User`s guide

RIGOL Chapter 15 Accessibility Setting
15-8 MSO1000Z/DS1000Z User’s Guide
Self-calibration
The self-calibration program can quickly make the oscilloscope reach the best
working state to get the most precise measurement values. You can perform
self-calibration at any time especially when the change of the environment
temperature is up to or more than 5. Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.
Disconnect all the input channel connections, and then press Utility Self-Cal and
the self-calibration interface as shown in the figure below is displayed.
Figure 15-2 Self-calibration Interface
Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit to give up the self-calibration operation at any time and return to the
previous menu.
Note: Most of the keys are disabled during the self-calibration.
Print Setting
PictBridge is a new print standard. If your oscilloscope
and the printer both comply with the PictBridge
standard, you can connect the oscilloscope to the
printer using a USB cable to print the screen image
directly. Devices complies with the PictBridge standard
always have the icon as shown in the figure on the right
side.