Instructions
RIGOL Chapter 2 Command System
2-104 DS1000Z-E Programming Guide
:MEASure:CLEar
Syntax
:MEASure:CLEar <item>
Description
Clear one or all of the last five measurement items enabled.
Parameter
Name
Type
Range
Default
<item>
Discrete
{ITEM1|ITEM2|ITEM3|ITEM4|ITEM5|ALL}
--
Explanation You can use the :MEASure:ITEM command to enable the desired parameters of the
37 waveform parameters. Note that the last five parameters are determined
according to the order in which you turned them on and they will not change as you
delete one or more measurement items.
You can send the
:MEASure:RECover command to recover the item(s) when a certain
item or all the items are cleared.
Example
:MEASure:CLEar ITEM1 /*Clear ITEM1*/
:MEASure:RECover
Syntax
:MEASure:RECover <item>
Description
Recover the measurement item which has been cleared.
Parameter
Name
Type
Range
Default
<item>
Discrete
{ITEM1|ITEM2|ITEM3|ITEM4|ITEM5|ALL}
--
Explanation You can use the :MEASure:ITEM command to enable the desired parameters of the
37 waveform parameters. Note that the last five parameters are determined
according to the order in which you turned them on and they will not change as you
delete one or more measurement items.
You can send the
:MEASure:CLEar command to clear the items again after one or all
of the measurement items are recovered.
:MEASure:ADISplay
Syntax :MEASure:ADISplay <bool>
:MEASure:ADISplay?
Description Enable or disable the all measurement function, or query the status of the all
measurement function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Explanation The all measurement function can measure the following 29 parameters of the
source at the same time:
Voltage Parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vupper, Vmid, Vlower,
Vavg, Vrms, Overshoot, Preshoot, Period Vrms, and Variance
Time Parameters: Period, Frequency, Rise Time, Fall Time, +Width, -Width, +Duty,
-Duty, tVmax, and tVmin
Other Parameters: +Rate, -Rate, Area, and Period Area.
The all measurement function can measure CH1, CH2, and the MATH channel at the
same time. You can send the
:MEASure:AMSource command to set the source of the
all measurement function.
Return
The query returns 1 or 0.