RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-28  MSO1000Z/DS1000Z User’s Guide 
  Setup: the oscilloscope triggers when the setup time is less than the specified 
setup time. After selecting this type, press Setup to set the setup time and the 
range is from 8 ns to 1 s. 
  Hold: the oscilloscope triggers when the hold time is less than the specified 
hold time. After selecting this type, press Hold to set the hold time and the 
range is from 8 ns to 1 s. 
  StpHold: the oscilloscope triggers when the setup time and hold time are less 
than the specified times. After selecting this type, press Setup and Hold to set 
the setup time and hold time respectively and the range is from 8 ns to 1 s. 
Trigger Mode: 
Press Sweep to open the trigger mode list and select auto, normal or single. For the 
details, please refer to “
Trigger Mode”. The corresponding status light of the 
current trigger mode turns on. 
Trigger Setting: 
Press Setting to set the trigger parameters (trigger holdoff and noise rejection) 
under this trigger type. 
Trigger Level:  
Press DataSource and use TRIGGER   LEVEL to modify the trigger level of the 
data source channel. Press ClkSource and use TRIGGER    LEVEL to modify the 
trigger level of the clock source channel. For the details, please refer to the 
introduction of “
Trigger Level”. 