu. 2. Press Mode to select Video mode 3. Press Source to select CH1 as trigger source 4. Press Polarity to select 5. Press Sync to select Line Num 6. Turn ( ) knob to trigger on a specified line number 7. Adjust the to set trigger level at the video sync pulse to get stable trigger. 8.
RIGOL Example 8: FFT Cursor measurement FFT measurements include: amplitude measurement (Vrms or dBVrms) and Frequency measurement (Hz). Do these steps as follow: 1. Press Cursor→Manual 2. Press Type to select X or Y 3. Press Source to select FFT 4. Turn ( ) knob to move the cursor to an interesting point. Figure 3-9 Figure 3-10 3-12 © Copyright RIGOL Technologies, Inc. 2007.
RIGOL Example 9: Pass/Fail Test The Pass/Fail Test is one of enhanced special functions based on DS5000 series. When running test function the oscilloscope automatically examines the input signal, compared to the already built waveform mask. If the waveform “touches” the mask, a “Fail” occurs, otherwise the test passes. When needed, a programmable output can be used for external automatic control applications, such as the automatic failure product kick-out in a production line.
RIGOL Example 10: Triggering on a Digital Signal The triggering way, especially to the Digital Signal, which the Pattern and Duration triggering are use for to carry on when triggering system was being used. The two triggering ways are only allowed to a digital signal. Pattern Do the steps as following: 1. Press the MENU key in the triggering control area to see the trigger menu 2. Press Mode to select Pattern mode 3. Turn ( ))knob to select the channel you want to set 4.
RIGOL Duration Do the steps as following: 1. Press the MENU key in the triggering control area to see the trigger menu 2. Press Mode to select Duration mode 3. Turn ( ))knob to select the channel you want to set 4. Press Code to set Code Setting (H, L, or X) 5. Press Qualifier to set time limit terms 6. Press Time to set duration and limit symbol time 7. Press Sweep to select trigger mode to: Auto, Normal, or Single 8. Press Set Up to adjust the Holdoff time.
RIGOL Chapter 4 : Prompt messages & troubleshooting Prompting Message Setting at limit: Prompts that the setting is at limit, can not be adjusted any further. Trigger level at limit: Prompts that the trigger level is at limit while turning the knob. Trigger position at limit: Prompts that the trigger position is at the start or the end point of the memory space while turning the horizontal adjust the trigger offset.
RIGOL Memory position at limit: Prompts that the Memo Offset is at the full range of the memory. Save finished: Prompts that the storage progress has finished. The storage is empty: Prompts that current position of the memory is empty to recall saved waveforms or settings. Measurement already selected: Prompts that the measurement parameter you pressed is already displayed on screen. Dot display only: Prompts that you could only use Dots type for display under this setting.
RIGOL Troubleshooting 1. After the oscilloscopes is powered on, the screen remains dark, please inspect the instrument following these steps: (1) (2) (3) (4) Check the power cable’s connection Ensure the power switch is turned on After the inspections above, restart the oscilloscope If the problem still remains, please contact RIGOL for help 2.
RIGOL the trigger level is out of the signal range. If yes, you should set the trigger level in proper range by turning the knob or pressing the 50% button. You could also set the Trigger Mode as “AUTO”. Moreover, you could push AUTO button to display the waveform on screen. 6. After the Acquisition is set to “Averages” or Display Persistence is set ON, the waveform refreshes slowly. It is normal in these settings. 7. The signal is displayed as ladder like waveform. (1) The time base maybe is too slow.
RIGOL Chapter 5 : Support & Service Warranty (DS1000 Series Digital Oscilloscopes) RIGOL warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment from an authorized RIGOL distributor. If a product proves defective within the respective period, RIGOL will provide repair or replacement as described in the complete warranty statement.
RIGOL Contact RIGOL If you meet any inconvenience during the use of our products, please contact RIGOL Technologies, Inc. or contact your local distributor. Domestic: Please call Tel: (86-10)80706688 Fax: (86-10)80720067 9:00 am –5: 00 pm from Monday to Friday Or by e- mail: support@rigol.com Or mail to: RIGOL Technologies, Inc. 156# CaiHe Village, ShaHe Town, ChangPing District, Beijing, China Post Code: 102206 Overseas: Contact local RIGOL distributor or sales office.
RIGOL Chapter 6 : Appendix Appendix A: Specifications All specifications apply to the DS1000 Series Oscilloscopes and a probe with the Attenuation switch set to 10X unless noted otherwise. To meet these specifications, two conditions must first be met: 1 The instrument must have been operating continuously for thirty minutes within the specified operating temperature. 2 You must perform the Self Cal operation, accessible through the Utility menu, if the operating temperature changes by more than 5 °C.
RIGOL Horizontal Sample Rate Range 1Sa/s-400MSa/s (Real-Time), 25GSa/s (Equivalent) Waveform interpolation Sin(x)/x Record Length 1M samples for single channel, 512K samples for each channel 512K samples ** Scan speed Range (Sec/div) 5ns/div-50s/div, DS1102XX, DS1062XX 10ns/div-50s/div, DS1042XX 20ns/div-50s/div, DS1022XX 1-2-5 Sequence Sample Rate and Delay Time Accuracy ±100ppm (over any Delta Time Measurement Accuracy (Full Bandwidth) Single-shot: ±(1 sample interval + 100ppm × reading + 0.
RIGOL Vertical A/D converter 8-bit resolution, each channel samples simultaneously* Volts/div Range 2mV/div-5V/div at input BNC Offset Range ±40V(200mV-5V), Analog Bandwidth 100MHz(DS1102CD, DS1102C, DS1102MD, DS1102M) 60MHz(DS1062CD, DS1062C, DS1062MD, DS1062M) 40MHz(DS1042CD, DS1042C, DS1042MD, DS1042M) 25MHz(DS1022CD, DS1022C, DS1022MD, DS1022M) Single-shot Bandwidth 80MHz(DS1102CD, 60MHz(DS1062CD, 40MHz(DS1042CD, 25MHz(DS1022CD, Selectable Analog Bandwidth Limit (typical) 20MHz Lower Freque
RIGOL Trigger Trigger Sensitivity 0.1div-1.0div (adjustable) Trigger Level Range Internal ±12 divisions from center of screen EXT ±1.6V EXT/5 ±8V Internal ±(0.
RIGOL Measurements Manual Voltage difference between cursors (∆V) Time difference between cursors (∆T) Reciprocal of ∆T in Hertz (1/∆T) Track Voltage value for Y-axis waveform Time value for Y-axis waveform Auto Cursors are visible for Automatic Measurement Cursor Auto Measure Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vavg, Vrms, Overshoot, Preshoot, Freq, Period, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty, Delay1→2 , Delay1→2 General Specifications Display Display Type 5.7 in.
RIGOL Environmental Ambient Temperature Operating 10℃~ 40℃ Cooling Method Fan force air flow Humidity +35℃ or below: ≤90% relative humidity Non-operating -20℃~ +60℃ +35℃~ +40℃: ≤60% relative humidity Altitude Operating 3,000 m or below Non-operating 15,000 m or below Mechanical Size Heavy Width 303mm Height 154mm Depth 133 mm Without package 2.4 kg Packaged 3.
RIGOL Appendix B: DS1000 Series Accessories Standard Accessories: 1 Probe×2 (1.5m), 1:1, (10:1) Passive Probes The passive probes have a 6MHz bandwidth with a rating of 150V CAT II when the switch is in the 1X position, and a Full oscilloscope bandwidth with a rating of 300 V CAT II when the switch is in the 10X position.
RIGOL Appendix C: General Care and Cleaning General Care Do not store or leave the instrument in where the LCD display will be exposed to direct sunlight for long periods of time. ! CAUTION: To avoid damage to the instrument or probes, do not expose them to sprays, liquids, or solvents. Cleaning If this instrument requires cleaning, disconnect it from all power sources and clean it with a mild detergent and water. Make sure the instrument is completely dry before reconnecting it to a power source.
RIGOL Index Number 50% ................................1-18, 2-36 A AC Coupling ............................... 2-4 AC Line.................................... 2-56 Accessories ........................... VI, 6-7 Acquire .................................... 2-59 Acquisition ................................. 6-1 Alternative Trigger.................... 2-46 Appendix ................................... 6-1 AUTO ..................1-12, 2-105, 2-106 Auto Trigger.............................
Measurements ............................ 6-5 N Normal Trigger ......................... 2-57 Nyquist Frequency .................... 2-15 O Optional ..................................... 6-7 P Pattern Trigger ......................... 2-50 Power ........................................ 6-5 Pretrigger................................. 2-57 Probe Attenuation ....................... 2-7 Probe compensation.................... 1-9 Probe Compensator .................... 6-5 Prompting messages ...................