User Guide
Chapter 13 Signal Source    RIGOL 
MSO1000Z/DS1000Z User’s Guide    13-9 
Figure 13-9 Built-in Waveform-Haversine 
3.  Set the frequency 
Press Frequency to set the frequency of the signal. For the setting method, 
please refer to the introduction in “
Parameter Setting Method”. The range is 
from 100 mHz to 1 MHz. 
4.  Set the amplitude 
Press Amplitude to set the amplitude of the signal. For the setting method, 
please refer to the introduction in “
Parameter Setting Method”. When the 
impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the 
impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp.  
5.  Set the DC offset voltage 
Press Offset to set the DC offset of the signal. For the setting method, please 
refer to the introduction in “
Parameter Setting Method”. When the 
impedance is HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.5 
V-the current amplitude/2); when the impedance is set to 50 Ω, the range is from 
(-1.25 V+the current amplitude/2) to (1.25 V-the current amplitude/2). 
6.  Set the start phase 
Press StartPhase to set the start phase of the signal. For the setting method, 
please refer to the introduction in “
Parameter Setting Method”. The range is 
from 0°to 360°. 
7.  Align Phase 
Pressing AlignPhase will re-configure the two channels to output according to 
the preset frequency and phase. For two signals of which the frequencies are the 
same or are multiples, this operation can align their phases. Acquire the 
waveforms of the two channels using the oscilloscope and display the waveforms 
stably. After switching the channel status, the phase deviation between the two 
waveforms is changed. At this point, press AlignPhase and the phase deviation 
shown on the oscilloscope will restore to the current phase deviation between 
the two waveforms automatically. 