User Guide
RIGOL  Chapter 2 To Set the Vertical System 
2-2  MSO1000Z/DS1000Z User’s Guide 
To Enable the Analog Channel 
MSO1000Z/DS1000Z provides 4 analog input channels (CH1-CH4). As the setting 
methods of the vertical systems of the four channels are the same, this chapter takes 
CH1 as an example to illustrate the setting method of the vertical system. 
Connect a signal to the channel connector of CH1 and then press CH1 in the vertical 
control area (VERTICAL) at the front panel to enable CH1. At this point, the channel 
setting menu is displayed at the right side of the screen and the channel status label 
at the bottom of the screen (as shown in the figure below) is highlighted. The 
information displayed in the channel status label is related to the current channel 
setting. 
After the channel is turned on, modify the parameters such as the vertical scale, 
horizontal time base, trigger mode and trigger level according to the input signal for 
easy observation and measurement of the waveform. 
Channel Coupling 
The undesired signals can be filtered out by setting the coupling mode. For example, 
the signal under test is a square waveform with DC offset. 
  When the coupling mode is “DC”: the DC and AC components of the signal under 
test can both pass the channel. 
  When the coupling mode is “AC”: the DC components of the signal under test 
are blocked. 
  When the coupling mode is “GND”: the DC and AC components of the signal 
under test are both blocked. 
Press CH1  Coupling and use   to select the desired coupling mode (the 
default is DC). The current coupling mode is displayed in the channel status label at 
the bottom of the screen as shown in the figure below. You can also press Coupling 
continuously to switch the coupling mode. 
  DC
    AC      GND 










