RIGOL Quick Guide DS1000Z Series Digital Oscilloscope May.
RIGOL Guaranty and Declaration Copyright © 2013 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL Technologies, Inc. Publication Number QGA17101-1110 Notices RIGOL products are protected by patent law in and outside of P.R.C.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injuries or damages to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual. Use Proper Power Cord. Only the power cord designed for the instrument and authorized by local country could be used. Ground The Instrument.
RIGOL Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel. Keep Well Ventilation. Inadequate ventilation may cause increasing of temperature or damages to the device. So please keep well ventilated and inspect the intake and fan regularly.
RIGOL Safety Terms and Symbols Terms in this Manual. These terms may appear in this manual: WARNING Warning statements indicate the conditions or practices that could result in injury or loss of life. CAUTION Caution statements indicate the conditions or practices that could result in damage to this product or other property. Terms on the Product. These terms may appear on the Product: DANGER indicates an injury or hazard may immediately happen.
RIGOL Measurement Category Measurement Category DS1000Z series digital oscilloscopes can make measurements in Measurement Category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits.
RIGOL Ventilation Requirement This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation. WARNING Inadequate ventilation may cause temperature increase which would damage the instrument.
RIGOL Altitude Operating: less than 3 km Non-operating: less than 15 km Installation (overvoltage) Category This product is powered by mains conforming to installation (overvoltage) category II. WARNING Make sure that no overvoltage (such as that caused by thunderbolt) can reach the product, or else the operator might expose to danger of electric shock.
RIGOL conductive dust, rain, or snow. For example: outdoor locations. Safety Class Class 1 – Grounded Product General Care and Cleaning General Care: Do not store or leave the instrument at places where the instrument will be exposed to direct sunlight for long periods of time. Cleaning: Clean the instrument regularly according to its operating conditions. To clean the exterior surface: 1. Disconnect the instrument from all power sources. 2.
RIGOL Environmental Considerations The following symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002/96/EC on waste electrical and electronic equipment (WEEE). Product End-of-Life Handling The equipment may contain substances that could be harmful to the environment or human health.
RIGOL Contents Guaranty and Declaration .................................................................. I Safety Requirement........................................................................... II General Safety Summary .................................................................... II Safety Terms and Symbols ................................................................ IV Measurement Category ......................................................................
RIGOL SINGLE .................................................................................... 15 Knob ........................................................................................ 15 MENU ...................................................................................... 16 Print ........................................................................................ 17 User Interface ..................................................................................
RIGOL Quick Start General Inspection 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the unit. 2. Inspect the instrument.
RIGOL Appearance and Dimensions 2 Figure 1 Front View Unit: mm Figure 2 Top View Unit: mm DS1000Z Quick Guide
RIGOL To Prepare for Operation To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the instrument as well as easier operation and observation of the instrument. Figure 3 To Adjust the Supporting Legs To Connect to AC Power Supply This oscilloscope can accept 100-240 V, 45-440 Hz AC power supply.
RIGOL Power-on Inspection at When the oscilloscope is energized, press the power key the lower-left corner of the front panel to turn on the oscilloscope. During the start-up process, the oscilloscope performs a series of self-test items.
RIGOL circuit point under test. Figure 5 To Connect the Probe Function Inspection 1. Press Storage Default to restore the oscilloscope to its default configuration. 2. Connect the earth alligator clip of the probe to the “Ground Terminal” under the probe compensation signal output terminal. 3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the “Compensation Signal Output Terminal” of the probe.
RIGOL Figure 7 Square Waveform 6. Use the same method to test the other channels. If the square waveforms actually shown do not match that in the figure above, please perform “Probe Compensation”. WARNING To avoid electric shock during the use of probe, please make sure that the insulated wire of the probe is in good condition and do not touch the metallic part of the probe when the probe is connected to high voltage source.
RIGOL Probe Compensation When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy and error. The probe compensation procedures are as follows: 1. Perform step 1, 2, 3 and 4 of “Function Inspection”. 2. Check the displayed waveforms and compare them with the following figures.
RIGOL Front Panel Overview 1 10 2 11 3 12 4 13 5 6 14 7 8 9 15 16 17 Figure 9 Front Panel Overview 8 No. Description No.
RIGOL Rear Panel Overview 1 4 5 2 6 3 7 8 Figure 10 Rear Panel Overview 1. Handle Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down. 2. LAN Connect the instrument to the network via this interface for remote control. This oscilloscope conforms to the LXI-C class instrument standards and can quickly build test system with other instruments.
RIGOL trigger via this interface, connect the signal to a waveform display instrument to measure the frequency of the signal and the result is equal to the current capture rate. Pass/Fail: there will be output a negative pulse signal when failed waveforms are detected by the oscilloscope. Connect this signal to other control systems to view the test result conveniently. There will be output a 3.3 V CMOS high level when no failed waveform is detected by the oscilloscope.
RIGOL oscilloscope is 100-240 V, 45-440 Hz. Use the power cord provided with the accessories to connect the instrument to AC power. Then, you can press the power key at the front panel to start the instrument. Front Panel Function Overview VERTICAL CH1, CH2, CH3, CH4: analog input channels. The 4 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors.
RIGOL current signal status (such as the frequency, amplitude and phase). POSITION: modify the vertical position of the Vertical current channel waveform. Turn clockwise to increase the position and turn counterclockwise to decrease. During the modification, the waveform would move up and down and the position ) at the lower-left corner of the screen message (e.g. would change accordingly. Press down this knob to quickly reset the vertical position to zero.
RIGOL screen would change accordingly. Press down this knob to quickly reset the horizontal position (or the delayed sweep position). MENU: press this key to open the horizontal control menu under which to turn on or off the delayed sweep function, switch between different time base modes. SCALE: modify the horizontal time base. HORIZONTAL Turn clockwise to reduce the time base and turn counterclockwise to increase the time base.
RIGOL trigger level to zero point. MENU: press this key to open the trigger operation menu. This oscilloscope provides various trigger types. FORCE: press this key to generate a trigger signal forcefully. CLEAR Press this key to clear all the waveforms on the screen. If the oscilloscope is in “RUN” state, new waveforms will still be displayed. AUTO Press this key to enable the waveform auto setting function.
RIGOL RUN/STOP Press this key to set the state of the oscilloscope to “RUN” or “STOP”. In “RUN” state, the key is illuminated in yellow. In “STOP” state, the key is illuminated in red. SINGLE Press this key to set the trigger mode to “Single”. In single trigger mode, press FORCE to generate a trigger signal immediately. Knob Adjust waveform brightness: In non-menu-operation mode, turn this knob to adjust the brightness of waveform display. The adjustable range is from 0% to 100%.
RIGOL In menu operation, press any menu softkey and turn the knob to select the submenus under this menu and then press down the knob to select the current submenu. It can also be used to modify parameters and input filename. In addition, for DS1000Z-S models oscilloscope, in the source interface, press the corresponding menu softkey and then press the knob; the numeric keyboard will pop-up on the screen and you can input the desired parameter value and unit directly using this knob.
RIGOL Storage: press this key to enter file store and recall interface. The storable file types include picture, traces, waveforms, setups, CSV and parameter. Internal and external storage as well as disk management are also supported. Cursor: press this key to enter cursor measurement menu. The oscilloscope provides three cursor modes: manual, track and auto.
RIGOL User Interface DS1000Z provides 7.0 inches, WVGA (800*480) 160,000 color TFT LCD. 1 2 3 4 12 13 5 14 6 7 15 8 16 9 10 11 17 18 19 20 Figure 11 User Interface 1. Auto Measurement Items Provide 12 horizontal (HORIZONTAL) and 12 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate the corresponding measurement item. Press MENU continuously to switch between the horizontal and vertical parameters. 2.
RIGOL 3. Status Available states include RUN, STOP, T’D (triggered), WAIT and AUTO. 4. Horizontal Time Base Represent the time per grid on the horizontal axis on the screen. SCALE to modify this parameter. Use HORIZONTAL The range available is from 5 ns to 50 s. 5. Sample Rate/Memory Depth Display the current sample rate and memory depth of the oscilloscope. SCALE to modify this Use HORIZONTAL parameter. 6.
RIGOL 10. Trigger Source Display the trigger source currently selected (CH1-CH4 or AC Line). Different labels are displayed when different trigger sources are selected and the color of the trigger parameter area will change accordingly. denotes that CH1 is selected as the trigger For example: source. 11. Trigger Level at the right of the screen is the trigger level label and the trigger level value is displayed at the upper-right corner of the screen.
RIGOL bandwidth limit (e.g. ). 15. CH4 Vertical Scale Display the voltage value per grid of CH4 waveform vertically. SCALE to modify this parameter. Use VIRTICAL The following labels will be displayed according to the current channel setting: channel coupling (e.g. ) and bandwidth limit (e.g. ). 16. Message Box Display prompt messages. 17. Source1 Waveform Display the type of waveform currently set for Source1.
RIGOL The following symbols might be displayed in the menu: Denote that at the front panel can be used to turns on select parameter items. The backlight of when parameter selection is valid. Denote that you can use to adjust the parameter to select the parameter. and then press down to input desired paramrter Denote that press values directly using the pop-up numeric keyboard. turns on when parameter input is The backlight of valid. Denote that the current menu has several options.
RIGOL To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the function keys (including menu keys) at the front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is “Help Options” and you can use “Button” or “Index” mode to select. The right is “Help Display Area”. Help Options Help Display Area Figure 13 Help Information Button: Default mode.
RIGOL Troubleshooting The commonly encountered failures and their solutions are listed below. When you encounter those failures, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (acquisition method: Utility System System Info). 1. The screen is still dark (no display) after power on: (1) Check if the power is correctly connected. (2) Check whether the fuse is burned out.
RIGOL complies with the attenuation ratio of the probe. 4. There is waveform display but not stable: (1) Check the trigger signal source: check whether the Source item at the trigger panel complies with the signal channel actually used. (2) Check the trigger type: general signals should use “Edge” trigger and video signal should use “Video” trigger. Only when the proper trigger type is used, can the waveform be displayed stably.
RIGOL 7. Fail to connect PC through USB: (1) Check if the USB data cable is correctly connected to the oscilloscope and PC. (2) Check if the USB data is in good condition and if needed, restart the oscilloscope. 8. The USB disk can not be recognized: (1) Check if the USB disk can work normally. (2) Make sure that the USB disk being used is flash type. This oscilloscope does not support hardware type USB disk. (3) Make sure if the capacity of the USB disk is too large.