RIGOL  Chapter 13 Signal Source 
13-6  MSO1000Z/DS1000Z User’s Guide 
To Output Noise 
Press Src1Conf to open the waveform setting menu. Press Wave and select “Noise”. 
At this point, you can output a noise signal with the specified amplitude, offset and 
impendance. 
1.  Source 1 output 
Press Output to turn on or off the signal output. 
2.  Set the amplitude 
Press Amplitude to set the amplitude of the signal. For the setting method, 
please refer to the introduction in “
Parameter Setting Method”. When the 
impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the 
impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp. 
3.  Set the offset 
Press Offset to set the DC offset of the noise signal. For the setting method, 
please refer to the introduction in “
Parameter Setting Method”. When the 
impedance is HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.5 
V-the current amplitude/2); when the impedance is set to 50 Ω, the range is from 
(-1.25 V+the current amplitude/2) to (1.25 V-the current amplitude/2). 
To Output Built-in Waveform 
MSO1000Z/DS1000Z series oscilloscope provides 7 kinds of built-in waveforms, 
including Sinc, ExpRise, ExpFall, ECG, Gauss, Lorentz and Haversine. Press 
Src1Conf to open the waveform setting menu. Press Wave and select “Built-in”. At 
this point, you can set the output signal parameters according to the type of the 
built-in waveform selected. 
1.  Source 1 output 
Press Output to turn on or off the signal output. 
2.  Select built-in waveform 
Press Built-In to select any of the 7 built-in waveforms (as shown in the figures 
below). 