User Guide
RIGOL  Chapter 15 Accessibility Setting 
15-8  MSO1000Z/DS1000Z User’s Guide 
Power-off Recall 
You can set the system configuration to be recalled when the oscilloscope is powered 
on again after power-off. Press Utility  System  Power Set to select “Last” or 
“Default”. 
  Last: return to the setting of the system at last power-off. 
  Default: return to the factory setting of the system.  
Self-calibration 
The self-calibration program can quickly make the oscilloscope reach the best 
working state to get the most precise measurement values. You can perform 
self-calibration at any time especially when the change of the environment 
temperature is up to or more than 5℃. Make sure that the oscilloscope has been 
warmed up or operated for more than 30 minutes before the self-calibration.  
Disconnect all the input channel connections, and then press Utility  Self-Cal and 
the self-calibration interface as shown in the figure below is displayed. 
Figure 15-3 Self-calibration Interface 
Press Start and the oscilloscope will start to execute the self-calibration program. 
Press Exit to give up the self-calibration operation at any time and return to the 
previous menu. 
Note: Most of the keys are disabled during the self-calibration. 