Chapter 6 MATH and Measurement    RIGOL 
MSO1000Z/DS1000Z User’s Guide    6-19 
Auto Measurement 
MSO1000Z/DS1000Z provides auto measurements of 33 waveform parameters and 
the statistics and analysis of the measurement results. What’s more, you can also 
use the frequency counter to realize more precise frequency measurement. 
Quick Measurement after AUTO 
When the oscilloscope is correctly connected and has detected valid input signal, 
press AUTO to enable the waveform auto setting function and open the following 
function menu: 
Single-period: set the oscilloscope to display single-period signal 
automatically. Measure the “Period” and “Frequency” of the 
current signal within a single period and display the measurement 
results at the bottom of the screen. 
Multi-period: set the oscilloscope to display multi-period signal 
automatically. Measure the “Period” and “Frequency” of the 
current signal within multiple periods and display the 
measurement results at the bottom of the screen. 
Rise Time: measure the “Rise Time” of the current signal and 
display the measurement result at the bottom of the screen. 
Fall Time: measure the “Fall Time” of the current signal and 
display the measurement result at the bottom of the screen. 
Back: return to the menu displayed at the last user setting. 
Undo: cancel the automatic settings and restore the parameters 
set by users lastly.  
Figure 6-2 Function Menu after AUTO  
Note: Waveform auto setting function requires that the frequency of sine is no lower 
than 41Hz; the duty cycle should be greater than 1% and the amplitude must be at 
least 20mVpp for square. Otherwise, the Waveform auto setting function may be 
invalid and the quick parameter measurement function displayed in the menu will 
also be unavailable. 